Clinical safety of magnetic resonance imaging in patients with implanted SynchroMed EL infusion pumps

被引:7
作者
Diehn, Felix E. [1 ]
Wood, Christopher P. [1 ]
Watson, Robert E., Jr. [1 ]
Mauck, William D. [2 ]
Burke, Michelle M. [3 ]
Hunt, Christopher H. [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] Mayo Clin, Dept Anesthesiol, Rochester, MN 55905 USA
[3] Mayo Clin, Dept Psychiat & Psychol, Rochester, MN 55905 USA
关键词
MRI; Safety; Spinal infusion pumps; Medtronic; SynchroMed; CHRONIC PAIN; TECHNICAL COMPLICATIONS; INTRATHECAL BACLOFEN; NEURAXIAL INFUSION; CANCER-PATIENTS; EFFICACY; MORPHINE; THERAPY; DEVICES; ORIGIN;
D O I
10.1007/s00234-010-0737-y
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Patients with implanted SynchroMed spinal infusion pumps (Medtronic, Inc., Minneapolis, MN) routinely undergo magnetic resonance imaging at our institution. In August 2008, Medtronic issued an urgent medical device correction report regarding several pumps. Because of the rare potential "for a delay in the return of proper drug infusion" and "for a delay in the logging of motor stall events," "a patient's pump must be interrogated after MRI exposure in order to confirm proper pump functionality." This is particularly important in patients receiving intrathecal baclofen, for whom a delay in return of proper pump infusion could lead to life-threatening baclofen withdrawal syndrome. The objective of this report is to present our experience and protocol of performing magnetic resonance imaging in patients with implanted SynchroMed EL pumps. We retrospectively reviewed records of 86 patients with implanted SynchroMed EL spinal infusion pumps who underwent 112 examinations on 1.5-T magnetic resonance imaging scanners from September 1, 1998 to July 7, 2004. No SynchroMed EL pumps were damaged by magnetic resonance imaging, and the programmable settings remained unchanged in all patients. Our data suggest that SynchroMed EL pump malfunction is indeed rare after routine clinical 1.5-T magnetic resonance imaging examinations. However, based on the Medtronic correction report, we perform pump interrogation before and after imaging.
引用
收藏
页码:117 / 122
页数:6
相关论文
共 21 条
[1]   Safety of active implantable devices during MRI examinations:: A finite element analysis of an implantable pump [J].
Buechler, Philippe ;
Simon, Anne ;
Burger, Juergen ;
Ginggen, Alec ;
Crivelli, Rocco ;
Tardy, Yanik ;
Luechinger, Roger ;
Olsen, Sigbjorn .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2007, 54 (04) :726-733
[2]   Abrupt withdrawal from intrathecal baclofen: Recognition and management of a potentially life-threatening syndrome [J].
Coffey, RJ ;
Edgar, TS ;
Francisco, GE ;
Graziani, V ;
Meythaler, JM ;
Ridgely, PM ;
Sadiq, SA ;
Turner, MS .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2002, 83 (06) :735-741
[3]  
CRUL BJP, 1991, REGION ANESTH, V16, P209
[4]   Efficacy and technical complications of long-term continuous intraspinal infusions of opioid and/or bupivacaine in refractory nonmalignant pain: A comparison between the epidural and the intrathecal approach with externalized or implanted catheters and infusion pumps [J].
Dahm, P ;
Nitescu, P ;
Appelgren, L ;
Curelaru, I .
CLINICAL JOURNAL OF PAIN, 1998, 14 (01) :4-16
[5]   Therapeutic applications of implantable drug delivery systems [J].
Dash, AK ;
Cudworth, GC .
JOURNAL OF PHARMACOLOGICAL AND TOXICOLOGICAL METHODS, 1998, 40 (01) :1-12
[6]   MRI safety review [J].
Dempsey, MF ;
Condon, B ;
Hadley, DM .
SEMINARS IN ULTRASOUND CT AND MRI, 2002, 23 (05) :392-401
[7]   Intrathecal drug therapy for chronic pain - From basic science to clinical practice [J].
Dougherty, PM ;
Staats, PS .
ANESTHESIOLOGY, 1999, 91 (06) :1891-1918
[8]  
Du Pen S, 1999, ONCOLOGY-NY, V13, P45
[9]   Alternative intrathecal agents for the treatment of pain [J].
Hassenbusch, SJ ;
Garber, J ;
Buchser, E ;
Du Pen, S ;
Nitescu, P .
NEUROMODULATION, 1999, 2 (02) :85-91
[10]  
Krames ES, 1999, ONCOLOGY-NY, V13, P37