Optimization of Contrast Material Delivery for Dual-Energy Computed Tomography Pulmonary Angiography in Patients With Suspected Pulmonary Embolism

被引:47
作者
Nance, John W., Jr. [2 ,3 ]
Henzler, Thomas [1 ]
Meyer, Mathias
Apfaltrer, Paul
Braunagel, Margarita
Krissak, Radko
Schoepf, Uwe J. [2 ]
Schoenberg, Stefan O.
Fink, Christian
机构
[1] Univ Heidelberg, Med Fac Mannheim, Univ Med Ctr Mannheim, Inst Clin Radiol & Nucl Med,Dept Clin Radiol & Nu, D-68167 Mannheim, Germany
[2] Med Univ S Carolina, Dept Radiol & Radiol Sci, Charleston, SC 29425 USA
[3] Med Univ S Carolina, Dept Family Med, Charleston, SC 29425 USA
关键词
dual-energy CT; iodine delivery rate; pulmonary perfusion; pulmonary embolism; MULTIDETECTOR-ROW CT; INJECTION PROTOCOL; ISO-OSMOLAR; ENHANCEMENT; LUNG; VISUALIZATION; PERFUSION;
D O I
10.1097/RLI.0b013e31821a2142
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To prospectively compare subjective and objective measures of image quality using 4 different contrast material injection protocols in dual-energy computed tomography pulmonary angiography (CTPA) studies of patients with suspected pulmonary embolism. Materials and Methods: A total of 100 consecutive patients referred for CTPA for the exclusion of pulmonary embolism were randomized into 1 of 4 contrast material injection protocols manipulating iodine concentration and iodine delivery rate (IDR, expressed as grams of iodine per second): Iomeprol 400 at 3 mL/s (IDR = 1.2 gI/s), iomeprol 400 at 4 mL/s (IDR = 1.6 gI/s), iomeprol 300 at 5.4 mL/s (IDR = 1.6 gI/s), or iomeprol 300 at 4 mL/s (IDR = 1.2 gI/s). Total iodine delivery was held constant. Dual-energy CTPA of the lungs were acquired and used to calculate virtual 120 kV CTPA images as well as iodine perfusion maps. Attenuation values in the thoracic vasculature and image quality of virtual 120 kV CTPAs were compared between groups. Iodine perfusion maps were also compared by identifying differences in the extent of beam-hardening artifacts and subjective image quality. Results: Protocols with an IDR of 1.6 gI/s provided the best attenuation profiles. CTPA image quality was greatest in the high concentration, high IDR (1.6 gI/s) protocol (P < 0.05 for all group comparisons) with no differences between the other groups (all P >= 0.05). Extent of beam-hardening artifacts and perfusion map image quality was significantly better using the high concentration, high IDR protocol as compared with all groups (P < 0.05 for all comparisons) and significantly worse using the low concentration, low IDR protocol as compared with all groups (all P >= 0.05); no difference was found between the high concentration, low IDR protocol and the low concentration, high IDR protocol (P = 0.73 for comparison of beam-hardening artifacts; P = 0.50 for comparison of perfusion map image quality). Conclusion: High iodine concentration and high IDR contrast material delivery protocols provide the best image quality of both CTPA and perfusion map images of the lung through high attenuation in the pulmonary arteries and minimization of beam-hardening artifacts.
引用
收藏
页码:78 / 84
页数:7
相关论文
共 27 条
  • [1] Effect of contrast material injection duration and rate on aortic peak time and peak enhancement at dynamic CT involving injection protocol with dose tailored to patient weight
    Awai, K
    Hiraishi, K
    Hori, S
    [J]. RADIOLOGY, 2004, 230 (01) : 142 - 150
  • [2] Pulmonary embolism:: Prospective comparison of iso-osmolar and low-osmolarity nonionic contrast agents for contrast enhancement at CT angiography
    Bédard, JP
    Blais, C
    Patenaude, YG
    Monga, E
    [J]. RADIOLOGY, 2005, 234 (03) : 929 - 933
  • [3] Intra-individual comparison of different contrast media concentrations (300 mg, 370 mg and 400 mg iodine) in MDCT
    Behrendt, Florian F.
    Pietsch, Hubertus
    Jost, Gregor
    Sieber, Martin A.
    Keil, Sebastian
    Plumhans, Cedric
    Seidensticker, Peter
    Guenther, Rolf W.
    Mahnken, Andreas H.
    [J]. EUROPEAN RADIOLOGY, 2010, 20 (07) : 1644 - 1650
  • [4] Effect of different saline chaser volumes and flow rates on intravascular contrast enhancement in CT using a circulation phantom
    Behrendt, Florian F.
    Bruners, Philipp
    Keil, Sebastian
    Plumhans, Cedric
    Mahnken, Andreas H.
    Das, Marco
    Ackermann, Diana
    Guenther, Rolf W.
    Muehlenbruch, Georg
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2010, 73 (03) : 688 - 693
  • [5] Contrast Enhancement in Chest Multidetector Computed Tomography: Intraindividual Comparison of 300 mg/ml versus 400 mg/ml Iodinated Contrast Medium
    Behrendt, Florian F.
    Plumhans, Cedric
    Keil, Sebastian
    Muehlenbruch, Georg
    Das, Marco
    Seidensticker, Peter
    Mutscher, Christina
    Guenther, Rolf W.
    Mahnken, Andreas H.
    [J]. ACADEMIC RADIOLOGY, 2009, 16 (02) : 144 - 149
  • [6] Dual-Energy CT for Assessment of the Severity of Acute Pulmonary Embolism: Pulmonary Perfusion Defect Score Compared With CT Angiographic Obstruction Score and Right Ventricular/Left Ventricular Diameter Ratio
    Chae, Eun Jin
    Seo, Joon Beom
    Jang, Yu Mi
    Krauss, Bernhard
    Lee, Choong Wook
    Lee, Hyun Joo
    Song, Koun-Sik
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (03) : 604 - 610
  • [7] Dual-Energy CT Angiography of the Lung in Patients with Suspected Pulmonary Embolism: Initial Results
    Fink, C.
    Johnson, T. R.
    Michaely, H. J.
    Morhard, D.
    Becker, C.
    Reiser, M.
    Nikolaou, K.
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2008, 180 (10): : 879 - 883
  • [8] First performance evaluation of a dual-source CT (DSCT) system
    Flohr, TG
    McCollough, CH
    Bruder, H
    Petersilka, M
    Gruber, K
    Süss, C
    Grasruck, M
    Stierstorfer, K
    Krauss, B
    Raupach, R
    Primak, AN
    Küttner, A
    Achenbach, S
    Becker, C
    Kopp, A
    Ohnesorge, BM
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (02) : 256 - 268
  • [9] CT of deep venous thrombosis and pulmonary embolus: Does iso-osmolar contrast agent improve vascular opacification?
    Goodman, LR
    Gulsun, M
    Nagy, P
    Washington, L
    [J]. RADIOLOGY, 2005, 234 (03) : 923 - 928
  • [10] BEAM HARDENING ERRORS IN POST-PROCESSING DUAL-ENERGY QUANTITATIVE COMPUTED-TOMOGRAPHY
    GOODSITT, MM
    [J]. MEDICAL PHYSICS, 1995, 22 (07) : 1039 - 1047