Rapid steroid hormone quantification for congenital adrenal hyperplasia (CAH) in dried blood spots using UPLC liquid chromatography-tandem mass spectrometry

被引:49
作者
Janzen, Nils [1 ,2 ]
Sander, Stefanie [2 ]
Terhardt, Michael [2 ]
Steuerwald, Ulrike [2 ]
Peter, Michael [2 ]
Das, Anibh M. [1 ]
Sander, Johannes [2 ]
机构
[1] Hannover Med Sch, Clin Pediat Kidney Liver & Metab Dis, D-30625 Hannover, Germany
[2] Screening Lab Hannover, D-30430 Hannover, Germany
关键词
Congenital adrenal hyperplasia; UPLC-tandem mass spectrometry; Second-tier testing; Extended newborn screening; Steroid profile; HUMAN SERUM; LC-MS/MS; NEWBORN; 17-ALPHA-HYDROXYPROGESTERONE; VALIDATION; EXTRACTION; ANDROSTENEDIONE; TESTOSTERONE;
D O I
10.1016/j.steroids.2011.07.013
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Newborn screening for congenital adrenal hyperplasia (CAH) is usually done by quantifying 17 alpha-hydroxyprogesterone using immunoassay. However, this test produces high rates of false positive results caused by cross reacting steroids. Therefore we have developed a selective and specific method with a short run time (1.25 min) for quantification of 17 alpha-hydroxyprogesterone, 21-deoxycortisol, 11-deoxycortisol, 11-deoxycorticosterone and cortisol from dried blood spots. The extraction procedure is very simple and steroid separation is ensured on a BEH C18 column and an ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS). Analysis was done in positive ionization mode (ESI+) and recorded in multiple reaction monitoring mode (MRM). The method gave linear results for all steroids over a range of 5-200 (cortisol: 12.5-500) nmol/L with coefficients of regression >0.992. Absolute recovery was >64.1%. Across the analytical range the inter-assay coefficient of variation (CV) was <3%. Newborn blood samples of patients with confirmed 21-CAH and 11-CAH could clearly be distinguished from samples of unaffected newborns falsely positive on immunoassay. The method is not influenced by cross reactions as found on immunoassay. Analysis of dried blood spots shows that this method is sensitive and fast enough to allow rapid analysis and can therefore improve the newborn screening program. (C) 2011 Elsevier Inc. All rights reserved.
引用
收藏
页码:1437 / 1442
页数:6
相关论文
共 32 条
[1]   Ion suppression in mass spectrometry [J].
Annesley, TM .
CLINICAL CHEMISTRY, 2003, 49 (07) :1041-1044
[2]   Quantitative determination of guanidinoacetate and creatine in dried blood spot by flow injection analysis-electrospray tandem mass spectrometry [J].
Carducci, C ;
Santagata, S ;
Leuzzi, V ;
Carducci, C ;
Artiola, C ;
Giovanniello, T ;
Batti, R ;
Antonozzi, I .
CLINICA CHIMICA ACTA, 2006, 364 (1-2) :180-187
[3]   Simultaneous quantitation of seven endogenous C-21 adrenal steroids by liquid chromatography tandem mass spectrometry in human serum [J].
Carvalho, Valdemir Melechco ;
Nakamura, Odete Hirata ;
Henriques Vieira, Jose Gilberto .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 872 (1-2) :154-161
[4]   Rapid quantification of steroid patterns in human serum by on-line solid phase extraction combined with liquid chromatography-triple quadrupole linear ion trap mass spectrometry [J].
Ceglarek, Uta ;
Kortz, Linda ;
Leichtle, Alexander ;
Fiedler, Georg M. ;
Kratzsch, Juergen ;
Thiery, Joachim .
CLINICA CHIMICA ACTA, 2009, 401 (1-2) :114-118
[5]  
Chace DH, 1999, CLIN CHEM, V45, P1269
[6]   Mass spectrometry in newborn and metabolic screening: historical perspective and future directions [J].
Chace, Donald H. .
JOURNAL OF MASS SPECTROMETRY, 2009, 44 (02) :163-170
[7]   Pilot proficiency testing study for second tier congenital adrenal hyperplasia newborn screening [J].
De Jesus, Victor R. ;
Simms, David A. ;
Schiffer, Jarad ;
Kennedy, Meredith ;
Mei, Joanne V. ;
Hannon, W. Harry .
CLINICA CHIMICA ACTA, 2010, 411 (21-22) :1684-1687
[8]   Liquid chromatography-tandem mass spectrometry determination of oxalate in spot urine [J].
Elgstoen, Katja B. P. ;
Woldseth, Berit ;
Hoie, Kari ;
Morkrid, Lars .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2010, 70 (03) :145-150
[9]   False positive rate in newborn screening for congenital adrenal hyperplasia (CAH)-ether extraction reveals two distinct reasons for elevated 17α-hydroxyprogesterone (17-OHP) values [J].
Fingerhut, Ralph .
STEROIDS, 2009, 74 (08) :662-665
[10]   Development and validation of a sensitive liquid chromatography-tandem mass spectrometry assay to simultaneously measure androgens and estrogens in serum without derivatization [J].
Harwood, D. Tim ;
Handelsman, David J. .
CLINICA CHIMICA ACTA, 2009, 409 (1-2) :78-84