Calculation of flip angles for echo trains with predefined amplitudes with the extended phase graph (EPG)-algorithm: Principles and applications to hyperecho and TRAPS sequences

被引:103
作者
Hennig, J
Weigel, M
Scheffler, K
机构
[1] Univ Hosp, Dept Diagnost Radiol, Sect Med Phys, D-79106 Freiburg, Germany
[2] Univ Basel Hosp, MR Phys, CH-4031 Basel, Switzerland
关键词
extended phase graph; hyperecho; TRAPS; TSE;
D O I
10.1002/mrm.10658
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The article presents an algorithm for calculation of flip angles in multiecho experiments to generate echoes with predefined amplitudes based on the extended phase graph algorithm. The algorithm can be used to optimize the echo envelope and thus the point spread function (PSF) in hyperecho and TRAPS (transition into the pseudosteady state) experiments while minimizing the total RF power. Implementations at 3 T using echo trains with Gaussian and Lorentzian PSF demonstrate a reduction in RF power by a factor of 3-5 while maintaining high image quality. (C) 2003 Wiley-Liss, Inc.