Comparison of the VersaTrek and BACTEC MGIT 960 systems for the contamination rate, time of detection and recovery of mycobacteria from clinical specimens

被引:0
作者
Yuksel, Pelin [1 ]
Saribas, Suat [1 ]
Bagdatli, Yasar [1 ]
机构
[1] Istanbul Univ Cerrahpasa, Fac Med, Dept Microbiol & Clin Microbiol, Istanbul, Turkey
关键词
VersaTrek; MGIT; 960; M; tuberculosis; GROWTH INDICATOR TUBE; LOWENSTEIN-JENSEN MEDIUM; TUBERCULOSIS; CULTURE;
D O I
暂无
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
The most commonly used conventional Lowenstein Jensen (L.J) culture method requires at least 6 to 10 weeks of incubation due to the slow growth rate of the Mycobacterium tuberculosis complex. Today, there are liquid automated systems to detect mycobacteria than faster conventional LJ culture method. In this study, we aimed to compare of the VersaTrek system and BACTEC MGIT 960 system for the contamination rate and recovery of mycobacteria from clinical specimens. The recovery rates of mycobacteria were detected as 4% (2/50) for both of the VersaTrek and MGIT 960 systems. No statistically significant difference was detected for the recovery rates of mycobacteria in both of the systems (p>0.05). Two strains were detected on 12 and 15 day in MGIT 960 system. Two strains were detected on 15 and 20 day in VersaTrek system. No statistically significant difference was detected for the time of detection of mycobacteria in both of the systems (p>0.05). The contamination rates were found as 10% (5/50) and 6% (3/50) in MGIT 960 and VersaTrek, systems respectively (p>0.05). No statistically significant difference was detected for the contamination rates for both of the systems. Antituberculosis resistance tests were performed using with antituberculosis agents in MGIT 960 and VersaTrek systems. The two strains were found sensitive against to streptomycin, isoniazid, rifampicin and ethambutol in MGIT 960 system. The two strains were found sensitive against to isoniazid, rifampicin and ethambutol in VersaTrek system. Streptomycin was not used in VersaTrek system because there is no standardization for streptomycin. We also examined the time of detection of 5 confirmed M. tuberculosis complex members in MGIT 960 and VersaTrek systems. When the time of detection in MGIT 960 and VersaTrek systems were compared, no statistically significant difference was found (p>0.05). In conclusion, no substantial difference has been observed between MGIT 960 and VersaTrek systems for M. tuberculosis isolation and contamination rate. We believe that VersaTrek media, which is going to be put into use in Turkey, is easy for application and follow up, gives rapid and sensitive results and that it can be used in routine alike MGIT media.
引用
收藏
页码:985 / 989
页数:5
相关论文
共 18 条
[1]  
Atas E, 2003, TUR TORKAS DER, V4, P138
[2]  
Augustynowicz-Kopec Ewa, 2002, Pneumonol Alergol Pol, V70, P450
[3]  
Casal M, 1997, INT J TUBERC LUNG D, V1, P81
[4]   Mycobacterial growth and bacterial contamination in the mycobacteria growth indicator tube and BACTEC 460 culture systems [J].
Cornfield, DB ;
Beavis, KG ;
Greene, JA ;
Bojak, M ;
Bondi, J .
JOURNAL OF CLINICAL MICROBIOLOGY, 1997, 35 (08) :2068-2071
[5]  
Diraa O, 2003, INT J TUBERC LUNG D, V7, P1010
[6]   Comparison of the VersaTREK system and Lowenstein-Jensen medium for the recovery of mycobacteria from clinical specimens [J].
Falconi, Francisco Quinones ;
Suarez, Lourdes Infante ;
Castillejos Lopez, Manuel de Jesus ;
Sancho, Cecilia Garcia .
SCANDINAVIAN JOURNAL OF INFECTIOUS DISEASES, 2008, 40 (01) :49-53
[7]  
Ghatole M, 2005, Indian J Med Microbiol, V23, P111
[8]  
Giampaglia CMS, 2007, INT J TUBERC LUNG D, V11, P803
[9]   Culture and susceptibility testing of mycobacteria with VersaTREK [J].
Gravet, A. ;
Souillard, N. ;
Habermacher, J. ;
Moser, A. ;
Lohmann, C. ;
Schmitt, F. ;
Delarbre, J. -M. .
PATHOLOGIE BIOLOGIE, 2011, 59 (01) :32-38
[10]  
ILGAZLI A, 1998, AKCIGER HASTALIKLARI, V9, P627