Implementation of a beam deflection system for studies of liquid interfaces on beamline I07 at Diamond

被引:39
作者
Arnold, Thomas [1 ]
Nicklin, Chris [1 ]
Rawle, Jonathan [1 ]
Sutter, John [1 ]
Bates, Trevor [1 ]
Nutter, Brian [1 ]
McIntyre, Gary [1 ]
Burt, Martin [1 ]
机构
[1] Diamond Light Source, Didcot OX11 0DE, Oxon, England
关键词
X-ray reflectivity; GIXD; GISAXS; air-water; liquid-liquid;
D O I
10.1107/S0909049512009272
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
X-ray optics, based on a double-crystal deflection scheme, that enable reflectivity measurements from liquid surfaces/interfaces have been designed, built and commissioned on beamline I07 at Diamond Light Source. This system is able to deflect the beam onto a fixed sample position located at the centre of a five-circle diffractometer. Thus the incident angle can be easily varied without moving the sample, and the reflected beam is tracked either by a moving Pilatus 100K detector mounted on the diffractometer arm or by a stationary Pilatus 2M detector positioned appropriately for small-angle scattering. Thus the system can easily combine measurements of the reflectivity from liquid interfaces (Qz > 1 angstrom-1) with off-specular data collection, both in the form of grazing-incidence small-angle X-ray scattering (GISAXS) or wider-angle grazing-incidence X-ray diffraction (GIXD). The device allows operation over the energy range 1028 keV.
引用
收藏
页码:408 / 416
页数:9
相关论文
共 12 条
[1]  
Daillant J, 2000, CURR SCI INDIA, V78, P1496
[2]   Theory of the use of more than two successive x-ray crystal reflections to obtain increased resolving power [J].
DuMond, JWM .
PHYSICAL REVIEW, 1937, 52 (08) :0872-0883
[3]   X-ray optics for liquid surface/interface spectrometers [J].
Honkimaeki, V. ;
Reichert, H. ;
Okasinski, J. S. ;
Dosch, H. .
JOURNAL OF SYNCHROTRON RADIATION, 2006, 13 :426-431
[4]   SHADOW - A SYNCHROTRON RADIATION RAY TRACING PROGRAM [J].
LAI, B ;
CERRINA, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 246 (1-3) :337-341
[5]   Following transient phases at the air/water interface [J].
Meron, M. ;
Gebhardt, J. ;
Brewer, H. ;
Viccaro, J. P. ;
Lin, B. .
EUROPEAN PHYSICAL JOURNAL-SPECIAL TOPICS, 2009, 167 :137-142
[6]   A new Diffractometer for Studies of Liquid - Liquid Interfaces [J].
Murphy, B. M. ;
Greve, M. ;
Runge, B. ;
Koops, C. T. ;
Elsen, A. ;
Stettner, J. ;
Seeck, O. H. ;
Magnussen, O. M. .
SRI 2009: THE 10TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION, 2010, 1234 :155-+
[8]  
Salah F, 2007, J APPL CRYSTALLOGR, V40, P813, DOI 10.1107/S0021889807030403813
[9]   A synchrotron x-ray liquid surface spectrometer [J].
Schlossman, ML ;
Synal, D ;
Guan, YM ;
Meron, M ;
Shea-McCarthy, G ;
Huang, ZQ ;
Acero, A ;
Williams, SM ;
Rice, SA ;
Viccaro, PJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (12) :4372-4384