Pressure and traction on a model fetal head and neck associated with the use of forceps, Kiwi™ ventouse and the BD Odon Device™ in operative vaginal birth: a simulation study

被引:18
作者
O'Brien, S. M. [1 ,2 ]
Winter, C. [1 ]
Burden, C. A. [1 ,2 ]
Boulvain, M. [3 ]
Draycott, T. J. [1 ,4 ]
Crofts, J. F. [1 ]
机构
[1] Southmead Hosp, Dept Obstet & Gynaecol, Bristol, Avon, England
[2] Univ Bristol, Sch Clin Sci, Bristol, Avon, England
[3] Hop Univ Geneve, Dept Obstet & Gynaecol, Geneva, Switzerland
[4] Univ Bristol, Sch Social & Community Med, Bristol, Avon, England
基金
比尔及梅琳达.盖茨基金会;
关键词
Forceps; Kiwi; Odon; pressure; traction; GESTATION; DELIVERY;
D O I
10.1111/1471-0528.14760
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective To determine the pressure and traction forces exerted on a model fetal head by the BD Odon Device, forceps and Kiwi ventouse during simulated births. Design Simulation study. Setting Simulated operative vaginal birth. Population or sample Eighty-four simulated operative vaginal births. Methods A bespoke fetal mannequin with pressure sensors around the head and strain gauge across the neck was used to investigate pressure applied over the head, and traction across the neck during 84 simulated births using the BD Odon Device, non-rotational forceps and Kiwi ventouse. Main outcome measures Peak pressure on the fetal face and lateral aspects of the head during correct use of the BD Odon Device and forceps. Peak pressure on orbits and neck during misplacement of the BD Odon Device and forceps. Peak traction force generated until instrument failure using the BD Odon Device, forceps and Kiwi ventouse. Results When correctly sited and using 80kPa inflation pressure on the cuff, the BD Odon Device generated a lower peak pressure on the fetal head than forceps (83 versus 146kPa). When instruments were purposefully misplaced over the orbits, the BD Odon Device generated a lower peak pressure on the orbits compared with forceps (70 versus 123kPa). When purposefully misplaced over the neck, the BD Odon Device, compared with forceps, generated a greater peak pressure on the anterio-lateral aspect of the neck (56 versus 17kPa) and a lower peak pressure on the posterior aspect of the neck (76 versus 93kPa) than forceps. In cases of true cephalic disproportion, the BD Odon Device popped-off' at a lower traction force than did forceps (208 versus 270N). Conclusions In simulated assisted vaginal birth with correctly placed instruments, the peak pressure exerted on the fetal head by a BD Odon Device is lower than the pressure exerted by non-rotational forceps. In cases in which delivery of the fetal head is not possible due to cephalo-pelvic disproportion, lower traction forces could be applied using the BD Odon Device than with forceps before the procedure was abandoned due to device failure.
引用
收藏
页码:19 / 25
页数:7
相关论文
共 13 条
[1]  
Crofts J, 2007, AM J OBSTET GYNECOL, V197, P156, DOI [10.3109/00016346009157836, DOI 10.3109/00016346009157836]
[2]  
Crofts JF, 2015, BJOG, V123, P111
[3]   Improving neonatal outcome through practical shoulder dystocia training [J].
Draycott, Timothy J. ;
Crofts, Joanna F. ;
Ash, Jonathan P. ;
Wilson, Louise V. ;
Yard, Elaine ;
Sibanda, Thabani ;
Whitelaw, Andrew .
OBSTETRICS AND GYNECOLOGY, 2008, 112 (01) :14-20
[4]   Maternal age and other predictors of newborn blood pressure [J].
Gillman, MW ;
Rich-Edwards, JW ;
Rifas-Shiman, SL ;
Lieberman, ES ;
Kleinman, KP ;
Lipshultz, SE .
JOURNAL OF PEDIATRICS, 2004, 144 (02) :240-245
[5]   Third- and fourth-degree perineal tears among primiparous women in England between 2000 and 2012: time trends and risk factors [J].
Gurol-Urganci, I. ;
Cromwell, D. A. ;
Edozien, L. C. ;
Mahmood, T. A. ;
Adams, E. J. ;
Richmond, D. H. ;
Templeton, A. ;
van der Meulen, J. H. .
BJOG-AN INTERNATIONAL JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 2013, 120 (12) :1516-1525
[6]   A new obstetrical polyurethane versus stainless steel forceps: a comparison of forces generated to the base of the fetal skull during simulated deliveries [J].
Hale, Theodore M. ;
Shi, Zhengzhuan ;
Idangodage, Hasitha ;
Mercado, Ray ;
Skupski, Daniel ;
Veilastegui, Zoila .
JOURNAL OF PERINATAL MEDICINE, 2009, 37 (06) :669-671
[7]  
Honig MA, 1996, CORNEA, V15, P463
[8]   Choice of instruments for assisted vaginal delivery [J].
O'Mahony, Fidelma ;
Hofmeyr, G. Justus ;
Menon, Vijay .
COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2010, (11)
[9]   Risk factors and morbidity associated with suboptimal instrument placement at instrumental delivery: observational study nested within the Instrumental Delivery & Ultrasound randomised controlled trial ISRCTN 72230496 [J].
Ramphul, M. ;
Kennelly, M. M. ;
Burke, G. ;
Murphy, D. J. .
BJOG-AN INTERNATIONAL JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 2015, 122 (04) :558-563
[10]  
RCOG, 2011, OP VAG DEL, P1