Corrosion resistance of stressed NiTi and stainless steel orthodontic wires in acid artificial saliva

被引:56
作者
Huang, HH [1 ]
机构
[1] Chung Shan Med Univ, Inst Dent Mat, Taichung 402, Taiwan
关键词
corrosion; NiTi; stainless steel; load; pH;
D O I
10.1002/jbm.a.10463
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to investigate the corrosion resistance of stressed NiTi and stainless steel orthodontic wires using cyclic potentiodynamic and potentiostatic tests in acid artificial saliva at 37degreesC. An atomic force microscope was used to measure the 3-D surface topography of as-received wires. Scanning electron microscope observations were carried out before and after the cyclic potentiodynamic tests. The surface chemical analysis was characterized using X-ray photoelectron spectroscopy and Auger electron spectroscopy after the potentiostatic tests. The cyclic potentiodynamic test results showed that the pH had a significant influence on the corrosion parameters of the stressed NiTi and stainless steel wires (p < 0.05). The pitting potential, protection potential, and passive range of stressed NiTi and stainless steel wires decreased on decreasing pH, whereas the passive current density increased on decreasing pH. The load had no significant influence on the above corrosion parameters (p > 0.05). For all pH and load conditions, stainless steel wire showed higher pitting potential and wider passive range than NiTi wire (p < 0.001), whereas NiTi wire had lower passive current density than stainless steel wire (p < 0.001). The corrosion resistance of the stressed NiTi and stainless steel wires was related to the surface characterizations, including surface defect and passive film. (C) 2003 Wiley Periodicals, Inc.
引用
收藏
页码:829 / 839
页数:11
相关论文
共 36 条
[1]   NICKEL HYPERSENSITIVITY IN THE ORTHODONTIC PATIENT [J].
BASS, JK ;
FINE, H ;
CISNEROS, GJ .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1993, 103 (03) :280-285
[2]  
BergerGorbet M, 1996, J BIOMED MATER RES, V32, P243, DOI 10.1002/(SICI)1097-4636(199610)32:2<243::AID-JBM14>3.0.CO
[3]  
2-K
[4]  
GEISGERSTORFER J, 1985, DEUT ZAHNAERZTL Z, V40, P87
[5]   REACTION TO STAINLESS-STEEL WIRE FOLLOWING ORTHOGNATHIC SURGERY [J].
GUYURON, B ;
LASA, CI .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1992, 89 (03) :540-542
[6]   CORROSION-RESISTANT ALLOYS IN CHLORIDE SOLUTIONS - MATERIALS FOR SURGICAL IMPLANTS [J].
HOAR, TP ;
MEARS, DC .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1966, 294 (1439) :486-&
[7]   Effect of chemical composition on the corrosion behavior of Ni-Cr-Mo dental casting alloys [J].
Huang, HH .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2002, 60 (03) :458-465
[8]  
Hunt NP, 1999, ANGLE ORTHOD, V69, P433
[9]  
Kim H, 1999, ANGLE ORTHOD, V69, P39
[10]   Nickel and molybdenum contact allergies in patients with coronary in-stent restenosis [J].
Köster, R ;
Vieluf, D ;
Kiehn, M ;
Sommerauer, M ;
Kähler, J ;
Baldus, S ;
Meinertz, T ;
Hamm, CW .
LANCET, 2000, 356 (9245) :1895-1897