Vanadium ionic species from degradation of Ti-6Al-4V metallic implants: In vitro cytotoxicity and speciation evaluation

被引:172
作者
Costa, Bruna C. [1 ,2 ]
Tokuhara, Cintia K. [2 ,3 ]
Rocha, Luis A. [2 ,4 ]
Oliveira, Rodrigo C. [2 ,3 ]
Lisboa-Filho, Paulo N. [2 ,4 ]
Pessoa, Joao Costa [5 ]
机构
[1] Sao Univ Estadual Paulista UNESP, Programa Posgrad Ciencia & Tecnol Mat POSMAT, Av Engn Luiz Edmundo Carrijo Coube 14-01, BR-17033360 Bauru, SP, Brazil
[2] Inst Biomat Tribocorrosao & Nanomed IBTN Br, Av Engn Luiz Edmundo Carrijo Coube 14-01, BR-17033360 Bauru, SP, Brazil
[3] Univ Sao Paulo, Fac Odontol Bauru, Alameda Dr Otavio Pinheiro Brisolla 9-75, BR-17012901 Sao Paulo, Brazil
[4] Dept Fis, Av Engn Luiz Edmundo Carrijo Coube 14-01, BR-17033360 Bauru, SP, Brazil
[5] Univ Lisbon, Inst Super Tecn, Ctr Quim Estrutural, Av Rovisco Pais, P-1049001 Lisbon, Portugal
来源
MATERIALS SCIENCE AND ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS | 2019年 / 96卷
基金
巴西圣保罗研究基金会;
关键词
Ti-6Al-4V; Tribocorrosion; Vanadate(V); Oxidovanadium(IV); Ionic speciation; Cytotoxicity; COMPLEXES; TITANIUM; ALLOY; VANADATE; ACTIVATION; BEHAVIOR; LIGANDS; BINDING; OXOVANADIUM(IV); MICROSTRUCTURES;
D O I
10.1016/j.msec.2018.11.090
中图分类号
TB3 [工程材料学]; R318.08 [生物材料学];
学科分类号
0805 ; 080501 ; 080502 ;
摘要
Among the metallic materials used in biomedical industry, the most common choice for orthopedics and dental implants is titanium (Ti) and its alloys, mainly due to their superior corrosion and tribocorrosion resistance and biocompatibility. Under different conditions in vivo, such as different pH levels, composition of body fluid and mechanical loads, metallic materials may suffer from degradation, resulting in the release of undesired wear particles and ions. In particular, the Ti-6Al-4V system represents almost half of the production of Ti as a biomaterial and many concerns have been raised about titanium, aluminum and vanadium ions releasing. This work evaluates the cytotoxic effects of vanadium ionic species generated from Ti-6Al-4V surfaces regarding mouse pre-osteoblasts and fibroblasts. In our cell viability tests, we noticed a significant decrease in the fibroblasts' cell viability with vanadium concentrations (23 mu M) close to those previously reported to be observed in vivo in patients with poor functioning of their medical devices based on Ti-6Al-4V (30 mu M). Speciation modelling was carried-out, for the first time, to this system. Results of the modelling reveal that vanadates(V), namely H2VO4- and HVO42-, are the main species present in cell culture media. Otherwise, in synovial fluids of individuals with poorly functioning implants, wherein the concentration of vanadium may go up to ca. 30 mu M, the tentative theoretical speciation data indicates a high occurrence probability for V-V- and V-IV-species bound to albumin and hyaluronic acid. In conclusion, even though relatively low concentrations of vanadium may be released from Ti-6Al-4V implants in vivo, the continuous contact with peri-implant cells for long periods of time may represent a potentially hazardous situation.
引用
收藏
页码:730 / 739
页数:10
相关论文
共 84 条
[1]   Hyperquad simulation and speciation (HySS): a utility program for the investigation of equilibria involving soluble and partially soluble species [J].
Alderighi, L ;
Gans, P ;
Ienco, A ;
Peters, D ;
Sabatini, A ;
Vacca, A .
COORDINATION CHEMISTRY REVIEWS, 1999, 184 :311-318
[2]   Tribocorrosion behaviour of anodic treated titanium surfaces intended for dental implants [J].
Alves, A. C. ;
Oliveira, F. ;
Wenger, F. ;
Ponthiaux, P. ;
Celis, J-P ;
Rocha, L. A. .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2013, 46 (40)
[3]   Metal in Total Hip Arthroplasty: Wear Particles, Biology, and Diagnosis [J].
Amanatullah, Derek F. ;
Sucher, Mark G. ;
Bonadurer, George F., III ;
Pereira, Gavin C. ;
Taunton, Michael J. .
ORTHOPEDICS, 2016, 39 (06) :371-379
[4]   HYALURONIC ACID IN SYNOVIAL FLUID .I. MOLECULAR PARAMETERS OF HYALURONIC ACID IN NORMAL AND ARTHRITIC HUMAN FLUIDS [J].
BALAZS, EA ;
WATSON, D ;
DUFF, IF ;
ROSEMAN, S .
ARTHRITIS AND RHEUMATISM, 1967, 10 (04) :357-&
[5]  
Ballazs E. A., 1974, PHYS PROPERTIES SYNO, P63
[6]   Vanadyl(IV) complexes with saccharides.: Bioactivity in osteoblast-like cells in culture [J].
Barrio, Daniel A. ;
Cattaneo, Elizabeth R. ;
Apezteguia, Maria C. ;
Etcheverry, Susana B. .
CANADIAN JOURNAL OF PHYSIOLOGY AND PHARMACOLOGY, 2006, 84 (07) :765-775
[7]  
Boas L. Vilas, 1987, VANADIUM COMPREHENSI, P453
[8]  
BRANCA M, 1989, DALTON T, P1283
[9]   Tribo-corrosion of materials: Interplay between chemical, electrochemical, and mechanical reactivity of surfaces [J].
Celis, J. -P. ;
Ponthiaux, P. ;
Wenger, F. .
WEAR, 2006, 261 (09) :939-946
[10]  
Chen Q., 2014, MAT CORROS