Warning!: Suctioning.: A lung model evaluation of closed suctioning systems

被引:39
作者
Stenqvist, O [1 ]
Lindgren, S [1 ]
Kárason, S [1 ]
Söndergaard, S [1 ]
Lundin, S [1 ]
机构
[1] Sahlgrens Univ Hosp, Dept Anaesthesia & Intens Care, S-41345 Gothenburg, Sweden
关键词
suctioning; closed system suctioning; airway pressure; intrinsic PEEP; ARDS; ALI; open lung;
D O I
10.1034/j.1399-6576.2001.450206.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Closed system suctioning, CSS, has been advocated to avoid alveolar collapse. However, ventilator manufacturers indicate that extreme negative pressure levels can be obtained during closed system suctioning, impeding the performance of the ventilator. Methods: Suctioning with a 12 or 14 Fr catheter with a vacuum level of -50 kPa was either performed with an open technology or a CSS, where the catheter is introduced through a tight-fitting connection through the endotracheal tube, ETT. The lung model was ventilated with a Servo 900C or 300 ventilator with an I:E ratio of 1:2, 1:1 and 2:1 and extrinsic positive end-expiratory pressure (PEEP) at 0 or 10 cm H2O. Respiratory volumes and alveolar pressure were measured at the lung model alveolus. Results: The initial suctioning flow was >40 l/min with a. 14 Pr catheter. When inserting the catheter through a no. 7 ETT, PEEP rose from 11 to 23 cm H2O during volume control ventilation with an I:E ratio 1:1. During suctioning the alveolar pressure fell to 10 cm H2O below the set PEEP level. CSS during pressure control ventilation had fewer effects. Low tidal volumes, inverse I:E ratio and secretions in the tube resulted in pressures down to -92 cm H2O. Conclusion: CSS should not be used in volume control ventilation due to risk of high intrinsic PEEP levels at insertion of the catheter and extreme negative pressures during suctioning. Pressure control ventilation produces less intrinsic PEEP effect. The continuous positive airway pressure (CPAP) mode offers the least intrinsic PEEP during insertion of the catheter and least sub-atmospheric pressure during suctioning.
引用
收藏
页码:167 / 172
页数:6
相关论文
共 16 条
[1]  
Adams D H, 1997, Intensive Crit Care Nurs, V13, P72, DOI 10.1016/S0964-3397(97)80135-8
[2]   Effect of a protective-ventilation strategy on mortality in the acute respiratory distress syndrome [J].
Amato, MBP ;
Barbas, CSV ;
Medeiros, DM ;
Magaldi, RB ;
Schettino, GDP ;
Lorenzi, G ;
Kairalla, RA ;
Deheinzelin, D ;
Munoz, C ;
Oliveira, R ;
Takagaki, TY ;
Carvalho, CRR .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 338 (06) :347-354
[3]   Closed tracheal suctioning systems and infection control in the intensive care unit [J].
Blackwood, B ;
Webb, CH .
JOURNAL OF HOSPITAL INFECTION, 1998, 39 (04) :315-321
[4]   CONSTANT-FLOW INSUFFLATION PREVENTS ARTERIAL OXYGEN DESATURATION DURING ENDOTRACHEAL SUCTIONING [J].
BROCHARD, L ;
MION, G ;
ISABEY, D ;
BERTRAND, C ;
MESSADI, AA ;
MANCEBO, J ;
BOUSSIGNAC, G ;
VASILE, N ;
LEMAIRE, F ;
HARF, A .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1991, 144 (02) :395-400
[5]   EVALUATION OF A CLOSED-TRACHEAL SUCTION SYSTEM [J].
CARLON, GC ;
FOX, SJ ;
ACKERMAN, NJ .
CRITICAL CARE MEDICINE, 1987, 15 (05) :522-525
[6]  
CLARK AP, 1990, HEART LUNG, V19, P552
[7]  
CRAIG KC, 1984, RESPIR CARE, V29, P1013
[8]  
Glass C, 1999, Am J Crit Care, V8, P93
[9]  
Harshbarger S A, 1992, Am J Crit Care, V1, P57
[10]   CLOSED VERSUS OPEN ENDOTRACHEAL SUCTIONING - COSTS AND PHYSIOLOGICAL CONSEQUENCES [J].
JOHNSON, KL ;
KEARNEY, PA ;
JOHNSON, SB ;
NIBLETT, JB ;
MACMILLAN, NL ;
MCCLAIN, RE .
CRITICAL CARE MEDICINE, 1994, 22 (04) :658-666