Effects of neck movements on stability and subsidence in cervical interbody fusion: an in vitro study

被引:51
作者
Kettler, A [1 ]
Wilke, HJ [1 ]
Claes, L [1 ]
机构
[1] Univ Ulm, Inst Orthoped Res & Biomech, D-89081 Ulm, Germany
关键词
cervical spine; interbody fusion; cyclic testing; flexibility; biomechanics;
D O I
10.3171/spi.2001.94.1.0097
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. The aim of this in vitro study was to determine the influence of simulated postoperative neck movements on the stabilizing effect and subsidence of four different anterior cervical interbody fusion devices. Emphasis was placed on the relation between subsidence and spinal stability. Methods. The flexibility of 24 human cervical spine specimens was tested before and directly after being stabilized with a WING, BAK/C, AcroMed I/F cage, or with bone cement in standard flexibility tests under 50 N axial preload. Thereafter, 700 pure moment loading cycles (+/- 2 Nm) were applied in randomized directions to simulate physiological neck movements. Additional flexibility tests in combination with measurements of the subsidence depth were conducted after 50, 100, 200, 300, 500, and 700 loading cycles. In all four groups, simulated postoperative neck movements caused an increase of the range of motion (ROM) ranging from 0.4 to 3.1 degrees and of the neutral zone from 0.1 to 4.2 degrees. This increase in flexibility was most distinct in extension followed by flexion, lateral bending, and axial rotation. After cyclic loading, ROM tended to be lower in the group fitted with AcroMed cages (3.3 degrees in right lateral bending, 3.5 degrees in left axial rotation, 7.8 degrees in flexion, 8.3 degrees in extension) and in the group in which bone cement was applied (5.4 degrees, 2.5 degrees, 7.4 degrees, and 8.8 degrees, respectively) than in those fu;ed with the WING (6.3 degrees, 5.4 degrees, 9.7 degrees, and 6.9 degrees, respectively) and BAK cages (6.2 degrees, 4.5 degrees, 10.2 degrees, and 11.6 degrees, respectively). Conclusions. Simulated repeated neck movements not only caused an increase of the flexibility but also subsidence of the implants into the adjacent vertebrae. The relation between flexibility increase and subsidence seemed to depend on the implant design: subsiding BAK/C cages partially supported stability whereas subsiding WING cages and AcroMed cages did not.
引用
收藏
页码:97 / 107
页数:11
相关论文
共 26 条
[1]   Biocompatible osteoconductive polymer versus iliac graft - A prospective comparative study for the evaluation of fusion pattern after anterior cervical discectomy [J].
AbouMadawi, A ;
Powell, M ;
Crockard, HA .
SPINE, 1996, 21 (18) :2123-2129
[2]  
BAGBY GW, 1988, ORTHOPEDICS, V11, P931
[3]  
Brooke NSR, 1997, BRIT J NEUROSURG, V11, P221
[4]  
BROWN MD, 1976, CLIN ORTHOP RELAT R, V119, P231
[5]  
FATHIE K, 1994, MT SINAI J MED, V61, P246
[6]   Bisegmental cervical interbody fusion using hydroxyapatite implants: surgical results and long-term observation in 70 cases [J].
Kim, P ;
Wakai, S ;
Matsuo, S ;
Moriyama, T ;
Kirino, T .
JOURNAL OF NEUROSURGERY, 1998, 88 (01) :21-27
[7]   DISCECTOMIES OF THE LOWER CERVICAL-SPINE USING INTERBODY BIO-POLYMER (BOP) IMPLANTS - ADVANTAGES IN THE TREATMENT OF COMPLICATED CERVICAL ARTHROSIS - A REVIEW OF 150 CASES [J].
LOZES, G ;
FAWAZ, A ;
CAMA, A ;
KRIVOSIC, I ;
DEVOS, P ;
HERLANT, M ;
SERTL, GO ;
CLARISSE, J ;
JOMIN, M .
ACTA NEUROCHIRURGICA, 1989, 96 (3-4) :88-93
[8]   Anterior interbody fusion with the BAK-cage in cervical spondylosis [J].
Matge, G .
ACTA NEUROCHIRURGICA, 1998, 140 (01) :1-8
[9]  
Moeini SMR, 1998, J SPINAL DISORD, V11, P129
[10]   BIOMECHANICAL TIME-TOLERANCE OF FRESH CADAVERIC HUMAN SPINE SPECIMENS [J].
PANJABI, MM ;
KRAG, M ;
SUMMERS, D ;
VIDEMAN, T .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1985, 3 (03) :292-300