Comparative analysis for five fixations of Pauwels-I by the biomechanical finite-element method

被引:2
|
作者
Jian-Qiao Peng, Matthew [1 ]
Chen, Hai-Yan [2 ]
Ju, XiangYang [3 ]
Hu, Yong [4 ]
Ayoub, Ashraf [3 ]
Khambay, Balvinder [3 ]
Liu, ZiQing [5 ]
Bai, Bo [1 ]
机构
[1] GuangZhou Med Univ, Affiliated Hosp 1, Orthoped Dept, 151 YanJiangXi Rd, Guangzhou 510120, Guangdong, Peoples R China
[2] HuiDong Peoples Hosp, Orthoped Dept, Huizhou, Guangdong, Peoples R China
[3] Univ Glasgow, Dept Clin Phys & Bioengn, Glasgow, Lanark, Scotland
[4] Univ Hong Kong, Neural Electrophysiol Lab, Hong Kong, Peoples R China
[5] SanShui Peoples Hosp, Traumat Orthoped Dept, 16 GuangHai DaDaoXI, Foshan 528100, Guangdong, Peoples R China
关键词
Pauwels-I; stress; displacement; fixation; FEA; INTERNAL-FIXATION; HIP-PROSTHESIS; FATIGUE BEHAVIOR; ARTHROPLASTY; OPTIMIZATION; FRACTURES; IMPLANTS;
D O I
10.1080/08941939.2018.1533054
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: Little is known about how biomechanics govern the five fixtures such as DHS, MLS, DHS + LS, LP, and HA are accepted as common therapeutic techniques. Aims and objectives: A series of numerical models for a femoral neck fracture of Pauwels-I will be constructed by innovative approach of finite element in order to determine the most optimized option in comparison with biomechanical performance. Method: Twenty sets of computer tomography scanned femora were imported onto Mimics to extract 3 D models; these specimens were transferred to Geomagic-Studio for a simulative osteotomy and kyrtograph; then, they underwent UG to fit simulative solid models; 5 sorts of fixture were then expressed by Pro-Engineer virtually. After processing with HyperMesh, all compartments (fracture model + internal implant) were assembled onto 5 systems: "Dynamic Hip Screw (DHS), Multiple Lag screw (MLS), DHS + LS, femoral Locking Plate (LP) and HemiArthroplasty (HA)." Eventually, numerical models of the finite-elemental analysis were exported to AnSys to determine the solution. Result: Four models of fixation and a simulation of HA for Pauwels-I were established, validated, and analyzed with the following findings: In term of displacement, these 5 fixtures ranged between 0.3801 and 0.7536 mm have no significant difference; in term of stress, the averages of peaks for integral assemblage are b(MLS) = 43.5766 approximate to< d(LP) = 43.6657 approximate to< e(Ha) = 43.6657 < c(DHS + LS) = 66.5494 < a(DHS) = 105.617 in MPa indicate that MLS, LP and HA are not significantly different, but less than DHS + LS or DHS in each. Conclusion: A fixture of MLS or LP with optional HA should be recommended to clinically optimize a Pauwels-I facture.
引用
收藏
页码:428 / 437
页数:10
相关论文
共 50 条
  • [1] Comparative Analysis for Internal Fixations of Pauwels II by Biomechanical Finite Element Method
    Peng, Matthew Jian-Qiao
    Ju, Xiangyang
    Chen, Hai-Yan
    Bo, Bai
    Li, XinXu
    INTERNATIONAL SURGERY, 2018, 103 (9-10) : 493 - 504
  • [2] Biomechanical analysis for five fixation techniques of Pauwels-III fracture by finite element modeling
    Peng, Matthew Jian-Qiao
    Xu, HongWen
    Chen, Hai-Yan
    Lin, Ze
    Li, XinXu
    Shen, ChuLong
    Lau, YongQiang
    He, ErXing
    Guo, YueMing
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2020, 193
  • [3] A Comparative Finite-Element Analysis of Bone Failure and Load Transfer of Osseointegrated Prostheses Fixations
    P. K. Tomaszewski
    N. Verdonschot
    S. K. Bulstra
    G. J. Verkerke
    Annals of Biomedical Engineering, 2010, 38 : 2418 - 2427
  • [4] A Comparative Finite-Element Analysis of Bone Failure and Load Transfer of Osseointegrated Prostheses Fixations
    Tomaszewski, P. K.
    Verdonschot, N.
    Bulstra, S. K.
    Verkerke, G. J.
    ANNALS OF BIOMEDICAL ENGINEERING, 2010, 38 (07) : 2418 - 2427
  • [5] FINITE-ELEMENT ANALYSIS FOR BIOMECHANICAL EFFECTS ON CRANIOFACIAL SKELETON
    MIYASAKA, J
    TANNE, K
    YAMAGATA, Y
    SAKUDA, M
    TSUTSUMI, S
    JOURNAL OF DENTAL RESEARCH, 1987, 66 : 323 - 323
  • [6] Biomechanical comparison of internal fixations in osteoporotic intertrochanteric fracture A finite element analysis
    Yuan, Gao-Xiang
    Shen, Yu-Hui
    Chen, Bo
    Zhang, Wei-Bin
    SAUDI MEDICAL JOURNAL, 2012, 33 (07) : 732 - 739
  • [7] Finite Element Analysis of Six Internal Fixations in the Treatment of Pauwels Type III Femoral Neck Fracture
    Sun, Xiang
    Han, Zhe
    Cao, Dongdong
    Han, Chao
    Xie, Mengqi
    Zeng, Xiantie
    Dong, Qiang
    ORTHOPAEDIC SURGERY, 2024, 16 (07) : 1695 - 1709
  • [8] HYDROTHERMAL ANALYSIS BY FINITE-ELEMENT METHOD
    LOZIUK, LA
    ANDERSON, JC
    BELYTCHK.T
    JOURNAL OF THE HYDRAULICS DIVISION-ASCE, 1974, 100 (NHY7): : 1075 - 1076
  • [9] A FINITE-ELEMENT METHOD FOR MICROSTRUCTURAL ANALYSIS
    MUELLER, AC
    COMPOSITES ENGINEERING, 1994, 4 (03): : 361 - 376
  • [10] Nonlinear finite-element analysis and biomechanical evaluation of the lumbar spine
    Wong, C
    Gehrchen, PM
    Darvann, T
    Kiær, T
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (06) : 742 - 746