Operation of NIM5 fountain with 1.5x10-15 uncertainty and design of new NIM6 in NIM

被引:4
作者
Fang, F. [1 ]
Liu, N. [1 ]
Liu, K. [1 ]
Chen, W. [1 ]
Suo, R. [1 ]
Li, Tianchu [1 ]
机构
[1] Natl Inst Metrol, Div Time & Frequency, Beijing, Peoples R China
来源
8TH SYMPOSIUM ON FREQUENCY STANDARDS AND METROLOGY 2015 | 2016年 / 723卷
关键词
1ST ACCURACY EVALUATION; CLOCK;
D O I
10.1088/1742-6596/723/1/012009
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The cesium fountain primary frequency standard NIM5 started to operate since 2008 and started to report to BIPM since 2014. The major constrains of NIM5 is a relatively large background signal at the detection and microwave leakages due to the Ramsey cavity. A new fountain clock NIM6 is under construction. Besides some improvements on the vacuum system, a new Ramsey cavity and a microwave synthesizer are made to reduce the Type B uncertainty. Another feature of NIM6 is collecting atoms from a MOT loading optical molasses to get more atoms with a more uniform density distribution. With a new frequency synthesizer based on cryogenic sapphire oscillator (CSO), NIM6 is aiming to reach the quantum projection noise, thus leading to a reduced Type A uncertainty compared with NIM5.
引用
收藏
页数:5
相关论文
共 13 条
[1]  
Fang F, 2013, JOINT UFFC EFTF PFM, V232
[2]   NIM5 Cs fountain clock and its evaluation [J].
Fang, Fang ;
Li, Mingshou ;
Lin, Pingwei ;
Chen, Weiliang ;
Liu, Nianfeng ;
Lin, Yige ;
Wang, Ping ;
Liu, Kun ;
Suo, Rui ;
Li, Tianchu .
METROLOGIA, 2015, 52 (04) :454-468
[3]   Uncertainty evaluation of the caesium fountain clock PTB-CSF2 [J].
Gerginov, V. ;
Nemitz, N. ;
Weyers, S. ;
Schroeder, R. ;
Griebsch, D. ;
Wynands, R. .
METROLOGIA, 2010, 47 (01) :65-79
[4]   Ultra-low-phase-noise cryocooled microwave dielectric-sapphire-resonator oscillators [J].
Hartnett, John G. ;
Nand, Nitin R. ;
Lu, Chuan .
APPLIED PHYSICS LETTERS, 2012, 100 (18)
[5]   First accuracy evaluation of NIST-F2 [J].
Heavner, Thomas P. ;
Donley, Elizabeth A. ;
Levi, Filippo ;
Costanzo, Giovanni ;
Parker, Thomas E. ;
Shirley, Jon H. ;
Ashby, Neil ;
Barlow, Stephan ;
Jefferts, S. R. .
METROLOGIA, 2014, 51 (03) :174-182
[6]   Accuracy evaluation of ITCsF2: a nitrogen cooled caesium fountain [J].
Levi, Filippo ;
Calonico, Davide ;
Calosso, Claudio E. ;
Godone, Aldo ;
Micalizio, Salvatore ;
Costanzo, Giovanni A. .
METROLOGIA, 2014, 51 (03) :270-284
[7]   Quantum projection noise in an atomic fountain: A high stability cesium frequency standard [J].
Santarelli, C ;
Laurent, P ;
Lemonde, P ;
Clairon, A ;
Mann, AG ;
Chang, S ;
Luiten, AN ;
Salomon, C .
PHYSICAL REVIEW LETTERS, 1999, 82 (23) :4619-4622
[8]   Spin polarization in a freely evolving sample of cold atoms [J].
Szymaniec, Krzysztof ;
Noh, Heung-Ryoul ;
Park, Sang Eon ;
Takamizawa, Akifumi .
APPLIED PHYSICS B-LASERS AND OPTICS, 2013, 111 (03) :527-535
[9]   Primary Frequency Standard NPL-CsF2: Optimized Operation Near the Collisional Shift Cancellation Point [J].
Szymaniec, Krzysztof ;
Park, Sang Eon .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (07) :2475-2481
[10]   First accuracy evaluation of the NPL-CsF2 primary frequency standard [J].
Szymaniec, Krzysztof ;
Park, Sang Eon ;
Marra, Giuseppe ;
Chalupczak, Witold .
METROLOGIA, 2010, 47 (04) :363-376