Treatment of seven patients with coarctation of the aorta treated using a Mexican-made platinum/iridium stent

被引:3
作者
Abundes-Velasco, Arturo
Zabal-Cerdeira, Carlos
Garcia-Montes, Jose Antonio
De los Rios-Ibarra, Manuel Odin
Gallegos, Joel Estrada
Pena, German Quintana
机构
[1] Hosp Cardiol, Hemodynam Unit, Ctr Med Nacl Siglo 21, Mexico City, DF, Mexico
[2] Inst Nacl Cardiol Ignacio Chavez, Pediat Cardiol Clin Serv, Mexico City, DF, Mexico
[3] Hosp Reg SSA Culiacan, Cardiol Clin Serv, Sinaloa, Mexico
关键词
coarctation of the aorta; platinum/iridium stent;
D O I
10.1016/j.arcmed.2007.06.002
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background. Nowadays, stenting is the treatment of choice in patients > 15 years of age with coarctation of the aorta without hypoplastic aortic isthmus. The platinum/iridium stent manufactured in Mexico may be an affordable alternative with the same benefits as the imported stent. Methods. This is a series of cases in which we present the immediate results of the first seven patients with coarctation of the aorta treated with the platinum/iridium stent manufactured in Mexico. The first four patients were selected and treated at the Cardiology Hospital of the National Medical Center during the year 2003, two more patients were selected and treated at the National Institute of Cardiology, and another patient was treated at the Military Hospital, Mexico City, in the year 2005. Results. A total of seven patients were successfully treated with the platinum/iridiurn stent without any complications. Average initial gradient was 56.4 mmHg (range: 30-90 mmHg). In six patients, the final gradient was 0.0 mmHg and in one it was 2 mmHg. In the four patients treated in the Cardiology Hospital of the National Medical Center, mean follow-up time was 17.5 +/- 2.5 months, with a gradient of 0.0 mmHg. All patients are in NYHA Class I. At 12 months follow-up, the gradient remains as 0 mmHg and the patients are still in NYHA Class I functional state. Conclusions. This stent can be a safe and effective alternative to other stents available on the market for the treatment of coarctation of the aorta. (c) 2007 IMSS. Published by Elsevier Inc.
引用
收藏
页码:853 / 857
页数:5
相关论文
共 18 条
[1]   Immediate and long-term results of implantation of the new platinum coronary stent (Atlas stent) in patients with coronary artery disease [J].
Abundes, A ;
Rivera, JD ;
Arizmendi, E ;
Farell, J ;
Ledesma, M ;
Montoya, S .
REVISTA ESPANOLA DE CARDIOLOGIA, 2002, 55 (11) :1205-1208
[2]  
ABUNDES VA, 2001, MULTIPLES FACETAS IN, P71
[3]   Results and mid-long-term follow-up of stent implantation for native and recurrent coarctation of the aorta [J].
Chessa, M ;
Carrozza, M ;
Butera, G ;
Piazza, L ;
Negura, DG ;
Bussadori, C ;
Bossone, E ;
Giamberti, A ;
Carminati, M .
EUROPEAN HEART JOURNAL, 2005, 26 (24) :2728-2732
[4]  
DIETHRICH EB, 1995, J ENDOVASC SURG, V2, P183, DOI 10.1583/1074-6218(1995)002<0183:ETIAAC>2.0.CO
[5]  
2
[6]   Use of balloon-expandable stents for coarctation of the aorta: Initial results and intermediate-term follow-up [J].
Ebeid, MR ;
Prieto, LR ;
Latson, LA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 30 (07) :1847-1852
[7]  
HIJAZI ZM, 1995, J INVASIVE CARDIOL, V7, P127
[8]  
Ledesma Mariano, 1997, Archivos del Instituto de Cardiologia de Mexico, V67, P399
[9]  
Ledesma Velasco M, 1992, Arch Inst Cardiol Mex, V62, P339
[10]  
Munayer Calderón Jaime, 2002, Arch. Cardiol. Méx., V72, P20