Dynamically Shaped Magnetic Fields Initial Animal Validation of a New Remote Electrophysiology Catheter Guidance and Control System

被引:72
作者
Gang, Eli S. [1 ]
Bich Lien Nguyen [1 ,4 ]
Shachar, Yehoshua [2 ]
Farkas, Leslie [2 ]
Farkas, Laszlo [2 ]
Marx, Bruce [2 ]
Johnson, David [2 ]
Fishbein, Michael C. [3 ]
Gaudio, Carlo [4 ]
Kim, Steven J. [5 ]
机构
[1] Cedars Sinai Med Ctr, Dept Med, Div Cardiol, Electrophysiol Sect, Los Angeles, CA 90048 USA
[2] Magnetecs Corp, Inglewood, CA USA
[3] Univ Calif Los Angeles, David Geffen Sch Med, Dept Pathol & Lab Med, Los Angeles, CA 90095 USA
[4] Sapienza Univ Rome, Heart & Great Vessels Dept, Umberto Hosp, Rome, Italy
[5] St Jude Med, St Paul, MN USA
关键词
catheter ablation; atrial fibrillation; remote magnetic navigation; mapping; imaging; electrophysiology; ATRIAL-FIBRILLATION ABLATION; NAVIGATION;
D O I
10.1161/CIRCEP.110.959692
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-To address some of the shortcomings of existing remote catheter navigation systems (RNS), a new magnetic RNS has been developed that provides real-time navigation of catheters within the beating heart. The initial experience using this novel RNS in animals is described. Methods and Results-A real-time, high-speed, closed-loop, magnetic RNS system (Catheter Guidance Control and Imaging) comprises 8 electromagnets that create unique dynamically shaped ("lobed") magnetic fields around the subject's torso. The real-time reshaping of these magnetic fields produces the appropriate 3D motion or change in direction of a magnetized electrophysiology ablation catheter within the beating heart. The RNS is fully integrated with the Ensite-NavX 3D electroanatomic mapping system (St Jude Medical) and allows for both joystick and automated navigation. Conventional and remote navigational mapping of the left atrium were performed using a 4-mm-tip ablation catheter in 10 pigs. A multielectrode transseptal sheath allowed for additional motion compensation. Linear and circumferential radiofrequency lesion sets were performed; in a subset of cases, selective pulmonary vein isolation was also performed. Recording and fluoroscopic equipments were unaffected by the magnetic fields generated by Catheter Guidance Control and Imaging. Automated mode navigation was highly reproducible (96 +/- 8.4% of attempts), accurate (1.9 +/- 0.4 mm from target site), and rapid (11.6 +/- 3.5 seconds to reach targets). At postmortem examination, radiofrequency lesion depth was 78.5 +/- 12.1% of atrial wall thickness. Conclusions-A new magnetic RNS using a dynamically shaped magnetic field concept can reproducibly and effectively reach target radiofrequency ablation points within the pig left atrium. Validation of the system in clinical settings is under way. (Circ Arrhythm Electrophysiol. 2011; 4: 770-777.)
引用
收藏
页码:770 / U287
页数:10
相关论文
共 9 条
[1]   Anatomy of the pig heart: comparisons with normal human cardiac structure [J].
Crick, SJ ;
Sheppard, MN ;
Ho, SY ;
Gepstein, L ;
Anderson, RH .
JOURNAL OF ANATOMY, 1998, 193 :105-119
[2]   Remote magnetic navigation - Human experience in pulmonary vein ablation [J].
Di Blase, Luigi ;
Fahmy, Tamer S. ;
Patel, Dimpi ;
Bai, Rong ;
Civello, Kenneth ;
Wazni, Oussama M. ;
Kanj, Mohamed ;
Elayi, Claude S. ;
Ching, Chi Keong ;
Khan, Mohamed ;
Popova, Lucie ;
Schweikert, Robert A. ;
Cummings, Jennifer E. ;
Burkhardt, J. David ;
Martin, David O. ;
Bhargava, Mandeep ;
Dresing, Thomas ;
Saliba, Walid ;
Arruda, Mauricio ;
Natale, Andrea .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (09) :868-874
[3]   Initial experience with remote catheter ablation using a novel magnetic navigation system -: Magnetic remote catheter ablation [J].
Ernst, S ;
Ouyang, FF ;
Linder, C ;
Hertting, K ;
Stahl, F ;
Chun, J ;
Hachiya, H ;
Bänsch, D ;
Antz, M ;
Kuck, KH .
CIRCULATION, 2004, 109 (12) :1472-1475
[4]   Novel, magnetically guided catheter for endocardial mapping and radiofrequency catheter ablation [J].
Faddis, MN ;
Blume, W ;
Finney, J ;
Hall, A ;
Rauch, J ;
Sell, J ;
Bae, KT ;
Talcott, M ;
Lindsay, B .
CIRCULATION, 2002, 106 (23) :2980-2985
[5]   Remote Magnetic Navigation With Irrigated Tip Catheter for Ablation of Paroxysmal Atrial Fibrillation [J].
Miyazaki, Shinsuke ;
Shah, Ashok J. ;
Xhaet, Olivier ;
Derval, Nicolas ;
Matsuo, Seiichiro ;
Wright, Matthew ;
Nault, Isabelle ;
Forclaz, Andrei ;
Jadidi, Amir S. ;
Knecht, Sebastien ;
Rivard, Lena ;
Liu, Xingpeng ;
Linton, Nick ;
Sacher, Frederic ;
Hocini, Meleze ;
Jais, Pierre ;
Haissaguerre, Michel .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2010, 3 (06) :585-589
[6]   Robotic magnetic navigation for atrial fibrillation ablation [J].
Pappone, C ;
Vicedomini, G ;
Manguso, F ;
Gugliotta, F ;
Mazzone, P ;
Gulletta, S ;
Sora, N ;
Sala, S ;
Marzi, A ;
Augello, G ;
Livolsi, L ;
Santagostino, A ;
Santinelli, V .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (07) :1390-1400
[7]   View-synchronized robotic image-guided therapy for atrial fibrillation ablation - Experimental validation and clinical feasibility [J].
Reddy, Vivek Y. ;
Neuzil, Petr ;
Malchano, Zachary J. ;
Vijaykumar, Ragu ;
Cury, Ricardo ;
Abbara, Suhny ;
Weichet, Jiri ;
McPherson, Christina D. ;
Ruskin, Jeremy N. .
CIRCULATION, 2007, 115 (21) :2705-2714
[8]   Remote Magnetic Catheter Navigation for Cavotricuspid Isthmus Ablation in Patients With Common-Type Atrial Flutter [J].
Vollmann, Dirk ;
Luethje, Lars ;
Seegers, Joachim ;
Hasenfuss, Gerd ;
Zabel, Markus .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2009, 2 (06) :603-610
[9]   Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus [J].
Yokoyama, Katsuaki ;
Nakagawa, Hiroshi ;
Shah, Dipen C. ;
Lambert, Hendrik ;
Leo, Giovanni ;
Aeby, Nicolas ;
Ikeda, Atsushi ;
Pitha, Jan V. ;
Sharma, Tushar ;
Lazzara, Ralph ;
Jackman, Warren M. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2008, 1 (05) :354-362