Physically Tailoring Ion Fluxes by Introducing Foamlike Structures into Polymeric Membranes of Solid Contact Ion-Selective Electrodes

被引:8
作者
Cheong, Yi Heng [1 ,2 ,3 ]
Lisak, Grzegorz [1 ,2 ]
机构
[1] Nanyang Technol Univ, Coll Engn, Sch Civil & Environm Engn, Singapore 639798, Singapore
[2] Nanyang Environm & Water Res Inst, Residues & Resource Reclamat Ctr, Singapore 637141, Singapore
[3] Robert Bosch South East Asia Pte Ltd, Singapore 573943, Singapore
关键词
ion-selective electrode; potential stability; potential reproducibility; membrane modification; ion fluxes; LOWER DETECTION LIMIT; WATER LAYER; POTENTIOMETRIC SENSORS; DIFFUSION; TRANSPORT; RECOMMENDATIONS; COEFFICIENTS; TRANSDUCER; CHALLENGES; GRAPHENE;
D O I
10.1021/acssensors.1c01413
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Transmembrane ion fluxes have earlier been identified as a source of potential instability in solid contact ion-selective electrodes (SC-ISEs). In this work, foamlike structures were intentionally introduced into a potassium-sensitive plasticized poly(vinyl chloride) ion-selective membrane (ISM) near the membrane vertical bar solid contact interface by controlling the temperature during membrane deposition. Foamlike structures in the ISM were shown to be effective at physically tailoring the transport of ions in the ion-selective membrane, greatly reducing the flux of interfering ions from the sample to the membrane vertical bar solid contact interface. The drifts during a conventional water layer test were hence able to be greatly mitigated, even with SC-ISEs incorporating a relatively hydrophilic poly(3,4-ethylenedioxythiophene) doped with poly(styrenesulfonate) (PEDOT:PSS) solid contact. In solutions with a high background concentration of interfering ions, equilibrated ion-selective electrodes with foamlike membranes were able to reproduce their initial potentials within 0.6 mV uncertainty (n = 3) from 0 to 18 h. This was achieved despite sensor exposure to solutions exceeding the selectivity limit of the ISEs in 3 h intervals, allowing improvement of the potential reproducibility of the sensors. Since the introduction of foamlike structures into ISM is linked to temperature-controlled membrane deposition, it is envisaged that the method is generally applicable to all solid contact ion-selective electrodes that are based on polymeric membranes and require membrane deposition from the cocktail solution.
引用
收藏
页码:3667 / 3676
页数:10
相关论文
共 70 条
[41]  
Makarychev-Mikhailov S, 2008, ELECTROCHEMICAL SENSORS, BIOSENSORS AND THEIR BIOMEDICAL APPLICATIONS, P71, DOI 10.1016/B978-012373738-0.50006-4
[42]   Effect of transmembrane electrolyte diffusion on the detection limit of carrier-based potentiometric ion sensors [J].
Mathison, S ;
Bakker, E .
ANALYTICAL CHEMISTRY, 1998, 70 (02) :303-309
[44]   Effect of interferents present in the internal solution or in the conducting polymer transducer on the responses of ion-selective electrodes [J].
Michalska, A ;
Ocypa, M ;
Maksymiuk, K .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 385 (01) :203-207
[46]   Highly selective all-plastic, disposable, Cu2+-selective electrodes [J].
Michalska, A ;
Ocypa, M ;
Maksymiuk, K .
ELECTROANALYSIS, 2005, 17 (04) :327-333
[47]   PEDOT films: multifunctional membranes for electrochemical ion sensing [J].
Michalska, A ;
Galuszkiewicz, A ;
Ogonowska, M ;
Ocypa, M ;
Maksymiuk, K .
JOURNAL OF SOLID STATE ELECTROCHEMISTRY, 2004, 8 (06) :381-389
[48]   Effects of ion transport on the potential response of ionophore-based membrane electrodes: A theoretical approach [J].
Morf, WE ;
Badertscher, M ;
Zwickl, T ;
de Rooij, NF ;
Pretsch, E .
JOURNAL OF PHYSICAL CHEMISTRY B, 1999, 103 (51) :11346-11356
[49]  
NIKOLSKII BP, 1985, ION SEL ELECTRODE R, V7, P3
[50]   The use of graphene oxide as a fixed charge carrier in ion-selective electrodes [J].
O'Neil, Glen D. ;
Fouskaki, Maria ;
Kounaves, Samuel P. ;
Chaniotakis, Nikos A. .
ELECTROCHEMISTRY COMMUNICATIONS, 2015, 55 :51-54