Development of a video-guided real-time patient motion monitoring system

被引:5
作者
Ju, Sang Gyu [1 ]
Huh, Woong [2 ]
Hong, Chae-Seon [1 ]
Kim, Jin Sung [1 ]
Shin, Jung Suk [1 ]
Shin, Eunhyuk [1 ]
Han, Youngyih [1 ]
Ahn, Yong Chan [1 ]
Park, Hee Chul [1 ]
Choi, Doo Ho [1 ]
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Radiat Oncol, Seoul 135710, South Korea
[2] Myongji Univ, Dept Elect Engn, Gyeonggido 449728, South Korea
关键词
IGRT; patient motion monitoring; patient care; real-time frame difference; radiation safety; RADIATION-THERAPY; POSITIONING SYSTEM;
D O I
10.1118/1.3700734
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The authors developed a video image-guided real-time patient motion monitoring (VGRPM) system using PC-cams, and its clinical utility was evaluated using a motion phantom. Methods: The VGRPM system has three components: (1) an image acquisition device consisting of two PC-cams, (2) a main control computer with a radiation signal controller and warning system, and (3) patient motion analysis software developed in-house. The intelligent patient motion monitoring system was designed for synchronization with a beam on/off trigger signal in order to limit operation to during treatment time only and to enable system automation. During each treatment session, an initial image of the patient is acquired as soon as radiation starts and is compared with subsequent live images, which can be acquired at up to 30 fps by the real-time frame difference-based analysis software. When the error range exceeds the set criteria (delta(movement)) due to patient movement, a warning message is generated in the form of light and sound. The described procedure repeats automatically for each patient. A motion phantom, which operates by moving a distance of 0.1, 0.2, 0.3, 0.5, and 1.0 cm for 1 and 2 s, respectively, was used to evaluate the system performance. The authors measured optimal delta(movement) for clinical use, the minimum distance that can be detected with this system, and the response time of the whole system using a video analysis technique. The stability of the system in a linear accelerator unit was evaluated for a period of 6 months. Results: As a result of the moving phantom test, the delta(movement) for detection of all simulated phantom motion except the 0.1 cm movement was determined to be 0.2% of total number of pixels in the initial image. The system can detect phantom motion as small as 0.2 cm. The measured response time from the detection of phantom movement to generation of the warning signal was 0.1 s. No significant functional disorder of the system was observed during the testing period. Conclusions: The VGRPM system has a convenient design, which synchronizes initiation of the analysis with a beam on/off signal from the treatment machine and may contribute to a reduction in treatment error due to patient motion and increase the accuracy of treatment dose delivery. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.3700734]
引用
收藏
页码:2396 / 2404
页数:9
相关论文
共 8 条
[1]   Technical aspects of image-guided respiration-gated radiation therapy [J].
Jiang, SB .
MEDICAL DOSIMETRY, 2006, 31 (02) :141-151
[2]   Initial clinical experience with a video-based patient positioning system [J].
Johnson, LS ;
Milliken, BD ;
Hadley, SW ;
Pelizzari, CA ;
Haraf, DJ ;
Chen, GTY .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (01) :205-213
[3]   SUCCESSFUL IMPLEMENTATION OF IMAGE-GUIDED RADIATION THERAPY QUALITY ASSURANCE IN THE TRANS TASMAN RADIATION ONCOLOGY GROUP 08.01 PROFIT STUDY [J].
Middleton, Mark ;
Frantzis, Jim ;
Healy, Brendan ;
Jones, Mark ;
Murry, Rebecca ;
Kron, Tomas ;
Plank, Ashley ;
Catton, Charles ;
Martin, Jarad .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (05) :1576-1581
[4]   Performance of a video-image-subtraction-based patient positioning system [J].
Milliken, BD ;
Rubin, SJ ;
Hamilton, RJ ;
Johnson, LS ;
Chen, GTY .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :855-866
[5]   AMERICAN SOCIETY FOR THERAPEUTIC RADIOLOGY AND ONCOLOGY (ASTRO) AND AMERICAN COLLEGE OF RADIOLOGY (ACR) PRACTICE GUIDELINES FOR IMAGE-GUIDED RADIATION THERAPY (IGRT) [J].
Potters, Louis ;
Gaspar, Laurie E. ;
Kavanagh, Brian ;
Galvin, James M. ;
Hartford, Alan C. ;
Hevezi, James M. ;
Kupelian, Patrick A. ;
Mohiden, Najeeb ;
Samuels, Michael A. ;
Timmerman, Robert ;
Tripuraneni, Prabhakar ;
Vlachaki, Maria T. ;
Xing, Lei ;
Rosenthal, Seth A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (02) :319-325
[6]   Analysis of head motion prior to and during proton beam therapy [J].
Schulte, RW ;
Fargo, RA ;
Meinass, HJ ;
Slater, JD ;
Slater, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (04) :1105-1110
[7]   A Survey of Image-Guided Radiation Therapy Use in the United States [J].
Simpson, Daniel R. ;
Lawson, Joshua D. ;
Nath, Sameer K. ;
Rose, Brent S. ;
Mundt, Arno J. ;
Mell, Loren K. .
CANCER, 2010, 116 (16) :3953-3960
[8]   An investigation of a video-based patient repositioning technique [J].
Yan, YL ;
Song, YL ;
Boyer, AL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 54 (02) :606-614