Voltammetric and amperometric determination of metoclopramide on boron-doped diamond film electrode

被引:9
作者
Dejmkova, Hana [1 ]
Dag, Candan [2 ]
Barek, Jiri [1 ]
Zima, Jiri [1 ]
机构
[1] Charles Univ Prague, Dept Analyt Chem, Fac Sci, UNESCO Lab Environm Electrochem, CZ-12843 Prague 2, Czech Republic
[2] Anadolu Univ, Dept Chem, Fac Sci, TR-26470 Eskisehir, Turkey
来源
CENTRAL EUROPEAN JOURNAL OF CHEMISTRY | 2012年 / 10卷 / 04期
关键词
Metoclopramide; Differential pulse voltammetry; Flow injection analysis; Boron-doped diamond film electrode; Passivation; GLASSY-CARBON ELECTRODE; INJECTION SPECTROPHOTOMETRIC DETERMINATION; PERFORMANCE LIQUID-CHROMATOGRAPHY; TANDEM MASS-SPECTROMETRY; ELECTROCHEMICAL DETECTION; PASTE ELECTRODE; POTENTIOMETRIC DETERMINATION; PHARMACEUTICAL FORMULATIONS; MEMBRANE SENSOR; HUMAN PLASMA;
D O I
10.2478/s11532-012-0056-8
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
New methods for the determination of metoclopramide, antiemetic and gastroprokinetic pharmaceutical, were developed, using differential pulse voltammetry (DPV) and flow injection analysis (FIA) with amperometric detection on a boron-doped diamond film electrode. Electrode pretreatment necessary to ensure the stable results was investigated and it was found, that while DPV requires frequent electrode cleaning, FIA with a sufficiently high flow rate can maintain a stable signal with no signs of electrode passivation. The calculated quantification limits of the DPV and FIA with amperometric detection were 0.13 mu mol L-1 and 0.015 mmol L-1, respectively. The applicability of the new methods was verified by the determination of metoclopramide in a pharmaceutical preparation. FIA with amperometic detection proved to be sensitive, accurate and, due to the resistance of the electrode to the passivation, also simple to handle.
引用
收藏
页码:1310 / 1317
页数:8
相关论文
共 39 条
[1]   Simultaneous quantification of doxorubicin, lorazepam, metoclopramide, ondansetron, and ranitidine in mixtures by liquid chromatography-tandem mass spectrometry [J].
Abdel-Hamid, ME ;
Sharma, D .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2004, 27 (04) :641-660
[2]   METOCLOPRAMIDE - PHARMACOLOGY AND CLINICAL-APPLICATION [J].
ALBIBI, R ;
MCCALLUM, RW .
ANNALS OF INTERNAL MEDICINE, 1983, 98 (01) :86-95
[3]  
Aleksic MM, 2012, CURR ANAL CHEM, V8, P133
[4]   Spectrofluorimetric Assessment of Metoclopramide Hydrochloride Using Terbium Doped in PMMA Matrix Optical Sensor [J].
Attia, M. S. ;
Othman, A. M. ;
Elraghi, E. ;
Aboul-Enein, Hassan Y. .
JOURNAL OF FLUORESCENCE, 2011, 21 (02) :739-745
[5]   Highly sensitive and selective spectrofluorimetric determination of metoclopramide hydrochloride in pharmaceutical tablets and serum samples using Eu3+ ion doped in sol-gel matrix [J].
Attia, M. S. ;
Aboaly, M. M. .
TALANTA, 2010, 82 (01) :78-84
[6]   DETERMINATION OF OPIATES AND OTHER BASIC DRUGS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ELECTROCHEMICAL DETECTION [J].
BESNER, JG ;
BAND, C ;
RONDEAU, JJ ;
YAMLAHI, L ;
CAILLE, G ;
VARIN, F ;
STEWART, J .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1989, 7 (12) :1811-1817
[7]   Sensitive quantification of chosen drugs by reversed-phase chromatography with electrochemical detection at a glassy carbon electrode [J].
Chmielewska, Aleksandra ;
Konieczna, Lucyna ;
Plenis, Alina ;
Lamparczyk, Henryk .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 839 (1-2) :102-111
[8]   A new sensitive reversed-phase high-performance liquid chromatography method for the quantitative determination of metoclopramide in canine plasma [J].
Cossu, Massimo ;
Sanna, Vanna ;
Gavini, Elisabetta ;
Rassu, Giovanna ;
Giunchedi, Paolo .
ANALYTICAL LETTERS, 2008, 41 (05) :767-778
[9]  
Dejmkova H, 2011, INT J ELECTROCHEM SC, V6, P3550
[10]   Determination of Total Phenols in Foods by Boron Doped Diamond Electrode [J].
Dejmkova, Hana ;
Scampicchio, Matteo ;
Zima, Jiri ;
Barek, Jiri ;
Mannino, Saverio .
ELECTROANALYSIS, 2009, 21 (09) :1014-1018