Development of a new electron ionization/field ionization ion source for gas chromatography/time-of-flight mass spectrometry

被引:8
|
作者
Miyamoto, Kenji [1 ]
Fujimaki, Susumu [2 ]
Ueda, Yoshihisa [2 ]
机构
[1] Naruto Univ Educ, Naruto, Tokushima 7728502, Japan
[2] JEOL Ltd, Tokyo 1968558, Japan
关键词
FIELD-IONIZATION; SPECTRA;
D O I
10.1002/rcm.4256
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We have developed a combined EI/FI source for gas chromatography/orthogonal acceleration time-of-flight mass spectrometry (GC/oaTOFMS). In general, EI (electron ionization) and FI (field ionization) mass spectra are complementary: the EI mass spectrum contains information about fragment ions, while the FI mass spectrum contains information about molecular ions. Thus, the comparative study of EI and FI mass spectra is useful for GC/MS analyses. Unlike the conventional ion sources for FI and EI measurements, the newly developed source can be used for both measurements without breaking the ion source vacuum or changing the ion source. Therefore, the combined EI/FI source is more preferable than the conventional EI or FI ion source from the viewpoint of the reliability of measurements and facility of operation. Using the combined EI/FI source, the complementarity between EI and FI mass spectra is demonstrated experimentally with n-hexadecane (100 pg): characteristic fragment ions for the n-alkane such as m/z 43, 57, 71, and 85 are obtained in the EI mass spectrum, while only the parent peak of m/z 226 (M(+)) without any fragment ions is observed in the FI mass spectrum. Moreover, the field desorption (FD) measurement is also demonstrated with poly(ethylene glycol)s M600 (10 ng) and M1000 (15 ng). Signals of [M+H](+), [M+Na](+) and [M+K](+) are clearly detected in the FD mass spectra. Copyright (C) 2009 John Wiley & Sons, Ltd.
引用
收藏
页码:3350 / 3354
页数:5
相关论文
共 50 条
  • [1] Compatibility of electron ionization and soft ionization methods in gas chromatography/orthogonal time-of-flight mass spectrometry
    Hejazi, Leila
    Ebrahimi, Diako
    Hibbert, D. Brynn
    Guilhaus, Michael
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2009, 23 (14) : 2181 - 2189
  • [2] Gas sampling glow discharge: A versatile ionization source for gas chromatography time-of-flight mass spectrometry
    Guzowski, JP
    Hieftje, GM
    ANALYTICAL CHEMISTRY, 2000, 72 (16) : 3812 - 3820
  • [3] Gas Chromatography/Femtosecond Multiphoton Ionization/Time-of-Flight Mass Spectrometry of Dioxins
    Watanabe-Ezoe, Yuka
    Li, Xing
    Imasaka, Tomoko
    Uchimura, Tomohiro
    Imasaka, Totaro
    ANALYTICAL CHEMISTRY, 2010, 82 (15) : 6519 - 6525
  • [4] Gas chromatography/multiphoton ionization/time-of-flight mass spectrometry of polychlorinated biphenyls
    Matsui, Taiki
    Uchimura, Tomohiro
    Imasaka, Totaro
    ANALYTICA CHIMICA ACTA, 2011, 694 (1-2) : 108 - 114
  • [5] Superimposition of a magnetic field around an ion guide for electron ionization time-of-flight mass spectrometry
    Yue, BF
    Lee, ED
    Rockwood, AL
    Lee, ML
    ANALYTICAL CHEMISTRY, 2005, 77 (13) : 4167 - 4175
  • [6] Determination of elemental compositions by gas chromatography/time-of-flight mass spectrometry using chemical and electron ionization
    Abate, Salvatore
    Ahn, Yun Gyong
    Kind, Tobias
    Cataldi, Tommaso R. I.
    Fiehn, Oliver
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2010, 24 (08) : 1172 - 1180
  • [7] Gas chromatography/multiphoton ionization/time-of-flight mass spectrometry using a femtosecond laser
    Imasaka, Totaro
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2013, 405 (22) : 6907 - 6912
  • [8] Gas chromatography/multiphoton ionization/time-of-flight mass spectrometry using a femtosecond laser
    Totaro Imasaka
    Analytical and Bioanalytical Chemistry, 2013, 405 : 6907 - 6912
  • [9] Development of a tandem time-of-flight mass spectrometer with an electrospray ionization ion source
    Nagao, Hirofumi
    Shimma, Shuichi
    Hayakawa, Shigeo
    Awazu, Kunio
    Toyoda, Michisato
    JOURNAL OF MASS SPECTROMETRY, 2010, 45 (08): : 937 - 943
  • [10] An electrodynamic ion funnel for electrospray ionization source based time-of-flight mass spectrometry
    Bhushan, K. G.
    Rao, K. C.
    Sule, U.
    Reddy, P.
    Rodrigues, S. M.
    Gaikwad, D. T.
    Mukundhan, R.
    Gupta, S. K.
    JOURNAL OF INSTRUMENTATION, 2016, 11