An alternative derivation of Katsevich's cone-beam reconstruction formula

被引:55
作者
Chen, GH [1 ]
机构
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53792 USA
关键词
D O I
10.1118/1.1628413
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this paper an alternative derivation of Katsevich's cone-beam image reconstruction algorithm is presented. The starting point is the classical Tuy's inversion formula. After (i) using the hidden symmetries of the intermediate functions, (ii) handling the redundant data by weighting them, (iii) changing the weighted average into an integral over the source trajectory parameter, and (iv) imposing an additional constraint on the weighting function, a filtered backprojection reconstruction formula from cone beam projections is derived. The following features are emphasized in the present paper: First, the nontangential condition in Tuy's original data sufficiency conditions has been relaxed. Second, a practical regularization scheme to handle the singularity is proposed. Third, the derivation in the cone beam case is in the same fashion as that in the fan-beam case. Our final cone-beam reconstruction formula is the same as the one discovered by Katsevich in his most recent paper. However, the data sufficiency conditions and the regularization scheme of singularities are different. A detailed comparison between these two methods is presented. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:3217 / 3226
页数:10
相关论文
共 16 条
[1]   3-DIMENSIONAL RECONSTRUCTION FROM CONE-BEAM DATA IN O(N-3 LOG-N) TIME [J].
AXELSSON, C ;
DANIELSSON, PE .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (03) :477-491
[2]   A new framework of image reconstruction from fan beam projections [J].
Chen, GH .
MEDICAL PHYSICS, 2003, 30 (06) :1151-1161
[3]  
CLACK R, 1994, P SOC PHOTO-OPT INS, V2299, P230, DOI 10.1117/12.179253
[4]   A CONE-BEAM RECONSTRUCTION ALGORITHM USING SHIFT-VARIANT FILTERING AND CONE-BEAM BACKPROJECTION [J].
DEFRISE, M ;
CLACK, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) :186-195
[5]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[6]  
GRANGEAT P, 1991, LECT NOTES MATH, V1497, P66
[7]   Theoretically exact filtered backprojection-type inversion algorithm for spiral CT [J].
Katsevich, A .
SIAM JOURNAL ON APPLIED MATHEMATICS, 2002, 62 (06) :2012-2026
[8]   Analysis of an exact inversion algorithm for spiral cone-beam CT [J].
Katsevich, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (15) :2583-2597
[9]  
Katsevich A, 2003, INT J MATH MATH SCI, V2003, DOI DOI 10.1155/S0161171203209315
[10]   DERIVATION AND IMPLEMENTATION OF A CONE-BEAM RECONSTRUCTION ALGORITHM FOR NONPLANAR ORBITS [J].
KUDO, H ;
SAITO, T .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) :196-211