Vessel membrane segmentation and calcification location in intravascular ultrasound images using a region detector and an effective selection strategy

被引:7
作者
Huang, Yi [1 ]
Yan, Wenjun [1 ]
Xia, Menghua [1 ]
Guo, Yi [1 ,2 ]
Zhou, Guohui [1 ,2 ]
Wang, Yuanyuan [1 ,2 ]
机构
[1] Fudan Univ, Dept Elect Engn, Shanghai 200433, Peoples R China
[2] Fudan Univ, Key Lab Med Imaging Comp & Comp Assisted Interven, Shanghai 200433, Peoples R China
关键词
Region detector; Selection strategy; Vessel membrane segmentation; Calcification location; ADVENTITIA BORDER DETECTION; AUTOMATIC SEGMENTATION; ARTERIAL-WALLS; LUMEN; IVUS;
D O I
10.1016/j.cmpb.2020.105339
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Background and objective: Segmenting vessel membranes and locating the calcific region in intravascular ultrasound (IVUS) images aid physicians in the diagnosis of atherosclerosis. However, the manual extraction of the media adventitia (MA)/lumen border and calcification location are cumbersome due to the excessive number of IVUS frames. Moreover, most existing (semi-)automatic detection methods cannot achieve both vessel membrane extraction and calcification location simultaneously, and they are unable to detect vessel membranes in IVUS frames from different acquisition systems. Method: A fully automatic approach is proposed based on extremal regions and a flexible selection strategy to extract vessel membranes in different IVUS frames and locate the calcific region in high-frequency ones. Three main steps are included in the algorithm. First, a region detector is employed to extract extremal regions from an IVUS image. Then, according to the selection strategy, a part of the extracted regions is selected. At the same time, the calcification is located according to its special acoustic properties. Next, approximate MA and lumen border segmentation is achieved based on the selected extremal regions and the located calcification in polar coordinates. Finally, the final segmentation results are obtained by smoothing the approximate values. Result: To demonstrate the feasibility of the method, it was evaluated based on a standard public dataset. Furthermore, to quantitatively evaluate the segmentation performance, the Hausdorff distance (HD), Jaccard measure (JM) and percentage of area difference (PAD) were used. The results show that a mean HD of 1.13/1.21 mm, a mean JM of 0.83/0.77 and a mean PAD of 0.11/0.23 are achieved for MA/lumen border detection in 77 40-MHz IVUS images. For MA/lumen border extraction in 435 20-MHz IVUS frames, the average HD, JM and PAD values are 0.47/0.28 mm, 0.84/0.89 and 0.13/0.10, respectively. In addition, the approach successfully achieves calcification location in 40-MHz IVUS frames. In comparison with other published methods, the method proposed in this study is competitive. Conclusion: According to these results, our strategy can extract MA/lumen borders in different IVUS frames and effectively locate calcification in high-frequency IVUS frames. (C) 2020 Elsevier B.V. All rights reserved.
引用
收藏
页数:13
相关论文
共 34 条
[1]   DEEP LEARNING-BASED FEMORAL CARTILAGE AUTOMATIC SEGMENTATION IN ULTRASOUND IMAGING FOR GUIDANCE IN ROBOTIC KNEE ARTHROSCOPY [J].
Antico, M. ;
Sasazawa, F. ;
Dunnhofer, M. ;
Camps, S. M. ;
Jaiprakash, A. T. ;
Pandey, A. K. ;
Crawford, R. ;
Carneiro, G. ;
Fontanarosa, D. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2020, 46 (02) :422-435
[2]   Standardized evaluation methodology and reference database for evaluating IVUS image segmentation [J].
Balocco, Simone ;
Gatta, Carlo ;
Ciompi, Francesco ;
Wahle, Andreas ;
Radeva, Petia ;
Carlier, Stephane ;
Unal, Gozde ;
Sanidas, Elias ;
Mauri, Josepa ;
Carillo, Xavier ;
Kovarnik, Tomas ;
Wang, Ching-Wei ;
Chen, Hsiang-Chou ;
Exarchos, Themis P. ;
Fotiadis, Dimitrios I. ;
Destrempes, Francois ;
Cloutier, Guy ;
Pujol, Oriol ;
Alberti, Marina ;
Mendizabal-Ruiz, E. Gerardo ;
Rivera, Mariano ;
Aksoy, Timur ;
Downe, Richard W. ;
Kakadiaris, Ioannis A. .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2014, 38 (02) :70-90
[3]   Intravascular ultrasound image segmentation:: A three-dimensional fast-marching method based on gray level distributions [J].
Cardinal, MHR ;
Meunier, J ;
Soulez, G ;
Maurice, RL ;
Therasse, É ;
Cloutier, G .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (05) :590-601
[4]   Improving airway segmentation in computed tomography using leak detection with convolutional networks [J].
Charbonnier, Jean-Paul ;
van Rikxoort, Eva M. ;
Setio, Arnaud A. A. ;
Schaefer-Prokop, Cornelia M. ;
van Ginneken, Bram ;
Ciompi, Francesco .
MEDICAL IMAGE ANALYSIS, 2017, 36 :52-60
[5]   Lumen and media-adventitia border detection in IVUS images using texture enhanced deformable model [J].
Chen, Fang ;
Ma, Ruibin ;
Liu, Jia ;
Zhu, Mingyu ;
Liao, Hongen .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2018, 66 :1-13
[6]   HoliMAb: A holistic approach for Media-Adventitia border detection in intravascular ultrasound [J].
Ciompi, Francesco ;
Pujol, Oriol ;
Gatta, Carlo ;
Alberti, Marina ;
Balocco, Simone ;
Carrillo, Xavier ;
Mauri-Ferre, Josepa ;
Radeva, Petia .
MEDICAL IMAGE ANALYSIS, 2012, 16 (06) :1085-1100
[7]   Segmentation method of intravascular ultrasound images of human coronary arteries [J].
Destrempes, Francois ;
Cardinal, Marie-Helene Roy ;
Allard, Louise ;
Tardif, Jean-Claude ;
Cloutier, Guy .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2014, 38 (02) :91-103
[8]  
Dijkstra J, 2001, COMPUT CARDIOL, V28, P25
[9]   Automatic segmentation of cross-sectional coronary arterial images [J].
Essa, Ehab ;
Xie, Xianghua .
COMPUTER VISION AND IMAGE UNDERSTANDING, 2017, 165 :97-110
[10]  
Falk E., 1992, CIRCULATION, V2, P393