Instrument intercomparison in the high-energy field at the CERN-EU reference field (CERF) facility and comparison with the 2017 FLUKA simulations

被引:10
作者
Dinar, N. [1 ,2 ]
Pozzi, F. [1 ]
Silari, M. [1 ]
Puzo, P. [2 ]
Chiriotti, S. [3 ]
De Saint-Hubert, M. [3 ]
Vanhavere, F. [3 ]
Van Hoey, O. [3 ]
Orchard, G. M. [4 ]
Waker, A. J. [4 ]
机构
[1] CERN, CH-1211 Geneva 23, Switzerland
[2] Univ Paris Saclay, Univ Paris Sud, LAL, CNRS,IN2P3, Orsay, France
[3] Belgian Nucl Res Ctr SCK CEN, Inst Environm Hlth & Safety, Mol, Belgium
[4] Univ Ontario, Inst Technol, Fac Energy Syst & Nucl Sci, Oshawa, ON, Canada
关键词
High-energy; Neutron H*(10); Intercomparison; FLUKA; Rem counters; Bonner sphere spectrometer; TEPC; Microdosimetry; NEUTRON REM COUNTER; LATEST DEVELOPMENTS; RANGE; SPECTROMETER; CALIBRATION; WENDI; METER; LINUS; BEAMS; CODE;
D O I
10.1016/j.radmeas.2018.05.008
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
This paper discusses an instrument intercomparison performed in the high-energy field at the CERF facility at CERN between 2015 and 2017 (October 2015, May 2016, November 2016 and June 2017). Measurements were performed in several reference exposure locations with the CERN extended-range Bonner Sphere Spectrometer (BSS), a Berthold LB4611, the LINUS rem counter from CERN, the LUPIN rem counter from ELSE Nuclear, the FHT 762 Wide Energy Neutron Detection Instrument (WENDI-II) from Thermo Scientific, the LUDLUM MODEL 42-41 PRESCILA NEUTRON PROBE and two models of Tissue-equivalent proportional counters (TEPCs) from Far West Technologies. All results are compared with the latest FLUKA reference values from simulations performed in 2017.
引用
收藏
页码:24 / 34
页数:11
相关论文
共 27 条
  • [1] BIRATTARI C, 1992, RADIAT PROT DOSIM, V44, P193
  • [2] AN EXTENDED RANGE NEUTRON REM COUNTER
    BIRATTARI, C
    FERRARI, A
    NUCCETELLI, C
    PELLICCIONI, M
    SILARI, M
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 297 (1-2) : 250 - 257
  • [3] CALIBRATION OF THE NEUTRON REM COUNTER LINUS IN THE ENERGY-RANGE FROM THERMAL TO 19 MEV
    BIRATTARI, C
    ESPOSITO, A
    FERRARI, A
    PELLICCIONI, M
    SILARI, M
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 324 (1-2) : 232 - 238
  • [4] The extended range neutron rem counter 'LINUS': Overview and latest developments
    Birattari, C
    Esposito, A
    Ferrari, A
    Pelliccioni, M
    Rancati, T
    Silari, M
    [J]. RADIATION PROTECTION DOSIMETRY, 1998, 76 (03) : 135 - 148
  • [5] A Bonner Sphere Spectrometer with extended response matrix
    Birattari, C.
    Dimovasili, E.
    Mitaroff, A.
    Silari, M.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 620 (2-3) : 260 - 269
  • [6] The FLUKA Code: Developments and Challenges for High Energy and Medical Applications
    Boehlen, T. T.
    Cerutti, F.
    Chin, M. P. W.
    Fosso, A.
    Ferrari, A.
    Ortega, P. G.
    Mairani, A.
    Sala, P. R.
    Smirnov, G.
    Vlachoudisl, V.
    [J]. NUCLEAR DATA SHEETS, 2014, 120 : 211 - 214
  • [7] The neutron fluence and H*(10) response of the new LB 6411 rem counter
    Burgkhardt, B
    Fieg, G
    Klett, A
    Plewnia, A
    Siebert, BRL
    [J]. RADIATION PROTECTION DOSIMETRY, 1997, 70 (1-4) : 361 - 364
  • [8] A new version of the LUPIN detector: Improvements and latest experimental verification
    Caresana, M.
    Cassell, C.
    Ferrarini, M.
    Hohmann, E.
    Manessi, G. P.
    Mayer, S.
    Silari, M.
    Varoli, V.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (06)
  • [9] LUPIN, a new instrument for pulsed neutron fields
    Caresana, M.
    Ferrarini, M.
    Manessi, G. P.
    Silari, M.
    Varoli, V.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 712 : 15 - 26
  • [10] Single- and multi-foils 27Al(p,3pn)24Na activation technique for monitoring the intensity of high-energy beams
    Curioni, A.
    Froeschl, R.
    Glaser, M.
    Iliopoulou, E.
    La Torre, F. P.
    Pozzi, F.
    Ravotti, F.
    Silari, M.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 858 : 101 - 105