Comparison of phantom materials for use in quality assurance of microbeam radiation therapy

被引:27
|
作者
Cameron, Matthew [1 ]
Cornelius, Iwan [2 ]
Cutajar, Dean [1 ]
Davis, Jeremy [1 ]
Rosenfeld, Anatoly [1 ]
Lerch, Michael [1 ]
Guatelli, Susanna [1 ]
机构
[1] Univ Wollongong, CMRP, Wollongong, NSW 2522, Australia
[2] Australian Synchrotron, Melbourne, Vic 3168, Australia
基金
英国医学研究理事会;
关键词
microbeam radiation therapy; polymer phantoms; dosimetry; Monte Carlo; Geant4; WATER EQUIVALENCE; SOLID PHANTOMS; DOSIMETRY; SIMULATION; RAYS;
D O I
10.1107/S1600577517005641
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Microbeam radiation therapy (MRT) is a promising radiotherapy modality that uses arrays of spatially fractionated micrometre-sized beams of synchrotron radiation to irradiate tumours. Routine dosimetry quality assurance (QA) prior to treatment is necessary to identify any changes in beam condition from the treatment plan, and is undertaken using solid homogeneous phantoms. Solid phantoms are designed for, and routinely used in, megavoltage X-ray beam radiation therapy. These solid phantoms are not necessarily designed to be water-equivalent at low X-ray energies, and therefore may not be suitable for MRT QA. This work quantitatively determines the most appropriate solid phantom to use in dosimetric MRT QA. Simulated dose profiles of various phantom materials were compared with those calculated in water under the same conditions. The phantoms under consideration were RMI457 Solid Water (Gammex-RMI, Middleton, WI, USA), Plastic Water (CIRS, Norfolk, VA, USA), Plastic Water DT (CIRS, Norfolk, VA, USA), PAGAT (CIRS, Norfolk, VA, USA), RW3 Solid Phantom (PTW Freiburg, Freiburg, Germany), PMMA, Virtual Water (Med-Cal, Verona, WI, USA) and Perspex. RMI457 Solid Water and Virtual Water were found to be the best approximations for water in MRT dosimetry (within +/- 3% deviation in peak and 6% in valley). RW3 and Plastic Water DT approximate the relative dose distribution in water (within +/- 3% deviation in the peak and 5% in the valley). PAGAT, PMMA, Perspex and Plastic Water are not recommended to be used as phantoms for MRT QA, due to dosimetric discrepancies greater than 5%.
引用
收藏
页码:866 / 876
页数:11
相关论文
共 50 条
  • [41] Characterization of 3-D-Mesa Silicon Single Strip Detectors for Use in Synchrotron Microbeam Radiation Therapy
    Cameron, Matthew J.
    Davis, Jeremy A.
    Dipuglia, Andrew
    Chartier, Lachlan
    Tran, Linh T.
    Prokopovich, Dale A.
    Petasecca, Marco
    Perevertaylo, Vladimir L.
    Rosenfeld, Anatoly B.
    Lerch, Michael L. F.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2020, 4 (04) : 470 - 478
  • [42] The thermoluminescence response of Ge-doped silica fibres for synchrotron microbeam radiation therapy dosimetry
    Rahman, A. T. Abdul
    Bradley, D. A.
    Doran, S. J.
    Thierry, Brochard
    Braeuer-Krisch, Elke
    Bravin, A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 619 (1-3) : 167 - 170
  • [43] Dosimetric studies of microbeam radiation therapy (MRT) with Monte Carlo simulations
    Siegbahn, EA
    Bräuer-Krisch, E
    Stepanek, J
    Blattmann, H
    Laissue, JA
    Bravin, A
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 548 (1-2) : 54 - 58
  • [44] Memory and survival after microbeam radiation therapy
    Schultke, Elisabeth
    Juurlink, Bernhard H. J.
    Ataelmannan, Khalid
    Laissue, Jean
    Blattmann, Hans
    Brauer-Krisch, Elke
    Bravin, Alberto
    Minczewska, Joanna
    Crosbie, Jeffrey
    Taherian, Hadi
    Frangou, Evan
    Wysokinsky, Tomasz
    Chapman, L. Dean
    Griebel, Robert
    Fourney, Daryl
    EUROPEAN JOURNAL OF RADIOLOGY, 2008, 68 (03) : S142 - S146
  • [45] Hybrid dose calculation: a dose calculation algorithm for microbeam radiation therapy
    Donzelli, Mattia
    Brauer-Krisch, Elke
    Oelfke, Uwe
    Wilkens, Jan J.
    Bartzsch, Stefan
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (04)
  • [46] A Novel Simple Phantom for Verifying the Dose of Radiation Therapy
    Lee, J. H.
    Chang, L. T.
    Shiau, A. C.
    Chen, C. W.
    Liao, Y. J.
    Li, W. J.
    Lee, M. S.
    Hsu, S. M.
    BIOMED RESEARCH INTERNATIONAL, 2015, 2015
  • [47] A Monte Carlo intercomparison of peak-to-valley dose ratios and output factors for microbeam radiation therapy
    Reynard, Dimitri
    Hugtenburg, Richard P.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 176
  • [48] The evaluation of a 2D diode array in "magic phantom" for use in high dose rate brachytherapy pretreatment quality assurance
    Espinoza, A.
    Petasecca, M.
    Fuduli, I.
    Howie, A.
    Bucci, J.
    Corde, S.
    Jackson, M.
    Lerch, M. L. F.
    Rosenfeld, A. B.
    MEDICAL PHYSICS, 2015, 42 (02) : 663 - 673
  • [49] Development of a fast Monte Carlo dose calculation system for online adaptive radiation therapy quality assurance
    Wang, Yuhe
    Mazur, Thomas R.
    Park, Justin C.
    Yang, Deshan
    Mutic, Sasa
    Li, H. Harold
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (12) : 4970 - 4990
  • [50] MOSFET dosimetry for microbeam radiation therapy at the European Synchrotron Radiation Facility
    Bräuer-Krisch, E
    Bravin, A
    Lerch, M
    Rosenfeld, A
    Stepanek, J
    Di Michiel, M
    Laissue, JA
    MEDICAL PHYSICS, 2003, 30 (04) : 583 - 589