Flash glucose monitoring in type 1 diabetes: A comparison with self-monitoring blood glucose

被引:10
作者
Babaya, Naru [1 ]
Noso, Shinsuke [1 ]
Hiromine, Yoshihisa [1 ]
Taketomo, Yasunori [1 ]
Niwano, Fumimaru [1 ]
Yoshida, Sawa [1 ]
Yasutake, Sara [1 ]
Kawabata, Yumiko [1 ]
Ikegami, Hiroshi [1 ]
机构
[1] Kindai Univ, Fac Med, Dept Endocrinol Metab & Diabet, Osaka, Japan
关键词
Flash glucose monitoring; Self-monitoring blood glucose; Type; 1; diabetes; DIAGNOSTIC-CRITERIA; FREESTYLE LIBRE; ACCURACY; COMMITTEE; FULMINANT; MELLITUS; SYSTEMS; SOCIETY;
D O I
10.1111/jdi.13229
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims/Introduction A flash glucose monitoring (FGM) system has become available. To clarify the relationship between FGM and self-monitoring blood glucose (SMBG) values, we compared the two values after simultaneous measurement in Japanese patients with type 1 diabetes, under daily life settings. Materials and Methods A total of 20 outpatients with type 1 diabetes were analyzed. When FGM and SMBG were carried out simultaneously (within +/- 3 min), the values were adopted and each FGM value was matched and compared with the corresponding SMBG value. In addition, we analyzed other cases of simultaneity defined as "within +/- 2 min," "within +/- 1 min" and "at the exact same time." Results The percentage of SMBG and FGM values in the clinically acceptable zone A + B in Clarke and consensus error grid analyses were 97.9 and 99.2%, respectively. Deming regression (x-axis: FGM values, y-axis: SMBG values) determined a slope of 0.9128 (95% confidence interval 0.9008-0.9247) and an intercept of +15.94 mg/dL (95% confidence interval 14.05-17.84). FGM values were lower than SMBG values in the lower glucose range, and higher in the higher glucose range. The shorter the time lag between measurements, the higher the rate of concordance between FGM and SMBG values. Conclusions The results of this study provided evidence on the reliability of FGM in Japanese patients with type 1 diabetes in home conditions. Based on the results, if an abnormal glucose value is detected by FGM, SBMG should then be used to confirm the result.
引用
收藏
页码:1222 / 1229
页数:8
相关论文
共 19 条
[1]   Evaluation of subcutaneous glucose monitoring systems under routine environmental conditions in patients with type 1 diabetes [J].
Aberer, Felix ;
Hajnsek, Martin ;
Rumpler, Markus ;
Zenz, Sabine ;
Baumann, Petra M. ;
Elsayed, Hesham ;
Puffing, Adelheid ;
Treiber, Gerlies ;
Pieber, Thomas R. ;
Sourij, Harald ;
Mader, Julia K. .
DIABETES OBESITY & METABOLISM, 2017, 19 (07) :1051-1055
[2]   The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System [J].
Bailey, Timothy ;
Bode, Bruce W. ;
Christiansen, Mark P. ;
Klaff, Leslie J. ;
Alva, Shridhara .
DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (11) :787-794
[3]   Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range [J].
Battelino, Tadej ;
Danne, Thomas ;
Bergenstal, Richard M. ;
Amiel, Stephanie A. ;
Beck, Roy ;
Biester, Torben ;
Bosi, Emanuele ;
Buckingham, Bruce A. ;
Cefalu, William T. ;
Close, Kelly L. ;
Cobelli, Claudio ;
Dassau, Eyal ;
DeVries, J. Hans ;
Donaghue, Kim C. ;
Dovc, Klemen ;
Doyle, Francis J. ;
Garg, Satish ;
Grunberger, George ;
Heller, Simon ;
Heinemann, Lutz ;
Hirsch, Irl B. ;
Hovorka, Roman ;
Jia, Weiping ;
Kordonouri, Olga ;
Kovatchev, Boris ;
Kowalski, Aaron ;
Laffel, Lori ;
Levine, Brian ;
Mayorov, Alexander ;
Mathieu, Chantal ;
Murphy, Helen R. ;
Nimri, Revital ;
Norgaard, Kirsten ;
Parkin, Christopher G. ;
Renard, Eric ;
Rodbard, David ;
Saboo, Banshi ;
Schatz, Desmond ;
Stoner, Keaton ;
Urakami, Tatsuiko ;
Weinzimer, Stuart A. ;
Phillip, Moshe .
DIABETES CARE, 2019, 42 (08) :1593-1603
[4]   Limitations of conventional methods of self-monitoring of blood glucose - Lessons learned from 3 days of continuous glucose sensing in pediatric patients with type 1 diabetes [J].
Boland, E ;
Monsod, T ;
Delucia, M ;
Brandt, CA ;
Fernando, S ;
Tamborlane, WV .
DIABETES CARE, 2001, 24 (11) :1858-1862
[5]   Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial [J].
Bolinder, Jan ;
Antuna, Ramiro ;
Geelhoed-Duijvestijn, Petronella ;
Kroeger, Jens ;
Weitgasser, Raimund .
LANCET, 2016, 388 (10057) :2254-2263
[6]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[7]  
Diabetes C, 2016, Diabetes Care, V39, P686, DOI [DOI 10.2337/DC15-1990, 10.2337/dc15-1990]
[8]   Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus [J].
Fokkert, M. J. ;
van Dijk, P. R. ;
Edens, M. A. ;
Abbes, S. ;
de Jong, D. ;
Slingerland, R. J. ;
Bilo, H. J. G. .
BMJ OPEN DIABETES RESEARCH & CARE, 2017, 5 (01)
[9]   User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria and Calculation of Insulin Dosing Errors [J].
Freckmann, Guido ;
Jendrike, Nina ;
Baumstark, Annette ;
Pleus, Stefan ;
Liebing, Christina ;
Haug, Cornelia .
DIABETES THERAPY, 2018, 9 (02) :683-697
[10]   Report of the Committee of the Japan Diabetes Society on the Research of Fulminant and Acute-onset Type 1 Diabetes Mellitus: New diagnostic criteria of fulminant type 1 diabetes mellitus (2012) [J].
Imagawa, Akihisa ;
Hanafusa, Toshiaki ;
Awata, Takuya ;
Ikegami, Hiroshi ;
Uchigata, Yasuko ;
Osawa, Haruhiko ;
Kawasaki, Eiji ;
Kawabata, Yumiko ;
Kobayashi, Tetsuro ;
Shimada, Akira ;
Shimizu, Ikki ;
Takahashi, Kazuma ;
Nagata, Masao ;
Makino, Hideichi ;
Maruyama, Taro .
JOURNAL OF DIABETES INVESTIGATION, 2012, 3 (06) :536-539