Endoscopy-related infection: relic of the past?

被引:34
作者
Seoane-Vazquez, Enrique [1 ,2 ]
Rodriguez-Monguio, Rosa [3 ,4 ]
机构
[1] Ohio State Univ, Coll Pharm, Columbus, OH 43210 USA
[2] Ohio State Univ, Coll Publ Hlth, Ctr Hlth Outcomes Policy & Evaluat Studies, Columbus, OH 43210 USA
[3] Univ Massachusetts, Inst Global Hlth, Amherst, MA 01003 USA
[4] Univ Massachusetts, Sch Publ Hlth & Hlth Sci, Amherst, MA 01003 USA
关键词
cross-infection; endoscopy; equipment contamination; infection control;
D O I
10.1097/QCO.0b013e328301396b
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Purpose of review Recent outbreaks of nosocomial infection and pseudoinfection have been linked to contaminated endoscopes. This review summarizes the recent literature, analyzes the latest published information related to the epidemiology, examines potential causes for the outbreaks, and discusses current alternatives for preventing infection. Recent findings A systematic follow-up of patients revealed that the risk of infection attributed to inadequate endoscope reprocessing was very low. Nevertheless, inadequate reprocessing practices are still considered the main culprit underlying contamination from endoscopy procedures. Moreover, standards of care are difficult to maintain given the numerous inconsistencies that exist among reprocessing guidelines and manufacturer's recommended practices. Exposure to contaminated equipment could be prevented through better reprocessing practices and adherence to decontamination guidelines. Recent literature reinforces the need for endoscopy drying after each reprocessing cycle, endoscope reprocessing after short periods of disuse, surveillance, and for a coordinated approach to handle postcontamination responses. Additional analyses such as health technology assessment and cost analysis are needed to identify control alternatives that are most effective. Summary Although the risk of endoscopy-related infection is very low, continued efforts are needed to ensure that quality is maintained during endoscope reprocessing to reduce the incidence of endoscopy-related infections.
引用
收藏
页码:362 / 366
页数:5
相关论文
共 44 条
[1]   Reprocessing failure [J].
Banerjee, Subhas ;
Nelson, Douglas B. ;
Dominitz, Jason A. ;
Ikenberry, Steven O. ;
Anderson, Michelle A. ;
Cash, Brooks D. ;
Gan, Seng-Ian ;
Harrison, M. Edwyn, III ;
Shen, Bo ;
Baron, Todd H. ;
Van Guilder, Trina ;
Lee, Kenneth K. .
GASTROINTESTINAL ENDOSCOPY, 2007, 66 (05) :869-871
[2]   ESGE-ESGENA guideline for quality assurance in reprocessing: Microbiological surveillance testing in endoscopy [J].
Beilenhoff, U. ;
Neumann, C. S. ;
Rey, J. F. ;
Biering, H. ;
Blum, R. ;
Schmidt, V. .
ENDOSCOPY, 2007, 39 (02) :175-181
[3]   A prospective study of the efficacy of routine decontamination for gastrointestinal endoscopes and the risk factors for failure [J].
Bisset, Linda ;
Cossart, Yvonne E. ;
Selby, Warwick ;
West, Richard ;
Catterson, Denise ;
O'Hara, Kate ;
Vickery, Karen .
AMERICAN JOURNAL OF INFECTION CONTROL, 2006, 34 (05) :274-280
[4]   Nosocomial outbreak of Pseudomonas aeruginosa infections related to a flexible bronchoscope [J].
Bou, R. ;
Aguilar, A. ;
Perpinan, J. ;
Ramos, P. ;
Peris, M. ;
Lorente, L. ;
Zuniga, A. .
JOURNAL OF HOSPITAL INFECTION, 2006, 64 (02) :129-135
[5]   Unusual implication of biopsy forceps in outbreaks of Pseudomonas aeruginosa infections and pseudo-infections related to bronchoscopy [J].
Corne, P ;
Godreuil, S ;
Jean-Pierre, H ;
Jonquet, O ;
Campos, J ;
Jumas-Bilak, E ;
Parer, S ;
Marchandin, H .
JOURNAL OF HOSPITAL INFECTION, 2005, 61 (01) :20-26
[6]  
DeFrances Carol J, 2006, Adv Data, P1
[7]   Hospital disinfection: efficacy and safety issues [J].
Dettenkofer, M ;
Block, C .
CURRENT OPINION IN INFECTIOUS DISEASES, 2005, 18 (04) :320-325
[8]  
Eisen Glenn M, 2007, Gastrointest Endosc Clin N Am, V17, P209, DOI 10.1016/j.giec.2007.01.002
[9]   Quality indicators for gastrointestinal endoscopic procedures: an introduction [J].
Faigel, DO ;
Pike, IM ;
Baron, TH ;
Chak, A ;
Cohen, J ;
Deal, SE ;
Hoffman, B ;
Jacobson, BC ;
Mergener, K ;
Petersen, BT ;
Petrini, JL ;
Rex, DK ;
Safdi, MA .
GASTROINTESTINAL ENDOSCOPY, 2006, 63 (04) :S3-S9
[10]  
*FDA, FDA ENF REP