摘要
A sensor incorporating a three electrodes configuration have been fabricated using low cost screen-printing technology. These electrodes couples with Square Wave Stripping Voltammetry (SWSV) has provided a convenient screening tool for on-site detection of trace levels of Pb(II), Cd(II) and Cu(II). Modification of the graphite carbon surface based on in situ deposition of mercury film has been carried out. By appropriate choice of supporting medium and optimized parameters setting such as amount of mercury used the deposition potential, deposition time, frequency and scan rate, well resolved and reproducible response for Pb(II), Cd(II) and Cu(II) were obtained. The performance characteristics of the developed mercury film screen printed carbon electrode (MFSPCE) for 120 s deposition time showed that the linear range for Cd(II), Pb(II) and Cu(II) were 10 to 200 mu g L(-1). The detection limit recorded for Cd(II), Pb(II) and Cu(II) were 2, 1 and 5 mu g L(-1) with relative standard deviation (RSD) of 6.5%, 6.9% and 7.5%, respectively. Successful applications of the sensing device to waste-water and extracted soil samples were demonstrated.
引用
收藏
页码:1153 / 1163
页数:11
相关论文
共 50 条
  • [1] Simultaneous determination of cadmium(II), lead(II) and copper(II) by using a screen-printed electrode modified with mercury nano-droplets
    Song, Wei
    Zhang, Lei
    Shi, Lei
    Li, Da-Wei
    Li, Yang
    Long, Yi-Tao
    MICROCHIMICA ACTA, 2010, 169 (3-4) : 321 - 326
  • [2] Simultaneous determination of cadmium(II), lead(II) and copper(II) by using a screen-printed electrode modified with mercury nano-droplets
    Wei Song
    Lei Zhang
    Lei Shi
    Da-Wei Li
    Yang Li
    Yi-Tao Long
    Microchimica Acta, 2010, 169 : 321 - 326
  • [3] Determination of lead(II) on a copper/mercury-plated screen-printed electrode
    Zen, JM
    Chung, HH
    Kumar, AS
    ANALYTICA CHIMICA ACTA, 2000, 421 (02) : 189 - 197
  • [4] A Disposable Chitosan Sensor with Carbon Screen-Printed Electrodes for the Detection of Copper (II)
    Chang, Shwu-Jen
    Chen, Ching-Jung
    Liu, Jen-Tsai
    Tseng, Ming-Liang
    BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2019, 124 : 192 - 193
  • [5] Highly sensitive direct simultaneous determination of zinc(II), cadmium(II), lead(II), and copper(II) based on in-situ-bismuth and mercury thin-film plated screen-printed carbon electrode
    Ceren Yıldız
    Dilek Eskiköy Bayraktepe
    Zehra Yazan
    Monatshefte für Chemie - Chemical Monthly, 2021, 152 : 1527 - 1537
  • [6] Highly sensitive direct simultaneous determination of zinc(II), cadmium(II), lead(II), and copper(II) based on in-situ-bismuth and mercury thin-film plated screen-printed carbon electrode
    Yildiz, Ceren
    Bayraktepe, Dilek Eskikoey
    Yazan, Zehra
    MONATSHEFTE FUR CHEMIE, 2021, 152 (12): : 1527 - 1537
  • [7] Sensitive Stripping Determination of Cadmium(II) and Lead(II) on Disposable Graphene Modified Screen-Printed Electrode
    Chao Huangfu
    Fu, Liang
    Li, Yuzhi
    Li, Xueling
    Du, Haijun
    Ye, Jianshan
    ELECTROANALYSIS, 2013, 25 (09) : 2238 - 2243
  • [8] Stripping chronopotentiometric measurements of lead(II) and cadmium(II) in soils extracts and wastewaters using a bismuth film screen-printed electrode assembly
    Rashid O. Kadara
    Ibtisam E. Tothill
    Analytical and Bioanalytical Chemistry, 2004, 378 : 770 - 775
  • [9] Stripping chronopotentiometric measurements of lead(II) and cadmium(II) in soils extracts and wastewaters using a bismuth film screen-printed electrode assembly
    Kadara, RO
    Tothill, IE
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 378 (03) : 770 - 775
  • [10] FOAM SEPARATION OF LEAD(II) AND CADMIUM(II) FROM WASTE-WATER
    FERGUSON, BB
    HINKLE, C
    WILSON, DJ
    SEPARATION SCIENCE, 1974, 9 (02): : 125 - 145