Implementation of a frequency measurement circuit for high-accuracy QCM sensors

被引:3
作者
Dolores Valdes, Maria [1 ]
Moure, Maria Jose [1 ]
Rodriguez, Loreto [1 ]
Farina, Jose [1 ]
机构
[1] Univ Vigo, Dept Elect Technol, Vigo 36310, Spain
来源
2008 4TH SOUTHERN CONFERENCE ON PROGRAMMABLE LOGIC, PROCEEDINGS | 2008年
关键词
D O I
10.1109/SPL.2008.4547765
中图分类号
TP3 [计算技术、计算机技术];
学科分类号
0812 ;
摘要
This paper presents the design of a FPGA-based digital counter for the measurement of the oscillation frequency of Quartz Crystal Microbalance (QCM) sensors. This work is part of a bigger project focused on the development of a standalone QCM system for the analysis and characterization of organic coatings. The method used to implement the digital counter is the direct frequency measurement. Therefore, special attention is paid in the design of the time-base generator. Measurement errors due to the drift of the system oscillator are discussed, and several solutions are proposed taking advantage of the embedded PLL circuits available in the FPGAs. The designed digital counter is implemented and tested using the Altera's Stratix EP1S10F780C6 FPGA.
引用
收藏
页码:233 / 236
页数:4
相关论文
共 7 条
  • [1] *ALT, 446 ALT
  • [2] BEELEY J, 2003, P EUR, V17, P851
  • [3] LAGO E, 2005, IADAT J ADV TECHN AU, V1, P34
  • [4] Moure MJ, 2007, LAT AM APPL RES, V37, P25
  • [5] Peng DG, 2007, IEEE INT CONF CON AU, P131
  • [6] Resolution in quartz crystal oscillator circuits for high sensitivity microbalance sensors in damping media
    Rodriguez-Pardo, L
    Fariña, J
    Gabrielli, C
    Perrot, H
    Brendel, R
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2004, 103 (1-2) : 318 - 324
  • [7] A review of sensor sensitivity and stability
    Vig, JR
    Walls, FL
    [J]. PROCEEDINGS OF THE 2000 IEEE/EIA INTERNATIONAL FREQUENCY CONTROL SYMPOSIUM & EXHIBITION, 2000, : 30 - 33