Radionuclide calibrator measurements of 18F in a 3 ml plastic syringe

被引:16
作者
Cessna, J. T. [1 ]
Schultz, M. K. [1 ]
Leslie, T. [2 ]
Bores, N. [3 ]
机构
[1] NIST, Gaithersburg, MD 20899 USA
[2] Univ Tennessee, Med Ctr, PETNET Solut, Knoxville, TN 37920 USA
[3] Oak Ridge Natl Lab, Oak Ridge, TN 37831 USA
关键词
F-18; liquid scintillation counting; radionuclide calibrator;
D O I
10.1016/j.apradiso.2008.02.046
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
This paper discusses the calibration of Capintec radionuclide calibrators for the measurement of F-18 in the pharmacy and clinic. In support of a planned regional comparison, a secondary laboratory was set up at Oak Ridge National Laboratories in Oak Ridge, TN. The laboratory was used to prepare 1 mL F-18-fluorodeoxyglucose sources in 3 mL plastic BD syringes (Becton, Dickinson and Company, Franklin Lakes, New Jersey, USA) and to determine dial settings in three models of radionuclide calibrators. Measurements were verified by comparing radionuclide calibrator measurements of a 5 mL ampoule source with previous results. The CIEMAT/NIST method of tritium efficiency tracing was used for activity determinations. (c) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:988 / 993
页数:6
相关论文
共 11 条
[1]  
*BIPM, 2004, MON BIPM 5
[2]  
GUNTHER E, 2004, COMMUNICATION
[3]   Comprehensive review and critical evaluation of the half-life of tritium [J].
Lucas, LL ;
Unterweger, MP .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 2000, 105 (04) :541-549
[4]   Assessment of radionuclidic impurities in 2-[18F]fluoro-2-deoxy-D-glucose ([18F]FDG) routine production [J].
Marengo, Mario ;
Lodi, Filippo ;
Magi, Silvia ;
Cicoria, Gianfranco ;
Pancaldi, Davide ;
Boschi, Stefano .
APPLIED RADIATION AND ISOTOPES, 2008, 66 (03) :295-302
[5]  
*NIST, 1991, CERT STAND REF MAT 4
[6]  
QUAIM SM, 2006, NUCL DATA SCI TECHNO, P817
[7]   A performance evaluation of 90Y dose-calibrator measurements in nuclear pharmacies and clinics in the United States [J].
Schultz, Michael K. ;
Cessna, Jeffrey T. ;
Anderson, Tamara L. ;
Ponto, James A. ;
Petry, Neil ;
Kowalsky, Richard J. ;
Palmer, Matthew R. ;
Beinlich, Uwe F. ;
Baker, William ;
Hinkle, George H. ;
Hung, Joseph C. ;
Quinton, Timothy ;
Rice, Peter A. ;
Divgi, Chaitanya ;
Norenberg, Jeffrey P. .
APPLIED RADIATION AND ISOTOPES, 2008, 66 (02) :252-260
[8]  
TAYLOR BN, 1996, 1297A NIST
[9]   Standardization of 63Ni by 4πβ liquid scintillation spectrometry with 3H-standard efficiency tracing [J].
Zimmerman, BE ;
Collé, R .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1997, 102 (04) :455-477
[10]   Experimental determinations of commercial 'dose calibrator' settings for nuclides used in nuclear medicine [J].
Zimmerman, BE ;
Cessna, JT .
APPLIED RADIATION AND ISOTOPES, 2000, 52 (03) :615-619