Surface ionization ion mobility spectrometry

被引:42
作者
Wu, C
Hill, HH [1 ]
Rasulev, UK
Nazarov, EG
机构
[1] Washington State Univ, Dept Chem, Pullman, WA 99164 USA
[2] Uzbek Acad Sci, Arifov Inst Elect, Tashkent 700143, Uzbekistan
关键词
D O I
10.1021/ac9712299
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A surface ionization (SI) source was designed and constructed for ion mobility spectrometry (IMS). Compared with a conventional Ni-63 source, the surface ionization source is as simple and reliable, has an extended dynamic response range, is more selective in response, and does not have regulatory problems associated with radioactive ionization sources. The performance of this SI-IMS was evaluated with several different classes df compounds. Triethylamine was employed for studying the behavior of the ionization source under different source conditions and gaseous environments. Amines, tobacco alkaloids, and triazine herbicides were also investigated. Picogram level detection limits were achieved for target compounds with a response dynamic range of 5 orders of magnitude. Selective monitoring by IMS was also demonstrated. While the surface ionization source does not have the universality of response that is obtained with a Ni-63 ionization source, it is an excellent nonradioactive alternative for the ionization and ion mobility detection of those compounds to which it responds.
引用
收藏
页码:273 / 278
页数:6
相关论文
共 24 条
[1]   OZONATION BY-PRODUCTS OF ATRAZINE IN SYNTHETIC AND NATURAL-WATERS [J].
ADAMS, CD ;
RANDTKE, SJ .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1992, 26 (11) :2218-2227
[2]  
BURYAKOV IA, 1993, ZH ANAL KHIM, V48, P156
[3]  
CARR TW, 1984, PLASMA CHROMATOGRAPH
[4]  
EATHERTON RL, 1987, THESIS WASHINGTON ST
[5]  
Eiceman G.A., 1993, ION MOBILITY SPECTRO
[6]   NEW SENSITIVE AND SELECTIVE DETECTOR FOR GAS-CHROMATOGRAPHY - SURFACE-IONIZATION DETECTOR WITH A HOT PLATINUM EMITTER [J].
FUJII, T ;
ARIMOTO, H .
ANALYTICAL CHEMISTRY, 1985, 57 (13) :2625-2628
[7]   MASS-SPECTROMETRIC STUDIES ON THE RESPONSE MECHANISM OF SURFACE-IONIZATION DETECTORS FOR GAS-CHROMATOGRAPHY [J].
FUJII, T ;
JIMBA, H ;
ARIMOTO, H .
ANALYTICAL CHEMISTRY, 1990, 62 (02) :107-111
[8]   ION MOBILITY SPECTROMETRY [J].
HILL, HH ;
SIEMS, WF ;
STLOUIS, RH ;
MCMINN, DG .
ANALYTICAL CHEMISTRY, 1990, 62 (23) :A1201-A1209
[9]   DETERMINATION OF NICOTINE N-OXIDE BY GAS-CHROMATOGRAPHY FOLLOWING THERMAL-CONVERSION TO 2-METHYL-6-(3-PYRIDYL)TETRAHYDRO-1,2-OXAZINE [J].
JACOB, P ;
BENOWITZ, NL ;
YU, L ;
SHULGIN, AT .
ANALYTICAL CHEMISTRY, 1986, 58 (11) :2218-2221
[10]   ION MOBILITY SPECTROMETRY OF ALIPHATIC AND AROMATIC-AMINES [J].
KARPAS, Z .
ANALYTICAL CHEMISTRY, 1989, 61 (07) :684-689