Fast Bound Pool Fraction Mapping Using Stimulated Echoes

被引:20
作者
Soellinger, M. [1 ]
Langkammer, C. [1 ]
Seifert-Held, T. [1 ]
Fazekas, F. [1 ]
Ropele, S. [1 ]
机构
[1] Med Univ Graz, Dept Neurol, A-8036 Graz, Austria
基金
奥地利科学基金会;
关键词
STEAM; myelin; bound pool fraction; two-pool model; MRI; magnetization transfer; brain; MAGNETIZATION-TRANSFER CONTRAST; WHITE-MATTER; MULTIPLE-SCLEROSIS; CROSS-RELAXATION; HUMAN BRAIN; INVERSION-RECOVERY; TISSUES; PARAMETERS; MRI; DIFFUSION;
D O I
10.1002/mrm.22846
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Magnetization transfer imaging advanced to an indispensible tool for investigating white matter changes. Quantitative magnetization transfer imaging methods allow the determination of the bound pool fraction (BPF), which is thought to be directly linked to myelin integrity. Long acquisition times and high specific absorption rates are still inhibiting broad in vivo utilization of currently available BPF mapping techniques. Herewith, a stimulated echoes amplitude modulation-based, single-shot echo planar imaging technique for BPF and T(1) quantification is presented at 3T. It allows whole brain mapping in 10-15 min and is low in specific absorption rates. The method was validated with different concentrations of bovine serum albumin (BSA) phantoms. Infra- and inter-subject variability was assessed in vivo. Phantom measurements verified linearity between bovine serum albumin concentrations and measured BPF, which was independent of T(1) variations. T(1) values in the phantoms correlated well with values provided by standard T(1) mapping methods. Intrasubject variability was minimal and mean regional BPFs of 10 volunteers (e.g., left frontal white matter = 0.135 +/- 0.003, right frontal white matter = 0.129 +/- 0.006) were in line with previously published data. Assessment of interhemispheric BPF differences revealed significantly higher BPF for the left brain hemisphere. To sum up, these results suggest the proposed method useful for crosssectional and longitudinal studies of white matter changes in the human brain. Magn Reson Med 66:717-724, 2011. (C)2011 Wiley-Liss, Inc.
引用
收藏
页码:717 / 724
页数:8
相关论文
共 35 条
[1]  
AXEL L, 1984, Magnetic Resonance Imaging, V2, P121, DOI 10.1016/0730-725X(84)90066-3
[2]   Investigation of quantitative magnetisation transfer parameters of lesions and normal appearing white matter in multiple sclerosis [J].
Cercignani, M. ;
Basile, B. ;
Spano, B. ;
Comanducci, G. ;
Fasano, F. ;
Caltagirone, C. ;
Nocentini, U. ;
Bozzali, M. .
NMR IN BIOMEDICINE, 2009, 22 (06) :646-653
[3]   High-resolution T1 and T2 mapping of the brain in a clinically acceptable time with DESPOT1 and DESPOT2 [J].
Deoni, SCL ;
Peters, TM ;
Rutt, BK .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (01) :237-241
[4]   High-resolution T1 mapping of the brain at 3T with driven equilibrium single pulse observation of T1 with high-speed incorporation of RF field inhomogeneities (DESPOT1-HIFI) [J].
Deoni, Sean C. L. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 26 (04) :1106-1111
[5]  
DORTCH RD, 2010, P INT SOC MAGN RESON, P335
[6]   CROSS RELAXATION AND SPIN DIFFUSION IN PROTON NMR OF HYDRATED COLLAGEN [J].
EDZES, HT ;
SAMULSKI, ET .
NATURE, 1977, 265 (5594) :521-523
[7]  
Fazekas F, 2000, J NEURAL TRANSM-SUPP, P45
[8]   LIMITATIONS OF STIMULATED ECHO ACQUISITION MODE (STEAM) TECHNIQUES IN CARDIAC APPLICATIONS [J].
FISCHER, SE ;
STUBER, M ;
SCHEIDEGGER, MB ;
BOESIGER, P .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (01) :80-91
[9]   A MULTISPECTRAL ANALYSIS OF BRAIN-TISSUES [J].
FLETCHER, LM ;
BARSOTTI, JB ;
HORNAK, JP .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (05) :623-630
[10]   Characterization of normal appearing brain structures using high-resolution quantitative magnetization transfer steady-state free precession imaging [J].
Garcia, M. ;
Gloor, M. ;
Wetzel, S. G. ;
Radue, E. -W. ;
Scheffler, K. ;
Bieri, O. .
NEUROIMAGE, 2010, 52 (02) :532-537