Reliability and validity of a smartphone-based application for the quantification of the sit-to-stand movement in healthy seniors

被引:44
|
作者
Cerrito, Adrien [1 ]
Bichsel, Lukas [2 ]
Radlinger, Lorenz [1 ]
Schmid, Stefan [1 ]
机构
[1] Bern Univ Appl Sci, Hlth Div, Discipline Physiotherapy, CH-3008 Bern, Switzerland
[2] Bern Univ Appl Sci, Inst Rehabil & Performance Technol, Dept Engn & Informat Technol, CH-3008 Bern, Switzerland
关键词
Chair rising; Fall prevention; Trunk accelerometry; Elderly; OLDER-ADULTS; FALLS; RISK; ACCELEROMETRY; BALANCE; TASK; SET;
D O I
10.1016/j.gaitpost.2014.11.001
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Introduction: Recently, it has been demonstrated that biomechanical parameters measured with accelerometers during everyday activities might provide early information about mobility in seniors. The use of smartphones has been suggested for this purpose. The aim of this study was to develop and validate a smartphone-based application for the quantification of the sit-to-stand movement (STS) in healthy seniors. Methods: Sixteen healthy seniors performed the STS with a smartphone first attached to the lower back and then to the sternum. The subjects' vertical ground reaction forces (VGRF) and vertical acceleration (VAcc) were recorded simultaneously using two force plates (reference standard) and the smartphone, respectively. VAcc data were converted to force; power was then calculated based on both force-time curves. Subsequently, the total movement duration (T-total), peak force (F-max), rate of force development (RFD) and peak power (P-max) were determined. Relative (intraclass correlation coefficient [ICC]) and absolute (standard error of measurement [SEM]) reliability were also calculated. For the variables showing strong reliability (ICC >= 0.75), Pearson correlation coefficients (r) between VGRF- and VAcc-derived data and one-sample T-tests for the mean differences were carried out. Results: In the sternal configuration, all variables showed strong reliability (range: 0.86-0.93), except the smartphone-derived parameter RFD (ICC = 0.43). F-max and T-total were found to be highly valid (r = 0.86 and 0.98, respectively). Conclusion: The results showed that the smartphone-based application can be used as a reliable and valid tool for the quantification of the STS in healthy seniors if the proposed protocol is followed. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:409 / 413
页数:5
相关论文
共 50 条
  • [1] Validation Of Smartphone-based Assessment Of Sit-to-stand Power
    Carnal, Matthew J.
    Beideman, William F.
    Andraca, Diego I.
    Singh, Pavarsdeep
    Tracy, Brian L.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2017, 49 (05): : 312 - 312
  • [2] Smartphone-Based Sit-to-Stand Analysis for Mobility Assessment in Middle Age
    Hayek, Roee
    Gutman, Itai
    Baranes, Guy
    Nudelman, Yaniv
    Springer, Shmuel
    INNOVATION IN AGING, 2024, 8 (10)
  • [3] Validity of accelerometry in assessing the duration of the sit-to-stand movement
    Janssen, Wim G. M.
    Bussmann, Johannes B. J.
    Horemans, Herwin L. D.
    Stam, Henk J.
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2008, 46 (09) : 879 - 887
  • [4] Validity of accelerometry in assessing the duration of the sit-to-stand movement
    Wim G. M. Janssen
    Johannes B. J. Bussmann
    Herwin L. D. Horemans
    Henk J. Stam
    Medical & Biological Engineering & Computing, 2008, 46 : 879 - 887
  • [5] RELIABILITY OF A CLINICAL KINEMATIC ASSESSMENT OF THE SIT-TO-STAND MOVEMENT
    JENG, SF
    SCHENKMAN, M
    ORILEY, P
    LIN, SJ
    PHYSICAL THERAPY, 1990, 70 (08): : 511 - 520
  • [6] An Experimental Study on the Validity and Reliability of a Smartphone Application to Acquire Temporal Variables during the Single Sit-to-Stand Test with Older Adults
    Marques, Diogo Luis
    Neiva, Henrique Pereira
    Pires, Ivan Miguel
    Zdravevski, Eftim
    Mihajlov, Martin
    Garcia, Nuno M.
    Ruiz-Cardenas, Juan Diego
    Marinho, Daniel Almeida
    Marques, Mario Cardoso
    SENSORS, 2021, 21 (06) : 1 - 15
  • [7] Reliability of Recurrence Quantification Analysis Measures for Sit-to-Stand and Stand-to-Sit Activities in Healthy Older Adults Using Wearable Sensors
    Nasim, Amnah
    Nchekwube, David C.
    Kim, Yoon Sang
    ELECTRONICS, 2021, 10 (19)
  • [8] Reliability and Validity of the Sit-to-Stand as a Muscular Power Measure in Older Adults
    Sherwood, Jennifer J.
    Inouye, Cathy
    Webb, Shannon L.
    Jenny, O.
    JOURNAL OF AGING AND PHYSICAL ACTIVITY, 2020, 28 (03) : 455 - 466
  • [9] Development of a Biomechanical Device for Parameter Quantification Associated with the Sit-to-Stand Movement
    Oliveira, Natacha
    Carvalho, Filipe
    Lains, Jorge
    Rasteiro, Deolinda
    Roseiro, Luis
    SENSORS, 2023, 23 (04)
  • [10] Validity and reliability of smartphone-based application for chronic ankle instability
    Yoon, Taelim
    Lee, Jihyun
    INTERNATIONAL JOURNAL OF THERAPY AND REHABILITATION, 2021, 28 (09):