Trace arsenic analysis in edible seaweeds by miniature in situ dielectric barrier discharge microplasma optical emission spectrometry based on gas phase enrichment

被引:0
作者
Zhang, Yaru [1 ,2 ]
Mao, Xuefei [2 ]
Tian, Di [1 ]
Liu, Jixin [2 ,3 ]
Li, Chunsheng [1 ]
机构
[1] Jilin Univ, Coll Instrumentat & Elect Engn, Changchun 130023, Peoples R China
[2] Chinese Acad Agr Sci, Inst Qual Standard & Testing Technol Agroprod, Minist Agr & Rural Affairs, Key Lab Agrofood Safety & Qual, Beijing 100081, Peoples R China
[3] Beijing Abil Tech Co Ltd, Beijing 100081, Peoples R China
关键词
ATOMIC-ABSORPTION-SPECTROMETRY; HYDRIDE GENERATION; ANTIMONY; PRECONCENTRATION; ATOMIZATION; MERCURY; RICE; LEAD; SB;
D O I
10.1039/d1ay01034d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work, a novel method using low-cost miniaturized hydride generation optical emission spectrometry equipment coupled with an in situ dielectric barrier discharge trap (HG-in situ DBD trap-OES) was established for the determination of As in edible seaweed samples. An improved peak volume algorithm, where the start time point and end time point of the spectrum at each concentration are determined according to the unified judgment criteria, was first proposed to extend the linear range from 1-100 mu g L-1 to 1-200 mu g L-1, and increase the sensitivity by about 30%. In addition, a modification was done on the DBD implementation, providing an enhancement of sensitivity by a factor of about 4 for As. All in all the detection limit (LOD) was improved from 0.5 mu g L-1 to 0.2 mu g L-1. By applying the method to seaweed samples, a method detection limit (MD) of 0.25 mg kg(-1) was achieved, with less than 3% relative standard deviations (RSDs). The calibration linearity reached R-2 > 0.990 in the 1.25-250 mg kg(-1) range. Results obtained by the proposed method showed good agreement with that of certified reference materials (CRMs), and spiked recoveries were 103% to 114%, indicating favorable accuracy. The proposed method is attractive in terms of instrumentation size (0.6 m x 0.5 m x 0.3 m), power consumption (<60 W), manufacturing cost, and gas consumption (300 measurements for 4 L compressed Ar/H-2 gas), and therefore more advantageous than conventional atomic spectrometric methods.
引用
收藏
页码:4079 / 4089
页数:11
相关论文
共 37 条
[1]   Preconcentration technique for manganese by adsorption onto a tantalum wire for tungsten tube atomizer electrothermal atomization atomic absorption spectrometry [J].
Amin, Md. Nurul ;
Kaneco, Satoshi ;
Nakano, Yukitake ;
Katsumata, Hideyuki ;
Suzuki, Tohru ;
Ohta, Kiyohisa .
MICROCHIMICA ACTA, 2008, 162 (1-2) :73-79
[2]   Determination of arsenic(III) by flow injection solid phase extraction coupled with on-line hydride generation atomic absorption spectrometry using a PTFE turnings-packed micro-column [J].
Anthemidis, Aristidis N. ;
Martavaltzoglou, Evdoxia K. .
ANALYTICA CHIMICA ACTA, 2006, 573 :413-418
[3]  
Borak J., 2007, J AGR FOOD CHEM, V47, P204
[4]   Spatially and Temporally Resolved Detection of Arsenic in a Capillary Dielectric Barrier Discharge by Hydride Generation High-Resolved Optical Emission Spectrometry [J].
Burhenn, Sebastian ;
Kratzer, Jan ;
Svoboda, Milan ;
Klute, Felix David ;
Michels, Antje ;
Veza, Damir ;
Franzke, Joachim .
ANALYTICAL CHEMISTRY, 2018, 90 (05) :3424-3429
[5]   Determination of arsenic and antimony in milk by hydride generation atomic fluorescence spectrometry [J].
Cava-Montesinos, P ;
Cervera, ML ;
Pastor, A ;
de la Guardia, M .
TALANTA, 2003, 60 (04) :787-799
[6]   Lead determination at ng/mL level by flame atomic absorption spectrometry using a tantalum coated slotted quartz tube atom trap [J].
Demirtas, Ilknur ;
Bakirdere, Sezgin ;
Ataman, O. Yavuz .
TALANTA, 2015, 138 :218-224
[7]   Solution-cathode glow discharge - optical emission spectrometry of a new design and using a compact spectrograph [J].
Doroski, Todd A. ;
King, Allison M. ;
Fritz, Michael P. ;
Webb, Michael R. .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2013, 28 (07) :1090-1095
[8]   Speciation of Antimony in Leaching Solution in Contact with Plastic by Novel Liquid-Liquid Microextraction and Graphite Furnace Atomic Absorption Spectrometry [J].
Fang, Lanyun ;
Wang, Li ;
Li, Jige ;
Gao, Qinlei .
ANALYTICAL LETTERS, 2015, 48 (12) :1954-1964
[9]   On the coupling of hydride generation (HG) with flowing liquid anode atmospheric pressure glow discharge (FLA-APGD) for determination of traces of As, Bi, Hg, Sb and Se by optical emission spectrometry (OES) [J].
Gorska, Monika ;
Greda, Krzysztof ;
Pohl, Pawel .
TALANTA, 2021, 222
[10]   Miniaturized dielectric barrier discharge-atomic emission spectrometer for pesticide: Sensitive determination of thiram after derivatization with mercurial ion [J].
Han, Bingjun ;
Li, Ya ;
He, Yan ;
Lv, Daizhu ;
Peng, Lixu ;
Yu, Huamei .
MICROCHEMICAL JOURNAL, 2018, 138 :457-464