PURPOSE: To analyze the changes in optic surface roughness before and after injection of various intraocular lens (IOL) models using atomic force microscopy (AFM). SETTINGS: Departments of Ophthalmology, Medical University of Graz, General Hospital Linz and University Hospital Basel; Upper Austria University, School of Applied Health and Social Sciences, Linz, Austria. DESIGN: Experimental study. METHODS: The morphology and surface roughness of 3 hydrophobic acrylic IOLs from different manufacturers were analyzed by AFM in liquid using the tapping mode. First, AFM was performed on IOLs taken from the original package without further manipulation. In a second step, under sterile conditions, an experienced cataract surgeon loaded the IOLs into the appropriate injection system and pushed them through a system resembling an IOL implantation in cataract surgery; this was followed by AFM evaluation. Finally, 3 samples of a preloaded hydrophilic acrylic IOL taken from the original cartridge were compared with 3 samples that were pushed through the implantation system. RESULTS: Comparison of the arithmetic mean, standard deviation, root mean square, and surface skewness of the IOLs before and after injection showed a significant increase in surface roughness (P<.05). CONCLUSIONS: Standard application procedures of IOLs may alter the IOL surface. Increases in the surface roughness of IOLs may influence postoperative posterior capsule opacification. Further studies are necessary to evaluate the interfacial properties of IOLs.
引用
收藏
页码:146 / 154
页数:9
相关论文
共 30 条
[1]   Uveal and capsular biocompatibility of 2 foldable acrylic intraocular lenses in patients with uveitis or pseudoexfoliation syndrome - Comparison to a control group [J].
Abela-Formanek, C ;
Amon, M ;
Schauersberger, J ;
Schild, G ;
Kolodjaschna, J ;
Barisani-Asenbauer, T ;
Kruger, A .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (07) :1160-1172
[2]   Eradication of posterior capsule opacification -: Documentation. of a marked decrease in Nd:YAG laser posterior capsulotomy rates noted in an analysis of 5416 pseudophakic human eyes obtained postmortem [J].
Apple, DJ ;
Peng, Q ;
Visessook, N ;
Werner, L ;
Pandey, SK ;
Escobar-Gomez, M ;
Ram, J ;
Auffarth, GU .
OPHTHALMOLOGY, 2001, 108 (03) :505-518
[3]  
Bhatia Sameer, 1997, CLAO Journal, V23, P264
[4]   Effect of an acrylic intraocular lens with a sharp posterior optic edge on posterior capsule opacification [J].
Buehl, W ;
Findl, O ;
Menapace, R ;
Rainer, G ;
Sacu, S ;
Kiss, B ;
Petternel, V ;
Georgopoulos, M .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (07) :1105-1111
[5]   Neodymium:YAG laser posterior capsulotomy after implantation of AcrySof intraocular lenses [J].
Davison, JA .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (07) :1492-1500
[6]   Optical and atomic force microscopy of an explanted AcrySof intraocular lens with glistenings [J].
Dogru, M ;
Tetsumoto, K ;
Tagami, Y ;
Kato, K ;
Nakamae, K .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (04) :571-575
[7]   Acrylic intraocular lens damage after folding using a forceps insertion technique [J].
Erie, Jay C. ;
Newman, Brant ;
Mahr, Michael A. ;
Khan, Amir R. ;
McIntosh, Malcolm .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (03) :483-487
[8]   Surface abnormalities on hydrophilic acrylic intraocular lenses implanted by an injector [J].
Faschinger, CW .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (06) :845-849
[9]   Effect of optic material on posterior capsule opacification in intraocular lenses with sharp-edge optics - Randomized clinical trial [J].
Findl, O ;
Menapace, R ;
Sacu, S ;
Buehl, W ;
Rainer, G .
OPHTHALMOLOGY, 2005, 112 (01) :67-72
[10]   The effect of polymethylmethacrylate, silicone, and polyacrylic intraocular lenses on posterior capsular opacification 3 years after cataract surgery [J].
Hollick, EJ ;
Spalton, DJ ;
Ursell, PG ;
Pande, MV ;
Barman, SA ;
Boyce, JF ;
Tilling, K .
OPHTHALMOLOGY, 1999, 106 (01) :49-54