Longitudinal assessment of quality assurance measurements in a 1.5T MR-linac: Part I-Linear accelerator

被引:12
作者
Subashi, Ergys [1 ]
Lim, Seng Boh [1 ]
Gonzalez, Xesus [2 ]
Tyagi, Neelam [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10065 USA
[2] Elekta Ltd, Stockholm, Sweden
关键词
longitudinal; MR-linac; 1; 5T; quality assurance; ADAPTIVE RADIATION-THERAPY; RELATIVE DOSIMETRY; DOSE CALCULATION; MAGNETIC-FIELDS; MONTE-CARLO; SYSTEM; RADIOTHERAPY;
D O I
10.1002/acm2.13418
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To describe and report longitudinal quality assurance (QA) measurements for the mechanical and dosimetric performance of an Elekta Unity MR-linac during the first year of clinical use in our institution. Materials and methods The mechanical and dosimetric performance of the MR-linac was evaluated with daily, weekly, monthly, and annual QA testing. The measurements monitor the size of the radiation isocenter, the MR-to-MV isocenter concordance, MLC and jaw position, the accuracy and reproducibility of step-and-shoot delivery, radiation output and beam profile constancy, and patient-specific QA for the first 50 treatments in our institution. Results from end-to-end QA using anthropomorphic phantoms are also included as a reference for baseline comparisons. Measurements were performed in water or water-equivalent plastic using ion chambers of various sizes, an ion chamber array, MR-compatible 2D/3D diode array, portal imager, MRI, and radiochromic film. Results The diameter of the radiation isocenter and the distance between the MR/MV isocenters was (mu +/- sigma) 0.39 +/- 0.01 mm and 0.89 +/- 0.05 mm, respectively. Trend analysis shows both measurements to be well within the tolerance of 1.0 mm. MLC and jaw positional accuracy was within 1.0 mm while the dosimetric performance of step-and-shoot delivery was within 2.0%, irrespective of gantry angle. Radiation output and beam profile constancy were within 2.0% and 1.0%, respectively. End-to-end testing performed with ion-chamber and radiochromic film showed excellent agreement with treatment plan. Patient-specific QA using a 3D diode array identified gantry angles with low-pass rates allowing for improvements in plan quality after necessary adjustments. Conclusion The MR-linac operates within the guidelines of current recommendations for linear accelerator performance, stability, and safety. The analysis of the data supports the recently published guidance in establishing clinically acceptable tolerance levels for relative and absolute measurements.
引用
收藏
页码:190 / 201
页数:12
相关论文
共 36 条
[1]  
[Anonymous], 2008, COMM RAD TREATM PLAN
[2]  
[Anonymous], 2001, Absorbed Dose Determination in External Beam Radiotherapy. No. 398 in Technical Reports Series
[3]   First clinical experiences with a high field 1.5 T MR linac [J].
Bertelsen, Anders S. ;
Schytte, Tine ;
Moller, Pia K. ;
Mahmood, Faisal ;
Riis, Hans L. ;
Gottlieb, Karina L. ;
Agergaard, Soren N. ;
Dysager, Lars ;
Hansen, Olfred ;
Gornitzka, Janne ;
Veldhuizen, Elisabeth ;
ODwyer, Dean B. ;
Christiansen, Rasmus L. ;
Nielsen, Morten ;
Jensen, Henrik R. ;
Brink, Carsten ;
Bernchou, Uffe .
ACTA ONCOLOGICA, 2019, 58 (10) :1352-1357
[4]   A daily end-to-end quality assurance workflow for MR-guided online adaptive radiation therapy on MR-Linac [J].
Chen, Xinfeng ;
Ahunbay, Ergun ;
Paulson, Eric S. ;
Chen, Guangpei ;
Li, X. Allen .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (01) :205-212
[5]   Measurement validation of treatment planning for a MR-Linac [J].
Chen, Xinfeng ;
Paulson, Eric S. ;
Ahunbay, Ergun ;
Sanli, Aydin ;
Klawikowski, Slade ;
Li, X. Allen .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (07) :28-38
[6]   Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM [J].
Das, Indra J. ;
Cheng, Chee-Wai ;
Watts, Ronald J. ;
Ahnesjo, Anders ;
Gibbons, John ;
Li, X. Allen ;
Lowenstein, Jessica ;
Mitra, Raj K. ;
Simon, William E. ;
Zhu, Timothy C. .
MEDICAL PHYSICS, 2008, 35 (09) :4186-4215
[7]   Reference dosimetry in MRI-linacs: evaluation of available protocols and data to establish a Code of Practice [J].
de Pooter, Jacco ;
Billas, Ilias ;
de Prez, Leon ;
Duane, Simon ;
Kapsch, Ralf-Peter ;
Karger, Christian P. ;
van Asselen, Bram ;
Wolthaus, Jochem .
PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (05)
[8]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[9]   The Rotating Biplanar Linac-Magnetic Resonance Imaging System [J].
Fallone, Biagio Gino .
SEMINARS IN RADIATION ONCOLOGY, 2014, 24 (03) :200-202
[10]   Task group 284 report: magnetic resonance imaging simulation in radiotherapy: considerations for clinical implementation, optimization, and quality assurance [J].
Glide-Hurst, Carri K. ;
Paulson, Eric S. ;
McGee, Kiaran ;
Tyagi, Neelam ;
Hu, Yanle ;
Balter, James ;
Bayouth, John .
MEDICAL PHYSICS, 2021, 48 (07) :E636-E670