Impedance analyser module for EIT and spectroscopy using undersampling

被引:24
作者
Dudykevych, T [1 ]
Gersing, E [1 ]
Thiel, F [1 ]
Hellige, G [1 ]
机构
[1] Univ Gottingen, Dept Anaesthesiol Res, Ctr Anaesthesiol Emergency & Intens Care Med TL195, D-37075 Gottingen, Germany
关键词
bioimpedance spectroscopy; EIT; undersampling;
D O I
10.1088/0967-3334/22/1/303
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In this paper we present the concept, the design and the test procedure for a DSP-based high-precision and high-performance wide-band (up to 10 MHz) bioimpedance analyser module for application in EIT or bioimpedance spectroscopy. The module implements a digital concept with appropriate signal conditioning hardware for voltage and current measurement, early signal digitization and subsequent digital signal processing in order to calculate the components of impedance (or admittance). At low frequencies, the module utilizes the conventional direct conversion method, whereas at high frequencies the undersampling technique is used. The advantages of the described system are the following: (a) the frequency range is extended to higher frequencies, (b) the number of data sampled per time interval is significantly reduced, and (c) the current consumption and the costs of the ADCs can be significantly reduced. The validation procedure is per-formed by comparing the measured and theoretical values of the magnitude and the phase of the impedance for a commonly used tissue model. The module offers an accuracy of better than 0.012% for the magnitude of impedance and better than 0.02 degrees for the phase.
引用
收藏
页码:19 / 24
页数:6
相关论文
共 13 条
[1]  
BRATU D, 1999, P EUR MED BIOL ENG C, P126
[2]   Dispersion and absorption in dielectrics I. Alternating current characteristics [J].
Cole, KS ;
Cole, RH .
JOURNAL OF CHEMICAL PHYSICS, 1941, 9 (04) :341-351
[3]  
DUDYKEVYCH T, 1999, P EUR MED BIOL ENG C, P130
[4]   A multichannel continuously selectable multifrequency electrical impedance spectroscopy measurement system [J].
Hartov, A ;
Mazzarese, RA ;
Reiss, FR ;
Kerner, TE ;
Osterman, KS ;
Williams, DB ;
Paulsen, KD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2000, 47 (01) :49-58
[5]  
KESTER W, 1996, ADCS HIGH SPEED DESI
[6]  
KESTER W, 1995, PRACTICAL ANALOG DES
[7]  
MILNES P, 1999, P EUR MED BIOL ENG C, P142
[8]  
NEBUYA S, 1999, P EUR MED BIOL ENG C, P84
[9]  
RIGUAD B, 1995, PHYSL MEAS S, V16, pA15
[10]   A model of artefacts produced by stray capacitance during whole body or segmental bioimpedance spectroscopy [J].
Scharfetter, H ;
Hartinger, P ;
Hinghofer-Szalkay, H ;
Hutten, H .
PHYSIOLOGICAL MEASUREMENT, 1998, 19 (02) :247-261