Development and characterization of a neutron tomography system based on image intensifier CCD system

被引:9
|
作者
Sinha, A [1 ]
Shaikh, AM
Shyam, A
机构
[1] Bhabha Atom Res Ctr, High Pressure Phys Div, Purnima Lab, Mumbai 400085, India
[2] Bhabha Atom Res Ctr, Div Solid State Phys, Mumbai 400085, India
关键词
neutron tomography; CCD; image intensifier; CBP method;
D O I
10.1016/S0168-583X(98)00263-8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A neutron tomography system has been developed using a low cost electronic imaging system which uses commercially available Image intensifrer, CCD, Frame grabber. This setup has been used for a series of experiments on neutron tomography at the 400 kW swimming pool type reactor APSARA which has a neutron flux of 10(6) n/cm(2)/s at L/D = 90. The main feature of this tomography system is integrated computer controlled imaging and stepper motor control system with online data acquisition, analysis and display of images for each projection angle. The density profile of the projection images and ray sum calculation can be carried out in offline or online manner at any slice and a series of reconstructed image for any slice can be obtained. An algorithm for reconstruction using Convolution Back Projection (CBP) technique has been developed and implemented on a Pentium system. The imaging system is modular in design. The reconstructed image has shown good quality and holes of less than 0.6 mm filled with wax can be revealed in a 60 mm aluminum matrix or 1 mm holes in 40 mm solid brass matrix have been resolved. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:425 / 431
页数:7
相关论文
共 50 条
  • [41] Comparison of image quality and radiation dose between an image-intensifier system and a newer-generation flat-panel detector system - technical phantom measurements and evaluation of clinical imaging in children
    Weis, Meike
    Hagelstein, Claudia
    Diehm, Theo
    Schoenberg, Stefan O.
    Neff, K. Wolfgang
    PEDIATRIC RADIOLOGY, 2016, 46 (02) : 286 - 292
  • [42] Comparison of image quality and radiation dose between an image-intensifier system and a newer-generation flat-panel detector system — technical phantom measurements and evaluation of clinical imaging in children
    Meike Weis
    Claudia Hagelstein
    Theo Diehm
    Stefan O. Schoenberg
    K. Wolfgang Neff
    Pediatric Radiology, 2016, 46 : 286 - 292
  • [43] Design of Image Processing System Based on FPGA
    Xu Guosheng
    MEMS, NANO AND SMART SYSTEMS, PTS 1-6, 2012, 403-408 : 1281 - 1284
  • [44] New film-based motion analysis system characterized by a data-matrix readout subsystem using CCD image processing
    Deng, NM
    Chen, LY
    Hou, X
    23RD INTERNATIONAL CONGRESS ON HIGH-SPEED PHOTOGRAPHY AND PHOTONICS, PT 1, 1999, 3516 : 349 - 353
  • [45] Characteristics of the neutron/X-ray tomography system at the SANRAD facility in South Africa
    de Beer, FC
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 542 (1-3) : 1 - 8
  • [46] CCD-based two-dimensional pneumatic arm control system
    Wang, Y-T
    Wong, R-H
    Lo, J-H
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART I-JOURNAL OF SYSTEMS AND CONTROL ENGINEERING, 2009, 223 (I5) : 585 - 590
  • [47] Image-based height measuring system
    Wang, Wei-Yen
    Lu, Ming-Chih
    Chuang, Chin-Tun
    Cheng, Jen-Chi
    PROCEEDINGS OF THE 7TH WSEAS INTERNATIONAL CONFERENCE ON SIGNAL PROCESSING, COMPUTATIONAL GEOMETRY AND ARTIFICIAL VISION (ISCGAV'-07), 2007, : 143 - +
  • [48] Projection-angle-dependent distortion correction in high-speed image-intensifier-based x-ray computed tomography
    Sanctorum, Joaquim G.
    Van Wassenbergh, Sam
    Nguyen, Van
    De Beenhouwer, Jan
    Sijbers, Jan
    Dirckx, Joris J. J.
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2021, 32 (03)
  • [49] Measurement System of the Simulated Missile-target Encounter Based on CCD Imaging
    Xu, Fengliang
    Chen, Keshan
    PROCEEDINGS OF THE 2009 2ND INTERNATIONAL CONGRESS ON IMAGE AND SIGNAL PROCESSING, VOLS 1-9, 2009, : 2774 - 2778
  • [50] Study of on-line CCD spectrum data acquisition system based on CPLD
    Li Gang
    Wang Yang
    Li Hai-Ian
    Lin Ling
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2007, 27 (10) : 1905 - 1909