NIST facility for spectral irradiance and radiance responsivity calibrations with uniform sources

被引:86
作者
Brown, SW [1 ]
Eppeldauer, GP [1 ]
Lykke, KR [1 ]
机构
[1] Natl Inst Stand & Technol, Opt Technol Div, Gaithersburg, MD 20899 USA
关键词
D O I
10.1088/0026-1394/37/5/53
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A laser-based facility has been developed to provide high-flux, monochromatic, Lambertian radiation over the spectral range 0.2 mum to 18 mum The facility was designed to reduce the uncertainties in a variety of radiometric applications, including irradiance and radiance responsivity calibrations. The operational characteristics of the facility are discussed and the results of detector responsivity calibrations over the spectral range 0.406 mum to 0.920 mum are presented.
引用
收藏
页码:579 / 582
页数:4
相关论文
共 13 条
[1]   HIGHLY STABLE, MONOCHROMATIC AND TUNABLE OPTICAL RADIATION SOURCE AND ITS APPLICATION TO HIGH-ACCURACY SPECTROPHOTOMETRY [J].
ANDERSON, VE ;
FOX, NP ;
NETTLETON, DH .
APPLIED OPTICS, 1992, 31 (04) :536-545
[2]   Optical characterization of diffuser-input standard irradiance meters [J].
Eppeldauer, G ;
Racz, M ;
Larason, T .
OPTIKA '98 - 5TH CONGRESS ON MODERN OPTICS, 1998, 3573 :220-224
[3]  
EPPELDAUERGP, 2000, METROLOGIA, V37, P531
[4]   High-accuracy aperture-area measurement facilities at the National Institute of Standards and Technology [J].
Fowler, JB ;
Durvasula, RS ;
Parr, AC .
METROLOGIA, 1998, 35 (04) :497-500
[5]   National Institute of Standards and Technology high-accuracy cryogenic radiometer [J].
Gentile, TR ;
Houston, JM ;
Hardis, JE ;
Cromer, CL ;
Parr, AC .
APPLIED OPTICS, 1996, 35 (07) :1056-1068
[6]  
GIBSON CE, 1998, NATL I STAND TECHNOL
[7]   Evaluating the performance of filter radiometers as a means of improving the uncertainty of ultraviolet measurements [J].
Hunt, TM ;
Fox, NP ;
Hartree, WS ;
Durant, NM .
METROLOGIA, 1998, 35 (04) :345-351
[8]  
LARASON TC, 1998, NATL I STAND TECHNOL
[9]   Domain-engineered pyroelectric radiometer [J].
Lehman, J ;
Eppeldauer, G ;
Aust, JA ;
Racz, M .
APPLIED OPTICS, 1999, 38 (34) :7047-7055
[10]   Development of a monochromatic, uniform source facility for calibration of radiance and irradiance detectors from 0.2 μm to 12 μm [J].
Lykke, KR ;
Shaw, PS ;
Hanssen, LM ;
Eppeldauer, GP .
METROLOGIA, 1998, 35 (04) :479-483