Comparison of ultra-performance liquid chromatography and ARK immunoassay for therapeutic drug monitoring of voriconazole

被引:6
|
作者
Pena-Lorenzo, Diego [1 ]
Rebollo, Noemi [1 ,2 ]
German Sanchez-Hernandez, Jose [1 ,2 ]
Zarzuelo-Castaneda, Aranzazu [2 ]
机构
[1] Univ Hosp Salamanca, Pharm Serv, Salamanca, Spain
[2] Univ Salamanca, Pharmaceut Sci Dept, Salamanca, Spain
关键词
Voriconazole; monitoring; immunoassay; UPLC;
D O I
10.1177/00045632211041887
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Therapeutic drug monitoring (TDM) of voriconazole is recommended for personalizing doses. The objective of this study was to compare the enzyme immunoassay developed by ARK Diagnostics Inc. for the quantification of voriconazole adapted to the Architect C4000 autoanalyzer (Abbott (R)) with ultra-performance liquid chromatography using ultraviolet detector (UPLC-UV) method. Methods Linearity, precision and accuracy of both methods were validated according to the Food and Drug Administration (FDA) and European Medicines Agency (EMA) guidelines. The limit of quantification (LOQ) of the UPLC-UV method was determined experimentally. Both methods were applied to the analysis of 62 samples from patients. Correlation was evaluated by Passing-Bablok analysis and the concordance by the Bland-Altman method. Dosage recommendations were generated; the discordances according to the technique were evaluated. Results All validation parameters determined for UPLC-UV met the criteria set out and LOQ of 0.1 mu g/mL was established. However, when the enzyme immunoassay was used to determine concentrations <= 1 mu g/mL, CVs were >20%. A linear correlation between both methods was found. However, an overestimation of immunoassay (systematic error of 0.39 mu g/mL) was detected. In 11.3% of the samples, the differences in concentrations when they were determined by different techniques would imply a different therapeutic regime. These samples had concentrations close to 1 mu g/mL. Conclusion Although both techniques can be used for TDM of voriconazole, when a value close to the lower limit of the therapeutic range is determined by the ARK(TM) immunoassay, it would be better to verify the result by a non-automated technique to avoid possible underdosing.
引用
收藏
页码:657 / 660
页数:4
相关论文
共 50 条
  • [41] Ultra-performance liquid chromatography/mass spectrometry of intact proteins
    Everley, Robert A.
    Croley, Timothy R.
    JOURNAL OF CHROMATOGRAPHY A, 2008, 1192 (02) : 239 - 247
  • [42] Fingerprint analysis of Resina Draconis by ultra-performance liquid chromatography
    Yudi Xue
    Lin Zhu
    Tao Yi
    Chemistry Central Journal, 11
  • [43] Determination of Pralsetinib in Human Plasma and Cerebrospinal Fluid for Therapeutic Drug Monitoring by Ultra-performance Liquid Chromatography-Tandem Mass Spectrometry (UPLC-MS/MS)
    Zhao, Zichen
    Pu, Qianlun
    Sun, Tonglin
    Huang, Qian
    Tong, Liping
    Fan, Ting
    Kang, Jingyue
    Chen, Yuhong
    Zhang, Yan
    ANTI-CANCER AGENTS IN MEDICINAL CHEMISTRY, 2024, 24 (11) : 867 - 877
  • [44] Ultra-Performance Liquid Chromatographic Method for Measurement of Voriconazole in Human Plasma and Oral Fluid
    Boeira, Paula
    Antunes, Marina V.
    Andreolla, Huander F.
    Pasqualotto, Alessandro C.
    Linden, Rafael
    JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2012, 23 (01) : 148 - 155
  • [45] Quantitation of levetiracetam concentrations in plasma and saliva samples by ultra-performance liquid chromatography-tandem mass spectrometry: Application to therapeutic drug monitoring for pregnant women with epilepsy
    Zhang, Mengyu
    Jin, Ying
    Li, Wanling
    He, Chaoqun
    Di, Xiangjie
    Duan, Yifei
    Chen, Lei
    Wang, Zhenlei
    BIOMEDICAL CHROMATOGRAPHY, 2024, 38 (02)
  • [46] Determination of melamine in pet food by enzyme immunoassay, high-performance liquid chromatography with diode array detection, and ultra-performance liquid chromatography with tandem mass spectrometry
    University of Maine, Department of Food Science and Human Nutrition, 5735 Hitchner Hall, Orono, ME 04469-5735
    不详
    不详
    J AOAC Int, 2008, 2 (408-413):
  • [47] Comparison of ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) and enzyme-multiplied immunoassay technique (EMIT) for quantification of voriconazole plasma concentration from Chinese patients
    Yu, Mingjie
    Yang, Jun
    Xiong, Lirong
    Zhan, Shipeng
    Cheng, Lin
    Chen, Yongchuan
    Liu, Fang
    HELIYON, 2023, 9 (11)
  • [48] Determination of melamine in pet food by enzyme immunoassay, high-performance liquid chromatography with diode array detection, and ultra-performance liquid chromatography with tandem mass spectrometry
    Kim, Byungchul
    Perkins, Lewis B.
    Bushway, Rodney J.
    Nesbit, Stephainie
    Fan, Titan
    Sheridain, Robert
    Greene, Virginia
    JOURNAL OF AOAC INTERNATIONAL, 2008, 91 (02) : 408 - 413
  • [49] Ultra-performance convergence chromatography, a more efficient method for chemical quality evaluation of Gaoben medicinal materials compared with the ultra-performance liquid chromatography
    Zhang, Xiaoyi
    Liu, Jiushi
    Gao, Shiman
    Chen, Rongchang
    Qi, Yaodong
    Zhang, Bengang
    Sun, Xiaobo
    Liu, Haitao
    Xiao, Peigen
    JOURNAL OF CHROMATOGRAPHY A, 2020, 1616
  • [50] Comparison of Novel Immunoassay With Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) for Therapeutic Drug Monitoring of Clozapine
    Buckley, Tiffany
    Kitchen, Christopher
    Vyas, Gopal
    Siegfried, Nathan A.
    Tefera, Eshetu
    Chen, Shuo
    DiPaula, Bethany A.
    Kelly, Deanna L.
    THERAPEUTIC DRUG MONITORING, 2020, 42 (05) : 771 - 777