Isotope ratios on transient signals with GC-MC-ICP-MS

被引:45
作者
Krupp, EA [1 ]
Donard, OFX [1 ]
机构
[1] LCABIE, CNRS, F-64053 Pau, France
关键词
MC-ICP-MS; gas chromatography; transient signals; isotope ratios;
D O I
10.1016/j.ijms.2004.11.026
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
ICP-MS instruments with multiple ion collectors (MC-ICP-MS) provide highly precise and accurate isotope ratio measurements. These instruments are easily coupled to chromatography, laser ablation or flow injection devices because of the ICP source, and thus attract general analytical interest for measurement of highly precise element isotope ratios on transient signals. Some few papers have been published today where MC-ICP-MS was employed for isotope ratios on transient signals, and in general, results are encouraging. Isotope ratio precision obtained is usually at least 10-fold superior compared to measurements with ICP-QMS. Nevertheless, most authors describe a drift of isotope ratios during the transient peak. This behaviour was either interpreted as a problem in the sample introduction system, or was related to instrumental mass bias drift, but has to date not been completely evaluated or explained. In this paper, we systematically investigate the drift observed on lead and mercury isotopes during short transient signals obtained from GC coupled to different MC-ICP-MS systems. We can clearly show that neither changes in instrumental mass bias, nor chromatographic fractionation effects are the source of the observed isotope ratio drift. An influence of analyte concentration on the observed drift was as well excluded as a source for this drift, because the slope on isotope ratios show the same values for different concentrations evaluated. In contrast, the peak width was found to influence the extent to which the isotope ratio drifts during peak elution. Thus, the relative intensity change per time seems to be an important factor for the measurements of transient signals with MC-ICP-MS. (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:233 / 242
页数:10
相关论文
共 11 条
[1]   Sources of mass bias and isotope ratio variation in multicollector ICP-MS:: optimization of instrumental parameters based on experimental observations [J].
Andrén, H ;
Rodushkin, I ;
Stenberg, A ;
Malinovsky, D ;
Baxter, DC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2004, 19 (09) :1217-1224
[2]   Uncertainty contributions to species specific isotope dilution analysis. Part 2. Determination of methylmercury by HPLC coupled with quadrupole and multicollector ICP-MS [J].
Clough, R ;
Belt, ST ;
Evans, EH ;
Fairman, B ;
Catterick, T .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2003, 18 (09) :1039-1046
[3]   Measurement of high precision isotope ratios for mercury from coals using transient signals [J].
Evans, RD ;
Hintelmann, H ;
Dillon, PJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (09) :1064-1069
[4]   Measurement of isotope ratios on transient signals by MC-ICP-MS [J].
Günther-Leopold, I ;
Wernli, B ;
Kopajtic, Z ;
Günther, D .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 378 (02) :241-249
[5]  
GUNTHERLEOPOLD I, IN PRESS INT J MASS
[6]   Improvements in precision of isotopic ratio measurements using laser ablation-multiple collector-ICP-mass spectrometry: reduction of changes in measured isotopic ratios [J].
Hirata, T ;
Hayano, Y ;
Ohno, T .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2003, 18 (10) :1283-1288
[7]   Precise isotope-ratio determination by CGC hyphenated to ICP-MCMS for speciation of trace amounts of gaseous sulfur, with SF6 as example compound [J].
Krupp, EM ;
Pécheyran, C ;
Meffan-Main, S ;
Donard, OFX .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 378 (02) :250-255
[8]   Precise isotope-ratio measurements of lead species by capillary gas chromatography hyphenated to hexapole Multicollector ICP-MS [J].
Krupp, EM ;
Pécheyran, C ;
Meffan-Main, S ;
Donard, OFX .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 370 (05) :573-580
[9]   Isotopic precision for a lead species (PbEt4) using capillary gas chromatography coupled to inductively coupled plasma-multicollector mass spectrometry [J].
Krupp, EM ;
Pécheyran, C ;
Pinaly, H ;
Motelica-Heino, M ;
Koller, D ;
Young, SMM ;
Brenner, IB ;
Donard, OFX .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2001, 56 (07) :1233-1240
[10]   Application of isotopically labeled methylmercury for isotope dilution analysis of biological samples using gas chromatography/ICPMS [J].
Martín-Doimeadios, RCR ;
Krupp, E ;
Amouroux, D ;
Donard, OFX .
ANALYTICAL CHEMISTRY, 2002, 74 (11) :2505-2512