Measurement of respirable crystalline silica concentration by X-ray diffraction: Evaluation of metrological performances

被引:1
|
作者
Nebbia, Rebecca [1 ]
Genta, Gianfranco [1 ]
Zuliani, Paolo [2 ]
Patrucco, Mario [3 ]
Pira, Enrico [3 ]
Galetto, Maurizio [1 ]
机构
[1] Politecn Torino, Dept Management & Prod Engn, Corso Duca Abruzzi 24, I-10129 Turin, Italy
[2] AOU Citta Salute & Sci, Corso Bramante 88, I-10126 Turin, Italy
[3] Univ Torino, Dept Publ Hlth & Pediat, Turin, Italy
关键词
Respirable crystalline silica; X-ray diffraction; Measurement uncertainty; OS&H; Occupational safety; health; EXPOSURE; RISK;
D O I
10.1016/j.measurement.2021.109839
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Worker exposure to Respirable Crystalline Silica (RCS) is related with increased risk for lung cancer, so it is important to correctly assess it, through measurements of RCS. There are several methods for the uncertainty evaluation of the concentration of RCS measured by X-ray diffraction, however, none involves the calculation of the sensitivity coefficients of the mathematical measurement model. Hence, the aim of this research work is to propose an improved method, in which the sensitivity coefficients are calculated as partial derivatives of the model. The method was applied to four measurements campaigns in a factory, and the results seem to be comparable with those published in literature. It arose that for low concentrations the major contribution of uncertainty is due to the X-ray diffraction calibration line. This may mean that the minimum number of six calibration test filters, as suggested by the standards, is not sufficient, especially for low concentrations.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Accurate Online Measurement of Iron Concentration in Galvannealed Coating Layers by Shift of X-ray Diffraction Peak
    Aoyama, Tomohiro
    Noro, Hisato
    ISIJ INTERNATIONAL, 2019, 59 (09) : 1599 - 1603
  • [22] Accurate Online Measurement of Iron Concentration in Galvannealed Coating Layers by Shift of X-ray Diffraction Peak
    Aoyama, Tomohiro
    Noro, Hisato
    TETSU TO HAGANE-JOURNAL OF THE IRON AND STEEL INSTITUTE OF JAPAN, 2018, 104 (07): : 363 - 368
  • [23] An improved X-ray diffraction method for cellulose crystallinity measurement
    Ju, Xiaohui
    Bowden, Mark
    Brown, Elvie E.
    Zhang, Xiao
    CARBOHYDRATE POLYMERS, 2015, 123 : 476 - 481
  • [24] On the measurement of austenite in supermartensitic stainless steel by X-ray diffraction
    Tolchard, Julian Richard
    Somme, Astri
    Solberg, Jan Ketil
    Solheim, Karl Gunnar
    MATERIALS CHARACTERIZATION, 2015, 99 : 238 - 242
  • [25] Stress measurement with two-dimensional X-ray diffraction
    He, BB
    THIRD INTERNATIONAL CONFERENCE ON EXPERIMENTAL MECHANICS, 2002, 4537 : 127 - 130
  • [26] Measurement of thin film elastic constants by X-ray diffraction
    Faurie, D
    Renault, PO
    Le Bourhis, E
    Villain, P
    Goudeau, P
    Badawi, F
    THIN SOLID FILMS, 2004, 469 : 201 - 205
  • [27] X-ray diffraction characteristics of five materials for stress measurement
    Zhang, Y-L
    Liu, J-Y
    Wang, J.
    Zeng, Y-J
    JOURNAL OF STRAIN ANALYSIS FOR ENGINEERING DESIGN, 2010, 45 (04): : 319 - 328
  • [28] Evaluation method of precipitated oxygen concentration in low resistivity silicon wafers using x-ray diffraction
    Takeno, Hiroshi
    Mizuno, Michihiro
    Ushio, Satoshi
    Takenaka, Takao
    Materials Science Forum, 1995, 196-201 (pt 4) : 1865 - 1870
  • [29] Evaluation method of precipitated oxygen concentration in low resistivity silicon wafers using X-ray diffraction
    Takeno, H
    Mizuno, M
    Ushio, S
    Takenaka, T
    ICDS-18 - PROCEEDINGS OF THE 18TH INTERNATIONAL CONFERENCE ON DEFECTS IN SEMICONDUCTORS, PTS 1-4, 1995, 196-2 : 1865 - 1869
  • [30] Structure of crystalline and amorphous Ge probed by X-ray absorption and diffraction techniques
    Di Cicco, A
    Principi, E
    Minicucci, M
    De Panfilis, S
    Filipponi, A
    Decremps, F
    Datchi, F
    Itié, JP
    Munsch, P
    Polian, A
    HIGH PRESSURE RESEARCH, 2004, 24 (01) : 93 - 99