Response of radiation protection dosemeters in mixed high-energy photon and electron radiation fields

被引:0
作者
Büermann, L [1 ]
Gargioni, E [1 ]
Kramer, HM [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
关键词
D O I
暂无
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The response of radiation protection dosemeters in terms of the phantom-related operational quantities H-p(10) and H ' (10,0 degrees) was measured for personal and area monitoring systems in mixed high-energy electron and photon radiation fields with energies up to 7 MeV. Using mixed radiation fields composed of different fractions of charged particle and photon fluence. three conditions were produced at the point of measurement: charged particle equilibrium (CPE) (a), a lack (b) and an excess (c) of charged particles relative to the conditions of CPE. Personal and area dosemeters of different types were investigated under conditions (a)-(c). A large variability of the response of the different dosemeter types was observed. The results are presented and discussed.