Value of SiPM PET in myocardial perfusion imaging using Rubidium-82

被引:9
作者
Koenders, S. S. [1 ,4 ]
van Dalen, J. A. [2 ]
Jager, P. L. [1 ]
Knollema, S. [1 ]
Timmer, J. R. [3 ]
Mouden, M. [3 ]
Slump, H. [4 ]
van Dijk, J. D. [1 ]
机构
[1] Isala Hosp, Dept Nucl Med, Zwolle, Netherlands
[2] Isala Hosp, Dept Med Phys, Zwolle, Netherlands
[3] Isala Hosp, Dept Cardiol, Zwolle, Netherlands
[4] Univ Twente, Tech Med Ctr, Enschede, Netherlands
关键词
SiPM PET; Rb-82; myocardial blood flow; PET; myocardial perfusion imaging; TIME-OF-FLIGHT; BLOOD-FLOW; POSITION PAPER; QUANTIFICATION; CT; MISREGISTRATION; ATTENUATION; ARTIFACTS; DIAGNOSIS; DISEASE;
D O I
10.1007/s12350-020-02141-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background PET scanners using silicon photomultipliers with digital readout (SiPM PET) have an improved temporal and spatial resolution compared to PET scanners using conventional photomultiplier tubes (PMT PET). However, the effect on image quality and visibility of perfusion defects in myocardial perfusion imaging (MPI) is unknown. Our aim was to determine the value of a SiPM PET scanner in MPI. Methods We prospectively included 30 patients who underwent rest and regadenoson-induced stress Rubidium-82 (Rb-82) MPI on the D690 PMT PET (GE Healthcare) and within three weeks on the Vereos SiPM PET (Philips Healthcare). Two expert readers scored the image quality and assessed the existence of possible defects. In addition, interpreter's confidence, myocardial blood flow (MBF), and myocardial flow reserve (MFR) values were compared. Results Image quality improved (P = 0.03) using the Vereos as compared to the D690. Image quality of the Vereos and the D690 was graded fair in 20% and 10%, good in 60% and 50%, and excellent in 20% and 40%, respectively. Defect interpretation and interpreter's confidence did not differ between the D690 and the Vereos (P > 0.50). There were no significant differences in rest MBF (P >= 0.29), stress MBF (P >= 0.11), and MFR (P >= 0.51). Conclusion SiPM PET provides an improved image quality in comparison with PMT PET. Defect interpretation, interpreter's confidence, and absolute blood flow measurements were comparable between both systems. SiPM PET is therefore a reliable technique for MPI using Rb-82.
引用
收藏
页码:204 / 212
页数:9
相关论文
共 31 条
[1]   Comparison of image quality and lesion detection between digital and analog PET/CT [J].
Alfonso Lopez-Mora, Diego ;
Flotats, Albert ;
Fuentes-Ocampo, Francisco ;
Camacho, Valle ;
Fernandez, Alejandro ;
Ruiz, Agusti ;
Duch, Joan ;
Sizova, Marina ;
Domenech, Anna ;
Estorch, Montserrat ;
Carrio, Ignasi .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (06) :1383-1390
[2]   Impact of point spread function modeling and time-of-flight on myocardial blood flow and myocardial flow reserve measurements for rubidium-82 cardiac PET [J].
Armstrong, Ian S. ;
Tonge, Christine M. ;
Arumugam, Parthiban .
JOURNAL OF NUCLEAR CARDIOLOGY, 2014, 21 (03) :467-474
[3]   Multisection CT Protocols: Sex- and Age-specific Conversion Factors Used to Determine Effective Dose from Dose-Length Product [J].
Deak, Paul D. ;
Smal, Yulia ;
Kalender, Willi A. .
RADIOLOGY, 2010, 257 (01) :158-166
[4]   Will 3-dimensional PET-CT enable the routine quantification of myocardial blood flow? [J].
dekemp, Robert A. ;
Yoshinaga, Keiichiro ;
Beanlands, Rob S. B. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (03) :380-397
[5]   Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: A definitive analysis of causes, consequences, and corrections [J].
Gould, K. Lance ;
Pan, Tinsu ;
Loghin, Catalin ;
Johnson, Nils P. ;
Guha, Ashrith ;
Sdringola, Stefano .
JOURNAL OF NUCLEAR MEDICINE, 2007, 48 (07) :1112-1121
[6]  
Grant K, 2015, J NUCL MED, V56
[7]   Studies of a Next-Generation Silicon-Photomultiplier-Based Time-of-Flight PET/CT System [J].
Hsu, David F. C. ;
Ilan, Ezgi ;
Peterson, William T. ;
Uribe, Jorge ;
Lubberink, Mark ;
Levin, Craig S. .
JOURNAL OF NUCLEAR MEDICINE, 2017, 58 (09) :1511-1518
[8]   Positron range in PET imaging: an alternative approach for assessing and correcting the blurring [J].
Jodal, L. ;
Le Loirec, C. ;
Champion, C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (12) :3931-3943
[9]   Routine Clinical Quantitative Rest Stress Myocardial Perfusion for Managing Coronary Artery Disease Clinical Relevance of Test-Retest Variability [J].
Kitkungvan, Danai ;
Johnson, Nils P. ;
Roby, Amanda E. ;
Patel, Monika B. ;
Kirkeeide, Richard ;
Gould, K. Lance .
JACC-CARDIOVASCULAR IMAGING, 2017, 10 (05) :565-577
[10]   Impact of regadenoson-induced myocardial creep on dynamic Rubidium-82 PET myocardial blood flow quantification [J].
Koenders, S. S. ;
van Dijk, J. D. ;
Jager, P. L. ;
Ottervanger, J. P. ;
Slump, C. H. ;
van Dalen, J. A. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2019, 26 (03) :719-728