Rapid determination of immunosuppressive drug concentrations in whole blood by coated blade spray-tandem mass spectrometry (CBS-MS/MS)

被引:52
作者
Gomez-Rios, German Augusto [1 ]
Tascon, Marcos [1 ]
Reyes-Garces, Nathaly [1 ]
Boyaci, Ezel [1 ]
Poole, Justen James [1 ]
Pawliszyn, Janusz [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Immunosuppressive drugs; Mass spectrometry; Coated blade spray; SPME-MS; SOLID-PHASE MICROEXTRACTION; LC-MS/MS; AMBIENT IONIZATION; COMPLEX MATRICES; TARGET ANALYTES; CYCLOSPORINE-A; HPLC-MS/MS; QUANTITATION; QUANTIFICATION; PERFORMANCE;
D O I
10.1016/j.aca.2017.10.016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Coated Blade Spray (CBS) is a technology that efficiently integrates sample preparation and direct coupling to mass spectrometry (MS) on a single device. In this article, we present CBS-tandem mass spectrometry (CBS-MS/MS) as a novel tool for the rapid and simultaneous determination of four commonly used immunosuppressive drugs (ISDs) in whole blood: tacrolimus (TAC) and cyclosporine-A (CycA), which are calcineurin inhibitors; and sirolimus (SIR) and everolimus (EVR), which are both mTOR (mechanistic target of rapamycin) inhibitors. Given that CBS extracts via free concentration, analytes that are largely bound to plasma proteins or red blood cells provide considerably lower extraction recovery rates. Therefore, we defy the solventless philosophy of SPME-based techniques, like CBS, by performing the analyte-enrichment step via direct immersion in a solvent-modified matrix. The assay was linear within the evaluated range of concentrations (between 1 and 100 ng/mL for EVR/SIR/TAC and 10-1000 ng/mL for CycA), and the limits of quantification were determined to be 10 ng/mL for CycA and 1 ng/mL for EVR/SIR/TAC. Good accuracy (87-119%) and linearity (r(2) >= 0.99) were attained over the evaluated range for all ISDs. Interassay imprecision (CV) determined from incurred sample reanalysis was <= 10% for all ISDs. Our method was validated using Liquichek (TM) whole blood immunosuppressant quality control (QC) standards purchased from Bio-Rad. Concentrations determined by CBS-MS/MS were inside the range specified by Bio-Rad and within 15% of the expected mean value for all ISDs at all QC levels. Furthermore, the effect of different hematocrit levels (20,45, and 70%) in the entire calibration range was carefully studied. No statistical differences (RSD <= 7%) in the calibration curve slopes of ISDs in blood were observed. CBS offers a simpler workflow than that of traditional methods; it eliminates the need for chromatographic separation and provides a clean extract that allows for long-term MS instrumental operation with minimal maintenance. Additionally, because CBS integrates all analytical steps into one device, it eliminates the risk of instrumental carry-over and can be used as a low-cost disposable device for sample preparation and analysis. Fully-automated sample preparation simplifies the method and allows for total analysis times as short as 3 min with turn-around times of less than 90 min. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:69 / 75
页数:7
相关论文
共 31 条
  • [1] Calibrant Free Sampling and Enrichment with Solid-Phase Microextraction: Computational Simulation and Experimental Verification
    Alam, Md. Nazmul
    Ricardez-Sandoval, Luis
    Pawliszyn, Janusz
    [J]. INDUSTRIAL & ENGINEERING CHEMISTRY RESEARCH, 2017, 56 (13) : 3679 - 3686
  • [2] Annesley T.M., 2013, CLIN CHEM, P59
  • [3] Buchwald Armin, 2012, BMC Clin Pharmacol, V12, P2, DOI 10.1186/1472-6904-12-2
  • [4] Quantification of Tacrolimus and Three Demethylated Metabolites in Human Whole Blood Using LC-ESI-MS/MS
    Dubbelboer, Ilse R.
    Pohanka, Anton
    Said, Rana
    Rosenborg, Staffan
    Beck, Olof
    [J]. THERAPEUTIC DRUG MONITORING, 2012, 34 (02) : 134 - 142
  • [5] Ambient Ionization Mass Spectrometry for Point-of-Care Diagnostics and Other Clinical Measurements
    Ferreira, Christina R.
    Yanne, Karen E.
    Jarmusch, Alan K.
    Pirro, Valentina
    Ouyang, Zheng
    Cooks, R. Graham
    [J]. CLINICAL CHEMISTRY, 2016, 62 (01) : 99 - 110
  • [6] Development of Coated Blade Spray Ionization Mass Spectrometry for the Quantitation of Target Analytes Present in Complex Matrices
    Gomez-Rios, German Augusto
    Pawliszyn, Janusz
    [J]. ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2014, 53 (52) : 14503 - 14507
  • [7] Grote-Koska D, 2015, THER DRUG MONIT, V37, P400, DOI 10.1097/FTD.0000000000000139
  • [8] High-throughput LC-MS/MS method for monitoring sirolimus and everolimus in the routine clinical laboratory
    Ivanova, Mariela
    Artusi, Carlo
    Polo, Giulia
    Zaninotto, Martina
    Plebani, Mario
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 (07) : 1151 - 1158
  • [9] Effective Use of Mass Spectrometry in the Clinical Laboratory
    Jannetto, Paul J.
    Fitzerald, Robert L.
    [J]. CLINICAL CHEMISTRY, 2016, 62 (01) : 92 - 98
  • [10] Ultra-fast cyclosporin A quantitation in whole blood by Laser Diode Thermal Desorption - Tandem Mass Spectrometry; comparison with High Performance Liquid Chromatography-Tandem Mass Spectrometry
    Jourdil, Jean-Francois
    Picard, Pierre
    Meunier, Cecile
    Auger, Serge
    Stanke-Labesque, Francoise