Dosimetric characteristics of the 192Ir high-dose-rate afterloading brachytherapy source

被引:10
作者
Sadeghi, Mahdi [1 ]
Taghdiri, Fatemah [2 ]
Saidi, Pooneh [2 ]
机构
[1] Nucl Sci & Technol Res Inst, Agr Med & Ind Res Sch, POB 31485-498, Tehran, Iran
[2] Islamic Azad Univ, Dept Nucl Engn, Sci & Res Branch, Tehran, Iran
关键词
Monte Carlo calculation; Brachytherapy; TG-43U1; MCNP5; Ir-192; MONTE-CARLO CALCULATION; PARAMETERS;
D O I
10.1007/s11604-011-0562-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose. For the treatment of some cancerous tumors using brachytherapy, an American Association of Physicists in Medicine (AAPM) Task Group No. 43U1 report recommends that the dosimetric parameters of a new brachytherapy source must be determined in two experimental and Monte Carlo theoretical methods before using each new source clinically. This study presents the results of Monte Carlo calculations of the dosimetric parameters for a Ir2.A85-2 brachytherapy source design. Materials and methods. Version 5 of the (MCNP) Monte Carlo radiation transport code was used to calculate the dosimetry parameters around the source. Results. The Monte Carlo calculated dose rate constant, I >, of the Ir2.A85-2 source was found to be 1.113 +/- 0.033 cGyU(-1)h(-1). Also in this study, the line-source radial dose function, g(l) (r) and the anisotropy function, F(r,theta), have been calculated at distances from 0.5 to 10 cm. The results of these calculations have been compared with the published data for the same source. Conclusion. All the results are in good concordance with previously published data, with a few exceptions in small angles and short distances. The dosimetric parameters calculated in this work can be used as input data in a treatment planning system (TPS) for exact brachytherapy treatment planning or to verify the calculations of the TPS used in brachytherapy.
引用
收藏
页码:324 / 329
页数:6
相关论文
共 21 条
[1]   Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source [J].
Angelopoulos, A ;
Baras, P ;
Sakelliou, L ;
Karaiskos, P ;
Sandilos, P .
MEDICAL PHYSICS, 2000, 27 (11) :2521-2527
[2]   Monte Carlo dosimetric study of best industries and alpha omega Ir-192 brachytherapy seeds [J].
Ballester, F ;
Granero, D ;
Pérez-Calatayud, J ;
Casal, E ;
Puchades, V .
MEDICAL PHYSICS, 2004, 31 (12) :3298-3305
[3]  
*CROSS SECT EV WOR, 2000, BNLNCS17541 NAT LAB
[4]   Monte Carlo study of the dose rate distributions for the Ir2.A85-2 and Ir2.A85-1 Ir-192 afterloading sources [J].
Granero, D. ;
Perez-Calatayud, J. ;
Ballester, F. .
MEDICAL PHYSICS, 2008, 35 (04) :1280-1287
[5]   Monte Carlo calculation of the TG-43 dosimetric parameters of a new BEBIG Ir-192HDR source [J].
Granero, D ;
Pérez-Calatayud, J ;
Ballester, F .
RADIOTHERAPY AND ONCOLOGY, 2005, 76 (01) :79-85
[6]   Monte Carlo dosimetry of a new 192Ir pulsed dose rate brachytherapy source [J].
Karaiskos, P ;
Angelopoulos, A ;
Pantelis, E ;
Papagiannis, P ;
Sakelliou, L ;
Kouwenhoven, E ;
Baltas, D .
MEDICAL PHYSICS, 2003, 30 (01) :9-16
[7]  
King R P, 2001, J Appl Clin Med Phys, V2, P69, DOI 10.1120/1.1349413
[8]   Dosimetric prerequisites for routine clinical use of photon emitting brachytherapy sources with average energy higher than 50 kev [J].
Li, Zuofeng ;
Das, Rupak K. ;
DeWerd, Larry A. ;
Ibbott, Geoffrey S. ;
Meigooni, Ali S. ;
Perez-Calatayud, Jose ;
Rivard, Mark J. ;
Sloboda, Ronald S. ;
Williamson, Jeffrey F. .
MEDICAL PHYSICS, 2007, 34 (01) :37-40
[9]   Dosimetric characteristics of a new high-intensity 192Ir source for remote afterloading [J].
Meigooni, AS ;
Kleiman, MT ;
Johnson, JL ;
Mazloomdoost, D ;
Ibbott, GS .
MEDICAL PHYSICS, 1997, 24 (12) :2008-2013
[10]  
*MONT CARL TEAM, MCNP A GEN MONT CARL