共 4 条
Elemental PGNAA analysis using gamma-gamma coincidence counting with the library least-squares approach
被引:12
作者:
Metwally, WA
[1
]
Gardner, RP
[1
]
Mayo, CW
[1
]
机构:
[1] N Carolina State Univ, CEAR, Raleigh, NC 27695 USA
来源:
关键词:
PGNAA;
gamma-gamma coincidence;
D O I:
10.1016/S0168-583X(03)01660-4
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
An accurate method for determining elemental analysis using gamma-gamma coincidence counting is presented. To demonstrate the feasibility of this method for PGNAA, a system of three radioisotopes (Na-24, Co-60 and Cs-134) that emit coincident gamma rays was used. Two HPGe detectors were connected to a system that allowed both singles and coincidences to be collected simultaneously. A known mixture of the three radioisotopes was used and data was deliberately collected at relatively high counting rates to determine the effect of pulse pile-up distortion. The results obtained, with the library least-squares analysis, of both the normal and coincidence counting are presented and compared to the known amounts. The coincidence results are shown to give much better accuracy. It appears that in addition to the expected advantage of reduced background, the coincidence approach is considerably more resistant to pulse pile-up distortion. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:394 / 399
页数:6
相关论文