Analysis of in-plane signal-to-noise ratio in computed tomography

被引:0
作者
Hara, Takanori [1 ,2 ]
Ichikawa, Katsuhiro [2 ]
Sanada, Shigeru [2 ]
Ida, Yoshihiro [3 ]
机构
[1] Nakatsugawa Municipal Gen Hosp, Dept Med Technol, 1522-1 Komanba, Gifu 5080011, Japan
[2] Kanazawa Univ, Grad Med Sci, Div Hlth Sci, Kanazawa, Ishikawa 9200942, Japan
[3] Fujita Hlth Univ Hosp, Dept Radiol, Toyoake, Aichi 4701192, Japan
来源
MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3 | 2008年 / 6913卷
关键词
computed tomography (CT); signal-to-noise ratio (SNR); modulation transfer function (MTF); noise power specta (NPS); in-plane; scanning area;
D O I
10.1117/12.768968
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purposes of this study are to analyze signal-to-noise ratio (SNR) changes for in-plane (axial plane) position and in-plane direction in X-ray computed tomography (CT) system and to verify those visual effects by using simulated small low-contrast disc objects. Three-models of multi detector-row CT were employed. Modulation transfer function (MTF) was obtained using a thin metal wire. Noise power spectrum (NPSs) was obtained using a cylindrical water phantom. The measurement positions were set to center and off-centered positions of 64mm, 128mm and 192mm. One-dimensional MTFs and NPSs for the x- and y-direction were calculated by means of a numerical slit scanning method. SNRs were then calculated from MTFs and NPSs. The simulated low-contrast disc objects with diameter of 2 to 10mm and contrast to background of 3.0%, 4.5% and 6.0% were superimposed on the water phantom images. Respective simulated objects in the images are then visually evaluated in degree of their recognition, and then the validity of the resultant SNRs are examined. Resultant in-plane SNRs differed between the center and peripheries and indicated a trend that the SNR values increase in accordance with distance from the center. The increasing degree differed between x- and y-direction, and also changed by the CT systems. These results suggested that the peripheries region has higher low-contrast detectability than the center. The properties derived in this study indicated that the depiction abilities at various in-plane positions are not uniform in clinical CT images, and detectability of the low contrast lesion may be influenced.
引用
收藏
页数:9
相关论文
共 13 条
[1]  
[Anonymous], 2006, J ICRU
[2]  
BISCHOF C J, 1977, Medical Physics (Woodbury), V4, P163, DOI 10.1118/1.594305
[3]   Determination of the presampled MTF in computed tomography [J].
Boone, JM .
MEDICAL PHYSICS, 2001, 28 (03) :356-360
[4]  
ENDO M, 1980, Nippon Acta Radiologica, V40, P43
[5]   Spiral interpolation algorithms for multislice spiral CT - Part II: Measurement and evaluation of slice sensitivity profiles and noise at a clinical multislice system [J].
Fuchs, T ;
Krause, J ;
Schaller, S ;
Flohr, T ;
Kalender, WA .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (09) :835-847
[6]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .1. MODULATION TRANSFER-FUNCTION [J].
GIGER, ML ;
DOI, K .
MEDICAL PHYSICS, 1984, 11 (03) :287-295
[7]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .2. NOISE WIENER SPECTRUM [J].
GIGER, ML ;
DOI, K ;
METZ, CE .
MEDICAL PHYSICS, 1984, 11 (06) :797-805
[8]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .7. NOISE WIENER SPECTRA OF II-TV DIGITAL IMAGING-SYSTEMS [J].
GIGER, ML ;
DOI, K ;
FUJITA, H .
MEDICAL PHYSICS, 1986, 13 (02) :131-138
[9]   ABSOLUTE MEASURES OF PHYSICAL IMAGE QUALITY - MEASUREMENT AND APPLICATION TO RADIOGRAPHIC MAGNIFICATION [J].
SANDRIK, JM ;
WAGNER, RF .
MEDICAL PHYSICS, 1982, 9 (04) :540-549
[10]   Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology [J].
Siewerdsen, JH ;
Antonuk, LE ;
ElMohri, Y ;
Yorkston, J ;
Huang, W ;
Boudry, JM ;
Cunningham, IA .
MEDICAL PHYSICS, 1997, 24 (01) :71-89