Augmented Reality Visualization With Image Overlay for MRI-Guided Intervention: Accuracy for Lumbar Spinal Procedures With a 1.5-T MRI System

被引:49
作者
Fritz, Jan [1 ]
U-Thainual, Paweena [2 ,3 ,4 ]
Ungi, Tamas [5 ]
Flammang, Aaron J. [6 ]
Cho, Nathan B. [2 ,3 ]
Fichtinger, Gabor [5 ]
Iordachita, Iulian I. [2 ,3 ]
Carrino, John A. [1 ]
机构
[1] Johns Hopkins Univ, Sch Med, Russell H Morgan Dept Radiol & Radiol Sci, Musculoskeletal Div, Baltimore, MD 21287 USA
[2] Johns Hopkins Univ, Dept Mech Engn, Baltimore, MD 21287 USA
[3] Johns Hopkins Univ, Lab Computat Sensing & Robot, Baltimore, MD 21287 USA
[4] Queens Univ, Dept Mech & Mat Engn, Kingston, ON, Canada
[5] Queens Univ, Sch Comp, Kingston, ON, Canada
[6] CAMI, Baltimore, MD USA
关键词
augmented reality guidance; diskography; epidural injection; interventional MRI; MR-guided spine injection; LOW-BACK-PAIN; INJECTION PROCEDURES; CT FLUOROSCOPY; FIELD ORIENTATION; NEEDLE INSERTION; GUIDANCE; FEASIBILITY; BIOPSY; TIME; ARTHROGRAPHY;
D O I
10.2214/AJR.11.6918
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to prospectively evaluate the accuracy of an augmented reality image overlay system in MRI-guided spinal injection procedures. MATERIALS AND METHODS. An augmented reality prototype was used in conjunction with a 1.5-T MRI system. A human lumbar spine phantom was used in which 62 targets were punctured to assess the accuracy of the system. Sixty anatomic targets (facet joint, disk space, and spinal canal) were punctured to assess how the accuracy of the system translated into practice. A visualization software interface was used to compare planned needle paths and final needle locations on coregistered CT images (standard of reference). Outcome variables included entry error, angle error, depth error, target error, successful access of anatomic targets, number of needle adjustments, and time requirements. RESULTS. Accuracy assessments showed entry error of 1.6 +/- 0.8 mm, angle error of 1.6 +/- 1.0, depth error of 0.7 +/- 0.5 mm, and target error of 1.9 +/- 0.9 mm. All anatomic targets (60 of 60 insertions) were successfully punctured, including all 20 facet joints, all 20 disks, and all 20 spinal canals. Four needle adjustments (6.7%) were required. Planning of a single needle path required an average of 55 seconds. A single needle insertion required an average of 1 minute 27 seconds. CONCLUSION. The augmented reality image overlay system evaluated facilitated accurate MRI guidance for successful spinal procedures in a lumbar spine model. It exhibited potential for simplifying the current practice of MRI-guided lumbar spinal injection procedures.
引用
收藏
页码:W266 / W273
页数:8
相关论文
共 37 条
  • [1] Blanco Sequeiros Roberto, 2005, Magn Reson Imaging Clin N Am, V13, P519
  • [2] Flexible Add-on Solution for MR Image-Guided Interventions in a Closed-Bore Scanner Environment
    Busse, Harald
    Garnov, Nikita
    Thoermer, Gregor
    Zajonz, Dirk
    Gruender, Wilfried
    Kahn, Thomas
    Moche, Michael
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (03) : 922 - 928
  • [3] Spinal injection procedures: Volume, provider distribution, and reimbursement in the US medicare population from 1993 to 1999
    Carrino, JA
    Morrison, WB
    Parker, L
    Schweitzer, ME
    Levin, DC
    Sunshine, JH
    [J]. RADIOLOGY, 2002, 225 (03) : 723 - 729
  • [4] Performance of interventions with manipulator-driven real-time MR guidance: implementation and initial in vitro tests
    Christoforou, Eftychios
    Akbudak, Erbil
    Ozcan, Alpay
    Karanikolas, Menelaos
    Tsekos, Nikolaos V.
    [J]. MAGNETIC RESONANCE IMAGING, 2007, 25 (01) : 69 - 77
  • [5] Sample size estimation: How many individuals should be studied?
    Eng, J
    [J]. RADIOLOGY, 2003, 227 (02) : 309 - 313
  • [6] Image overlay guidance for needle insertion in CT scanner
    Fichtinger, G
    Deguet, A
    Masamune, K
    Balogh, E
    Fischer, GS
    Mathieu, H
    Taylor, RH
    Zinreich, SJ
    Fayad, LM
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (08) : 1415 - 1424
  • [7] Fischer GS, 2006, STUD HEALTH TECHNOL, V119, P150
  • [8] Real-time MR fluoroscopy-navigated lumbar facet joint injections:: feasibility and technical properties
    Fritz, J.
    Clasen, S.
    Boss, A.
    Thomas, C.
    Koenig, C. W.
    Claussen, C. D.
    Pereira, P. L.
    [J]. EUROPEAN RADIOLOGY, 2008, 18 (07) : 1513 - 1518
  • [9] MR-Guided pain therapy: Principles and clinical applications
    Fritz, J.
    Pereira, P. L.
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2007, 179 (09): : 914 - 924
  • [10] Evaluation of MR imaging guided steroid injection of the sacroiliac joints for the treatment of children with refractory enthesitis-related arthritis
    Fritz, J.
    Tzaribachev, N.
    Thomas, C.
    Carrino, J. A.
    Claussen, C. D.
    Lewin, J. S.
    Pereira, P. L.
    [J]. EUROPEAN RADIOLOGY, 2011, 21 (05) : 1050 - 1057