Background/aim: The Heidelberg retina flowmeter (HRF) is designed to measure retinal capillary blood flow. Previous studies however showed weak reproducibility of data. The intraindividual reproducibility of circadian HRF measurements was examined in healthy subjects in three locations of the retina. Methods: 36 healthy volunteers (27.3 (SD 4.3) years) were examined by HRF seven times a day (t0 - t6). Using a default window of 10610 pixels, three consecutive measurements were performed in three precise focusing planes: superficial, intermediate and deep layer, peripapillary retina, neuroretinal rim and cup, respectively. Images of identical tissue locations identified by capillary landmarks of each layer were selected to quantify the retinal microcirculation of each volunteer. Means and standard deviations of all flow results of a given subject were calculated, at t0 - t6 and the coefficients of variation as a measure of reproducibility. Results: The coefficients of variation ranged between 8.4% and 41.0% in the superficial layer ( mean 19.8% ( SD 8.4%)), 10.6%, and 43.0% in the intermediate layer ( mean 24.0% ( SD 8.4%)), and 9.9% and 84.0% ( mean 29.6% ( SD 15.8%)) in the deep layer. Conclusions: These data show the best reproducibility of measurements in the superficial layer followed by the intermediate and the deep layer. Clinically, this is an unsatisfactory intraindividual reproducibility of flow values in each studied layer.
引用
收藏
页码:348 / 353
页数:6
相关论文
共 29 条
[11]   Heidelberg retinal flowmetry: factors affecting blood flow measurement [J].
Kagemann, L ;
Harris, A ;
Chung, HS ;
Evans, D ;
Buck, S ;
Martin, B .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1998, 82 (02) :131-136
[12]  
Kagemann L, 2001, INVEST OPHTH VIS SCI, V42, P354
[13]  
LESTER M, 2002, J GLAUCOMA, V11, P488
[14]  
Martin B., 1997, Investigative Ophthalmology and Visual Science, V38, pS1049
[15]  
Michael MS, 1996, T MET FINISH INDIA, V5, P21
[16]  
Michelson G, 1996, J GLAUCOMA, V5, P91
[17]   Automatic full field analysis of perfusion images gained by scanning laser Doppler flowmetry [J].
Michelson, G ;
Welzenbach, J ;
Pal, I ;
Harazny, J .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1998, 82 (11) :1294-1300
[18]  
Michelson G, 1996, J GLAUCOMA, V5, P99
[19]   Relationship between ocular pulse pressures and retinal vessel velocities [J].
Michelson, G ;
Harazny, J .
OPHTHALMOLOGY, 1997, 104 (04) :664-671
[20]  
Michelson G, 1995, Ger J Ophthalmol, V4, P21