Reproducibility of circadian retinal and optic nerve head blood flow measurements by Heidelberg retina flowmetry

被引:22
作者
Jonescu-Cuypers, CP
Harris, A
Bartz-Schmidt, KU
Kagemann, L
Boros, AS
Heimann, UE
Lenz, BH
Hilgers, RD
Krieglstein, GK
机构
[1] Univ Cologne, Dept Ophthalmol, Cologne, Germany
[2] Indiana Univ, Dept Ophthalmol, Glaucoma Res & Diagnost Ctr, Indianapolis, IN 46204 USA
[3] Univ Tubingen, Dept Ophthalmol, Tubingen, Germany
[4] Univ Aachen, Rhein Westfal TH Aachen, Inst Biometry, Aachen, Germany
关键词
D O I
10.1136/bjo.2003.024885
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background/aim: The Heidelberg retina flowmeter (HRF) is designed to measure retinal capillary blood flow. Previous studies however showed weak reproducibility of data. The intraindividual reproducibility of circadian HRF measurements was examined in healthy subjects in three locations of the retina. Methods: 36 healthy volunteers (27.3 (SD 4.3) years) were examined by HRF seven times a day (t0 - t6). Using a default window of 10610 pixels, three consecutive measurements were performed in three precise focusing planes: superficial, intermediate and deep layer, peripapillary retina, neuroretinal rim and cup, respectively. Images of identical tissue locations identified by capillary landmarks of each layer were selected to quantify the retinal microcirculation of each volunteer. Means and standard deviations of all flow results of a given subject were calculated, at t0 - t6 and the coefficients of variation as a measure of reproducibility. Results: The coefficients of variation ranged between 8.4% and 41.0% in the superficial layer ( mean 19.8% ( SD 8.4%)), 10.6%, and 43.0% in the intermediate layer ( mean 24.0% ( SD 8.4%)), and 9.9% and 84.0% ( mean 29.6% ( SD 15.8%)) in the deep layer. Conclusions: These data show the best reproducibility of measurements in the superficial layer followed by the intermediate and the deep layer. Clinically, this is an unsatisfactory intraindividual reproducibility of flow values in each studied layer.
引用
收藏
页码:348 / 353
页数:6
相关论文
共 29 条
[1]  
Bohdanecka Z, 1998, J GLAUCOMA, V7, P151
[2]  
Chauhan BC, 1997, J GLAUCOMA, V6, P237
[3]  
Chung HS, 1999, INVEST OPHTH VIS SCI, V40, P2448
[4]   Peripapillary retinal blood flow in normal tension glaucoma [J].
Chung, HS ;
Harris, A ;
Kagemann, L ;
Martin, B .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1999, 83 (04) :466-469
[5]   Heidelberg retina flowmeter parameters at the papilla in healthy subjects [J].
Griesser, SM ;
Lietz, A ;
Orgül, S ;
Schötzau, A ;
Hendrickson, P ;
Flammer, J ;
Haefliger, IO .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 1999, 9 (01) :32-36
[6]  
HARRIS A, 1997, INVEST OPHTHALMOL S, V38, P439
[7]   Evaluation of the peripapillary circulation in healthy and glaucoma eyes with scanning laser Doppler flowmetry [J].
Hollo, G ;
Greve, EL ;
vandenBerg, TJTP ;
Vargha, P .
INTERNATIONAL OPHTHALMOLOGY, 1997, 20 (1-3) :71-77
[8]  
Hollo Gabor, 1996, International Ophthalmology, V20, P63
[9]   Detector sensitivity influences blood flow sampling in scanning laser Doppler flowmetry [J].
Hosking, SL ;
Embleton, S ;
Kagemann, L ;
Chabra, A ;
Jonescu-Cuypers, C ;
Harris, A .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2001, 239 (06) :407-410
[10]   New neuroretinal rim blood flow evaluation method combining Heidelberg retina flowmetry and tomography [J].
Jonescu-Cuypers, CP ;
Chung, HS ;
Kagemann, L ;
Ishii, Y ;
Zarfati, D ;
Harris, A .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (03) :304-309