Visibility of microcalcification clusters and masses in breast tomosynthesis image volumes and digital mammography: A 4AFC human observer study

被引:53
作者
Timberg, P. [1 ]
Bath, M. [2 ,3 ]
Andersson, I. [1 ]
Mattsson, S. [1 ]
Tingberg, A. [1 ,4 ]
Ruschin, M. [5 ]
机构
[1] Lund Univ, Skane Univ Hosp, Dept Clin Sci Malmo, SE-20502 Malmo, Sweden
[2] Sahlgrens Univ Hosp, Dept Med Phys & Biomed Engn, SE-41345 Gothenburg, Sweden
[3] Univ Gothenburg, Sahlgrenska Acad, Dept Radiat Phys, Inst Clin Sci, SE-41345 Gothenburg, Sweden
[4] Skane Univ Hosp, Dept Radiat Phys, SE-20502 Malmo, Sweden
[5] Princess Margaret Hosp, Univ Hlth Network, Radiat Med Program, Toronto, ON M5G 2M9, Canada
关键词
FILTERED BACKPROJECTION; DETECTOR; PERFORMANCE; SYSTEM; PARAMETERS; QUALITY; LESIONS; NOISE; SIZE;
D O I
10.1118/1.3694105
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the visibility of simulated lesions in digital breast tomosynthesis (BT) image volumes compared with 2D digital mammography (DM). Methods: Simulated lesions (masses and microcalcifications) were added to images of the same women acquired on a DM system (Mammomat Novation, Siemens) and a BT prototype. The same beam quality was used for the DM and BT acquisitions. The total absorbed dose resulting from a 25-projection BT acquisition and reconstruction (BT25) was approximately twice that of a single DM view. By excluding every other projection image from the reconstruction (BT13), approximately the same dose as in DM was effected. Simulated microcalcifications were digitally added with varying contrast to the DM and BT images. Simulated masses with 8 mm diameter were also added to BT images. A series of 4-alternative forced choice (4AFC) human observer experiments were conducted. Four medical physicists participated in all experiments, each consisting of 60 trials per experimental condition. The observers interpreted the BT image volumes in cine-mode at a fixed image sequence speed. The required threshold contrast (S-t) to achieve a detectability index (d') of 2.5 (i.e., 92.5% correct decisions) was determined. Results: The St for mass detection in DM was approximately a factor of 2 higher than required in BT indicating that the detection of masses was improved under BT conditions compared to DM. St for microcalcification detection was higher for BT than for DM at both BT dose levels (BT25 and BT13), with a statistically significant difference in St between DM and BT13. These results indicate a dose-dependent decrease in detection performance in BT for detection of microcalcifications. Conclusions: In agreement with previous investigations, masses of size 8 mm can be detected with less contrast in BT than in DM indicating improved detection performance for BT. However, for the investigated microcalcifications, the results of this study indicate potentially worse performance for BT than for DM at the same dose level. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.3694105]
引用
收藏
页码:2431 / 2437
页数:7
相关论文
共 25 条
[1]  
[Anonymous], 2003, BREAST IM REP DAT SY
[2]  
[Anonymous], 1998, NEMA STAND PUBL PS, V20037
[3]  
BAETZ L, 2005, P 7 INT WORKSH DIG M, P23
[4]   Digital breast tomosynthesis using an amorphous selenium flat panel detector [J].
Bissonnette, M ;
Hansroul, M ;
Masson, E ;
Savard, S ;
Cadieux, S ;
Warmoes, P ;
Gravel, D ;
Agopyan, J ;
Polischuk, B ;
Haerer, W ;
Mertelmeier, T ;
Lo, JY ;
Chen, Y ;
Dobbins, JT ;
Jesneck, JL ;
Singh, S .
Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 :529-540
[5]   Search for lesions in mammograms: Statistical characterization of observer responses [J].
Bochud, FO ;
Abbey, CK ;
Eckstein, MP .
MEDICAL PHYSICS, 2004, 31 (01) :24-36
[6]   Estimation of the noisy component of anatomical backgrounds [J].
Bochud, FO ;
Valley, JF ;
Verdun, FR ;
Hessler, C ;
Schnyder, P .
MEDICAL PHYSICS, 1999, 26 (07) :1365-1370
[7]   Scatter/primary in mammography: Comprehensive results [J].
Boone, JM ;
Lindfors, KK ;
Cooper, VN ;
Seibert, JA .
MEDICAL PHYSICS, 2000, 27 (10) :2408-2416
[8]   Effect of detector element size on signal detectability in digital mammography [J].
Burgess, AE .
MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2005, 5745 :232-242
[9]   Effect of toto circle and mammographic background size on detection performance [J].
Burgess, AE ;
Judy, PF .
MEDICAL IMAGING 2002: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2002, 4686 :43-46
[10]   COMPARISON OF RECEIVER OPERATING CHARACTERISTIC AND FORCED-CHOICE OBSERVER PERFORMANCE-MEASUREMENT METHODS [J].
BURGESS, AE .
MEDICAL PHYSICS, 1995, 22 (05) :643-655